GB/T 11081-2005 Determination of Ultraviolet Absorbance of White Oil

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A Reference

Question Key Answer
Standard number and name GB/T 11081-2005 Determination of ultraviolet absorbance of white oil
Which standard it replaces Replaces GB/T 11081-1989 “Determination of ultraviolet absorbance of white oil”
Which foreign standard it adopts Modified adoption of ASTM D 2269:1999
Applicable objects Cosmetic, medical, and food-grade white oils
Non-applicable objects White oils containing additives that are soluble in dimethyl sulfoxide and can exhibit fluorescence or fluorescence quenching
Method principle Extract the sample with dimethyl sulfoxide, and measure the ultraviolet absorbance of the extract at 260 nm to 420 nm
Extractant Dimethyl sulfoxide
Dilution/reference solvent n-Hexane
Measurement wavelength range 260 nm to 420 nm
Absorption cell path length 1 cm (1.00 cm±0.005 cm)
Separatory funnel specifications 125 mL, pear-shaped, with glass stopper, polytetrafluoroethylene stopcock
Room temperature requirement Not lower than 20 ℃
Temperature of sample extract and reference solution Should be the same
Reporting precision Accurate to 3 decimal places
Repeatability Exceeds 0.014 absorbance units only once in 20 times
Reproducibility Exceeds 0.044 absorbance units only once in 20 times
Wavelength range of precision data 275 nm to 280 nm
Treatment of samples containing inhibitors Correct according to Appendix B (normative appendix)
Nature of Appendix A Informative appendix, list of technical differences and their reasons
Nature of Appendix B Normative appendix, correction of inhibitor content

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.