GB/T 12582-1990 Determination of electrical conductivity of liquid hydrocarbons (precision electrometer method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Place the sample in a clean conductivity measuring cell and connect it in series with a direct-current power supply and a sensitive direct-current ammeter. The electrical conductivity is automatically calculated based on Ohm's law from the conductivity measuring cell constant, the voltage applied to the measuring cell, and the nearly instantaneous peak current reading.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

8.1 Calibrate the electrometer

8.2 Verify the measuring cell constant

8.3 Clean the conductivity measuring cell

8.4 Clean auxiliary equipment

8.5 Test procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

Judge the reliability of the test results according to the following provisions (95% confidence level).

11. Report

The conductivity of the sample is expressed in pS/m, and the sample temperature is reported.

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
Standard number and name? GB/T 12582-90 Determination of conductivity of liquid hydrocarbons (precision electrometer method).
Card type? Test method.
Scope of application? "Static" conductivity of aviation fuels and other similar low-conductivity liquid hydrocarbons; laboratory or field.
Measurement range? 0.1~2 000 pS/m.
Principle of the method? The sample is placed in a clean measuring cell, connected in series with a DC power supply and a sensitive DC ammeter; based on Ohm's law, it is automatically calculated from the cell constant, voltage, and near-instantaneous current peak reading.
Sample amount? At least 0.7 L, clean and transparent.
Key instruments? Conductivity measuring cell, electrometer, triaxial cable, thermometer, glass tube, porcelain dish.
Electrometer range and accuracy? 0~20 000 pS/m; ±0.5%±1 digits.
Requirements for n-heptane? Analytical reagent grade, conductivity lower than 1 pS/m.
Requirements for cyclohexane? Analytical reagent grade, dielectric constant at the calibration temperature within the literature value range of 2.023±5%.
Silica gel specification? 100~200 mesh.
Nitrogen purity? Ordinary nitrogen, purity above 95%.
Acceptance criteria for cleaning the measuring cell? Low-conductivity sample: after treatment, the conductivity of n-heptane is lower than 0.05 pS/m; for samples greater than 1 pS/m: less than 0.1 pS/m.
When is the measuring cell constant verified? It must be verified only when the measuring cell is damaged.
Formula for the measuring cell constant? K = 8.854/CA; CA = (Cs-CE)/(εs-1).
New measuring cell constant? 1.0.
Standing time before testing low-conductivity samples? When expected to be less than 1 pS/m, let stand for 15 min before testing.
Reading stability and display? After pressing the key, the reading stabilizes after 3 s and continues to display 9 s.
Interval for repeated testing? 1 min.
Temperature reading accuracy? Read accurately to 1°C.
Expression of results? Conductivity is expressed in pS/m, and the sample temperature is reported.
Precision criteria? 95% confidence level; the difference between repeatability and reproducibility shall not exceed the value in Figure 2.
Zero error correction? When the zero error is greater than ±0.01 and has not been adjusted, subtract the positive zero error from the reading or add the negative zero error.
Referenced and adopted standard? ASTM D4308-83.
Drafting unit? Responsible for drafting: Research Institute of Petroleum Processing; main drafters: Lai Guangcheng, Bi Zaijun.