GB/T 15181-1994 Volumetric calibration method for spherical metal tanks (circumferential tape method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

No centralized statement of the principle chapter is provided in the available source. From Chapters 7 and 8, it can be summarized as follows: by directly measuring the equatorial circumference or vertical circumference with a steel circumferential tape (or indirectly deriving the equatorial radius by measuring a circumference deviating from the equator or by the optical plumb-line method), and combining parameters such as plate thickness, total internal height of the tank, span correction, temperature, and oil gauge tape correction, the equatorial internal diameter, vertical internal diameter, and total internal height of the tank are calculated, and then the total volume of the empty tank (20℃), pressure-bearing volume, etc. are calculated (Chapters 7 and 8).

4. Reagents and materials (purity / preparation)

Not provided in the available source (this standard is a geometric measurement method and does not list reagents).

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

Not provided in the available source (this is on-site calibration, with no sampling or preparation content).

8. Analysis procedure

9. Qualitative and quantitative determination and calculation formulas

10. Precision

11. Report

12. Reference documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
What tanks does this standard apply to? Spherical metal tanks that are newly built or have undergone major repair and have not contained liquid (Chapter 1)
What tanks does it not apply to? Double-layer spherical metal tanks (Chapter 1)
When should calibration be carried out? After passing the pressure test and before insulating the tank body (4.2)
What tension is applied during tape measurement? 50 N (7.1.1.2, 7.1.3.4, 7.2.1)
What is the tolerance for two tape measurements? Not exceeding one ten-thousandth of their arithmetic mean (7.1.1.3, 7.2.1)
How many measurements are averaged for direct measurement of the equatorial circumference? The arithmetic mean of two consecutive measurements (7.1.1.3)
How to measure when the support column is higher than the equator? Measure the circumference deviating from the equator Cₘ and the deviation distance h, and indirectly calculate the equatorial circumference (7.1.2)
How to measure the equatorial radius when tape measurement conditions are not available? Optical plumb line method (7.1.3)
Tolerance for optical plumb line inspection? The difference between two readings on the same horizontal straightedge shall not exceed 1 mm (7.1.3.1)
How to determine the ground projection of the tank center? 120° Set three points A, B, C at intervals, use a theodolite to draw tangent projection lines forming a triangle, and take the intersection of the angle bisectors O' (7.1.3.2)
How many measuring stations are arranged for equatorial radius measurement? Eight plumb line measuring stations are evenly arranged along the ground projection of the equator (7.1.3.3)
How many vertical circumferences are measured? The first is C₂, and intersecting with the first at 90° measure the second C₃ (7.2.1, 7.2.2)
How is the crossing correction defined? The difference between the unobstructed arc length and the obstructed arc length (7.3.1)
To what precision is the total height reading inside the tank? Read accurately to 1 mm (7.4.1, 7.4.2)
What is the reading precision for plate thickness and arc length? Thickness read to 0.1 mm, arc length estimated to 0.5 mm (7.5)
What value is taken for the linear expansion coefficient of the steel tape? For general steel tape 11.5×10⁻⁶ ℃⁻¹, for stainless steel tape 14×10⁻⁶ ℃⁻¹ (8.2.2)
What value is taken for Young's modulus of elasticity of carbon steel? 206×10³ MPa (8.5.1, original text “206×10” requires confirmation against the source standard (source-text extraction))
At what temperature is the total volume of the empty tank expressed? 20℃ (8.4, Appendix C)
What is the total uncertainty? Less than 0.2% (95% confidence level) (Chapter 10)
In what format is the report issued? Refer to Appendix C (Chapter 9)
How is the liquid level height correction value calculated? ΔH=H₁−(H₂+S_L) (Appendix A Formula A1)
What is the volume table temperature correction formula? V_t=V₂₀+π·h·R²·α(t−20)×10⁻⁶ (Appendix C, requires confirmation against the source standard (source-text extraction))
What should be done after major repair or severe deformation of the tank body? Recalibration must be applied for (Appendix C Note 4)
How are the volumes of accessories handled? May be carried out according to the provisions of GB 13235.1 (Appendix C Note 2)
Which standards are cited? GB 12337, GB/T 13235.1, GB 13236 (Chapter 2)

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.