GB/T 1884-2000 Standard Test Method for Density of Crude Petroleum and Liquid Petroleum Products by Laboratory Determination (Hydrometer Method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Bring the sample to the specified temperature, pour it into a density measuring cylinder at approximately the same temperature, place a suitable hydrometer into the temperature-adjusted sample, and let it stand. After the temperature has reached equilibrium, read the hydrometer scale reading and the sample temperature. Use the petroleum measurement tables to convert the observed hydrometer reading to standard density. If necessary, place the density measuring cylinder together with the contained sample in a constant-temperature bath to avoid excessive temperature variation during the determination.

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

Table 1 Technical requirements for hydrometers

Model Unit Density range Scale interval per unit Maximum scale error Meniscus correction value
SY-02 kg/m³ (20℃) 600~1 100 20 ±0.2 +0.3
SY-05 kg/m³ (20℃) 600~1 100 50 ±0.3 +0.7
SY-10 kg/m³ (20℃) 600~1 100 50 ±0.6 +1.4
SY-02 g/cm³ (20℃) 0.600~1.100 0.02 ±0.000 2 +0.000 3
SY-05 g/cm³ (20℃) 0.600~1.100 0.05 ±0.000 3 +0.000 7
SY-10 g/cm³ (20℃) 0.600~1.100 0.05 ±0.001 0 +0.001 4

Note: A SY-1 type or SY-Ⅱ type petroleum hydrometer may be used (in the original, source-text extraction is “SY-I type”, requires confirmation against the source standard (source-text extraction)).

Table 2 Technical requirements for thermometers

Range, ℃ Scale interval Maximum error range
-1~38 0.1 ±0.1
-20~102 0.2 ±0.15

Note: A resistance thermometer may be used, provided its accuracy is not lower than the uncertainty of the above thermometers.

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

Table 3 Repeatability

Petroleum product Temperature range, ℃ Unit Repeatability
Transparent -2~24.5 kg/m³ 0.5
Transparent -2~24.5 g/cm³ 0.000 5
Low viscosity -2~24.5 kg/m³ 0.6
Low viscosity -2~24.5 g/cm³ 0.000 6
Opaque -2~24.5 kg/m³ 0.6
Opaque -2~24.5 g/cm³ 0.000 6

Table 4 Reproducibility

Petroleum product Temperature range, ℃ Unit Reproducibility
Transparent -2~24.5 kg/m³ 1.2
Transparent -2~24.5 g/cm³ 0.001 2
Low viscosity -2~24.5 kg/m³ 1.5
Low viscosity -2~24.5 g/cm³ 0.001 5
Opaque -2~24.5 kg/m³ 1.5
Opaque -2~24.5 g/cm³ 0.001 5

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer
What does this standard measure? Density of crude oil, petroleum products, and mixtures of petroleum products and non-petroleum products at 20℃, with RVP less than 100 kPa.
Which international standard is equivalently adopted? ISO 3675:1998.
Which version does it replace? GB/T 1884—1992.
Issue/implementation dates? Issued on 2000-04-03, implemented on 2000-07-01.
What shall the hydrometer comply with? SH/T 0316 and the technical requirements in Table 1.
Inner diameter requirement for the hydrometer cylinder? At least 25 mm larger than the outer diameter of the hydrometer.
Distance between the bottom of the hydrometer and the bottom of the cylinder? At least 25 mm.
Temperature control accuracy of the constant-temperature bath? Within ±0.25℃.
Ambient temperature change requirement? Not greater than 2℃; if greater than ±2℃, a constant-temperature bath shall be used.
Temperature recording accuracy? To the nearest 0.1℃.
How to judge temperature stability? If the difference from the initial test temperature is greater than 0.5℃, re-read until stable within ±0.5℃.
Where to read for transparent liquids? The point where the lower meniscus of the liquid is tangent to the hydrometer scale.
Where to read for opaque liquids? The point where the upper meniscus of the liquid is tangent to the hydrometer scale, and apply meniscus correction according to Table 1.
To what precision is the reading taken? To the nearest 1/5 of the scale interval.
Recording precision of hydrometer reading? 0.1 kg/m³ (0.000 1 g/cm³).
Reporting precision of results? 0.1 kg/m³ (0.000 1 g/cm³), 20℃.
What is used to convert to standard density? GB/T 1885 Table 59A (crude oil), Table 59B (petroleum products), Table 59D (lubricating oils).
What is used for density conversion between 20℃ and 15℃? GB/T 1885 Table E1 and Table E2.
According to what standard is sampling performed? GB/T 4756.
What is the calibration interval for the hydrometer? Recalibrate at least once every five years.
Repeatability (transparent, -2~24.5℃)? 0.5 kg/m³ (0.000 5 g/cm³).
Reproducibility (transparent, -2~24.5℃)? 1.2 kg/m³ (0.001 2 g/cm³).
Repeatability (opaque, -2~24.5℃)? 0.6 kg/m³ (0.000 6 g/cm³).
Reproducibility (opaque, -2~24.5℃)? 1.5 kg/m³ (0.001 5 g/cm³).
Heating requirements for waxy crude oil? When the pour point is higher than 10℃ or the cloud point is higher than 15℃, heat to more than 9℃ above the pour point or more than 3℃ above the cloud point.
Heating requirements for waxy distillate oil? Heat to more than 3℃ above the cloud point.
Heating requirements for residual fuel oil? Before mixing, heat to the test temperature (see 7.2.1 and its Note 2).
How should volatile samples with RVP greater than 50 kPa be mixed? Mix in the original container and closed system.
Test temperature for crude oil? 20℃, or more than 9℃ above the pour point, or more than 3℃ above the cloud point, whichever is higher.
After high-temperature use of a lead-shot wax-sealed hydrometer? After use above 38℃, dry vertically and cool.
The test report shall at least include? Substance type and identification, national standard number, test results, hydrometer reading and corresponding test temperature, situations differing from this method, test date.