GB/T 262-2010 Standard Test Method for Aniline Point and Mixed Aniline Point of Petroleum Products and Hydrocarbon Solvents

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

A specified volume of aniline and the test specimen, or aniline and the test specimen plus n-heptane, is placed in a test tube, and the mixture is stirred. The mixture is heated at a controlled rate until the two phases in the mixture are completely miscible. Then the mixture is cooled at a controlled rate, and the temperature at which the two phases separate is recorded as the aniline point or mixed aniline point of the test specimen. (Chapter 4)

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and test specimen preparation

8. Analytical Procedure

Method One (7.3)

Method Two (8.3)

Method Three (9.3)

9. Qualitative and Quantitative Calculation Formulas

10. Precision

11. Report

The test report shall include at least the following contents (Chapter 12):

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Points of Answer
How many methods does this standard have? Three: Method One (manual, light-colored transparent), Method Two (automatic apparatus, light-colored and dark), Method Three (volatile samples).
What samples is Method One applicable to? Transparent samples whose initial boiling point is higher than room temperature, whose aniline point is lower than the bubble point of the aniline-sample mixture and higher than its condensation point, and whose color determined in accordance with GB/T 6540 is not greater than 6.5.
What samples is Method Two applicable to? Samples applicable to Method One and extremely dark samples that Method One cannot determine.
What samples is Method Three applicable to? Samples that volatilize significantly at the aniline point temperature, such as aviation gasoline.
Definition of aniline point? The minimum equilibrium solution temperature of a mixture of equal volumes of aniline and the sample to be tested, expressed as ℃.
Definition of mixed aniline point? The minimum equilibrium solution temperature of a mixture of two volumes of aniline, one volume of the sample to be tested, and one volume of n-heptane, expressed as ℃.
Definition of bubble point? The temperature at which bubbles first begin to appear in the mixture when heated under standard conditions.
Method one sample amount? 10 mL aniline + 10 mL test sample; mixed aniline point is 10 mL aniline + 5 mL test sample + 5 mL n-heptane.
Method three sample amount? 5 mL aniline + 5 mL test sample.
Aniline purity requirement? Analytical reagent; distilled on the day of use, discard the first and last 10% (volume fraction) fractions.
n-Heptane purity requirement? Not lower than 99.75%.
Potassium hydroxide specification? Chemical reagent, used for drying aniline.
What is the desiccant? Industrial anhydrous sodium sulfate or calcium sulfate, calcined, placed in a desiccator to cool.
Thermometer model? GB-75 (−38 ℃~42 ℃), GB-76 (25 ℃~105 ℃), GB-77 (90 ℃~170 ℃), graduation value all 0.2 ℃, immersion depth 50 mm.
Can water be used in the heating bath? No, aniline readily absorbs water, and moisture will lead to erroneous results.
Cooling rate? 0.5 ℃/min~1.0 ℃/min.
Heating rate? Approximately 1 ℃/min~3 ℃/min (when directly heating the jacket).
Result reporting precision? Accurate to 0.1 ℃.
Repeatability r? Light-colored transparent samples 0.2 ℃; dark-colored samples 0.3 ℃.
Reproducibility R (aniline point)? Light-colored transparent samples 0.5 ℃; dark-colored samples 1.0 ℃.
Reproducibility R (mixed aniline point)? Light-colored transparent samples 0.7 ℃; dark-colored samples 1.0 ℃.
What to do when automatic instrument results are disputed? Only manual instruments may be used for determination.
Automatic instrument correction formula? Xa= (Xs−A)/B, A, B are obtained by the least squares method according to 8.4.2.
When to report that the method is not applicable? When continuously observed temperature shows an increasing (decreasing) trend, or the range of change is greater than 0.2 ℃ for light-colored transparent samples and greater than 0.3 ℃ for dark-colored samples.
What are the referenced documents? GB/T 514, GB/T 6540.
Relationship with ISO 2977:1997? Modified adoption (MOD), structural differences see Appendix A (informative appendix).

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.