GB/T 265-1988 Kinematic viscosity of petroleum products and calculation of dynamic viscosity

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

At a certain constant temperature, measure the time for a certain volume of liquid to flow under gravity through a calibrated glass capillary viscometer; the product of the capillary constant of the viscometer and the flow time is the kinematic viscosity of the liquid measured at that temperature. At temperature t, the kinematic viscosity is denoted by the symbol ν_t. The product of the kinematic viscosity at that temperature and the density of the liquid at the same temperature is the dynamic viscosity of the liquid at that temperature. At temperature t, the dynamic viscosity is denoted by the symbol η_t.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formula

10. Precision

Use the following provisions to judge the reliability of test results (95% confidence level).

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick-reference Q&A

Question Key points of answer
What does this standard measure? The kinematic viscosity of liquid petroleum products (Newtonian liquids), and the dynamic viscosity can also be calculated.
Unit of kinematic viscosity? The legal unit is m²/s, while mm²/s is commonly used in practice.
Principle of the method? At constant temperature, measure the time for a certain volume of liquid to flow under gravity through a calibrated glass capillary viscometer, and multiply by the viscometer constant.
How is dynamic viscosity calculated? At the same temperature, multiply kinematic viscosity by density, η_t = ν_t · ρ_t.
What are the capillary inner diameters? 0.4, 0.6, 0.8, 1.0, 1.2, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 5.0, 6.0 mm.
Flow time requirement? Generally not less than 200 s; for viscometers with inner diameter 0.4 mm, not less than 350 s.
How to choose the constant-temperature bath liquid? Select according to Table 1 based on the determination temperature: 50–100°C use transparent mineral oil/glycerol/25% ammonium nitrate aqueous solution; 20–50°C use water; 0–20°C use ice water or ethanol dry ice; 0–-50°C use ethanol dry ice (if ethanol is unavailable, lead-free gasoline may be used).
Constant-temperature time? 80, 100°C is 20 min; 40, 50°C is 15 min; 20°C is 10 min; 0~-50°C is 15 min.
Temperature control accuracy? The test temperature must be kept constant to ±0.1°C.
Number of repeated determinations? At least four times, and take the arithmetic mean of not fewer than three times.
Flow time deviation requirement? 100–15°C: ±0.5%; below 15–-30°C: ±1.5%; below -30°C: ±2.5%.
Kinematic viscosity formula? ν_t = C · τ_t.
Density determination references? GB 1884 and GB 1885.
Repeatability? 1.0% (temperature interval requires confirmation against the source standard), 3.0% (below 15~-30°C), 5.0% (below -30~-60°C).
Reproducibility? 2.2% of the arithmetic mean (temperature interval requires confirmation against the source standard).
How many significant figures for the result? Four significant figures.
Which value is taken for the report? The arithmetic mean of two repeated determination results.
Thermometer correction formula? Δt = k·h(t₁-t₂), mercury k=0.00016, alcohol k=0.0011.
What is Appendix A? A reference item, a conversion table between kinematic viscosity and Engler viscosity (degrees of condition).