GB/T 28769-2012 Determination of free glycerol content in fatty acid methyl esters — Gas chromatography

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative Determination and Calculation Formulas

10. Precision

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
What does this standard measure? Free glycerol content in fatty acid methyl esters (FAME), gas chromatography
Applicable concentration range? Mass fraction 0.005%~0.07% (1.1)
What samples is it applicable to? Fatty acid methyl esters of pure biodiesel (1.2)
Relationship with the international standard adopted? Modified adoption by redrafting method of EN 14106: 2003 (foreword)
Main differences from EN 14106? 9.1 Added splitless injection chromatographic conditions to improve sensitivity (foreword)
Method principle? Ethanol/water/n-hexane + internal standard extraction into two phases, free glycerol transfers to the lower layer, GC determination of the lower layer solution (Chapter 4)
What is the internal standard? 1,4- butanediol, purity not less than 99% (5.2, 5.8)
How to prepare the internal standard solution? In a 100 mL volumetric flask, weigh approximately 80 mg (accurate to 0.1 mg) of 1,4- butanediol, dissolve with a small amount of distilled water, add 1 mL formic acid, dilute to the mark; stable at room temperature for 24 h (5.8)
Carrier gas requirements? Helium or hydrogen, generally purity not less than 99.99% (5.6)
Detector type? Flame ionization detector (FID) (6.1.3)
Recommended chromatographic column? PoraPLOT Q capillary column 10 m×0.32 mm×10 μm, or CHROMOSORB 101 packed column 4 mm×1 m (6.1.5)
Chromatographic column performance requirements? Complete separation of glycerol and internal standard peaks, analysis time <15 min, glycerol correction factor not exceeding 2.5 (6.1.5)
Recommended injection mode? Capillary column splitless injection mode is recommended (9.1.2 Note)
Column temperature program for splitless mode? 150 ℃ hold for 1 min, raise to 210 ℃ at 40 ℃/min and hold for 12 min (9.1.1)
Column temperature for split mode? 210 ℃ (9.1.1)
Column temperature for packed column? 200 ℃ (9.1.2)
Inlet/detector temperature? Capillary splitless: 250 ℃/250 ℃; capillary split: 230 ℃/250 ℃; packed column: 230 ℃/250 ℃ (9.1)
What standard for sampling? GB/T 4756 (7.1)
What to note about sample containers? Avoid glass containers; suitable plastic containers may be used (free glycerol has strong affinity for glass) (7.2)
What standard for sample preparation? GB/T 15687; samples must not be heated or filtered (Chapter 8)
How is the correction factor measured? About 100 mg glycerol + 100 mg 1,4- butanediol in a 100 mL volumetric flask, add 50 mL ethanol to dissolve, dilute to volume with distilled water; take 1 μL for injection, at least three parallel runs (9.3)
Correction factor limit? Not exceeding 2.5 (9.3, 6.1.5)
Sample amount for quantitative analysis? About 3.5 g (accurate to 0.1 mg, about 4 mL) (9.4.1)
Extraction procedure? Add 1 mL ethanol and shake well, add 1 mL internal standard solution and 4 mL n-hexane, shake vigorously for 5 min, centrifuge for 15 min, take the lower layer (9.4.1)
Injection technique? "Hot needle" technique, take about 1 μL of the lower layer solution, insert the needle and wait for 5 s, inject rapidly, withdraw after 5 s (9.4.2)
Calculation formula? w = (A2/A1) × F × (m1/m) × 100% (Formula 2)
Expression of results? Mass fraction (%), accurate to the 2th decimal place (Chapter 10)
Repeatability formula? r = 0.466 4w − 0.0012 (Table 1)
Reproducibility formula? R = 0.781 2w + 0.0032 (Table 1)
Precision confidence level? 95% (11.1)
What must the report state? Sample identification information, sampling method, analysis method and standard number, operational details not listed and influencing factors, measurement results (Chapter 12)
Nature of Annex A? Informative annex, European 5 country 7 laboratory 5 sample collaborative test, statistics according to ISO 4259 (Annex A)

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.