Origin Standard

GB/T 30514-2014 Glass capillary kinematic viscometers — Specifications and operating instructions

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-application

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and specimen preparation

8. Analytical procedure

9. Qualitative and quantitative and calculation formulas

10. Precision

11. Report

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference for questions and answers

Question Key points of answer
What type of standard is this? Test method standard (specifications and operating instructions, including calibration method).
Standard number and name? GB/T 30514-2014 Glass capillary kinematic viscometers—Specifications and operating instructions.
Relationship to international standards? Modified adoption using the redrafting method of ISO 3105:1994 (MOD).
Publication and implementation dates? Published on 2014-02-19, implemented on 2014-06-01.
Which standard does it replace? not provided in the available source.
Scope of application? Specifies the specifications and operating instructions for glass capillary kinematic viscometers, and also specifies calibration methods; applicable to glass capillary viscometers used in the procedure of GB/T 30515 for the determination of kinematic viscosity of petroleum products.
Types of viscometers specified? Improved Ostwald (Ostwald) viscometer (Appendix A), suspended-level viscometer (Appendix B), reverse-flow viscometer (Appendix C).
Can other viscometers be used? Other glass capillary viscometers that can meet the precision requirements of GB/T 30515 for kinematic viscosity determination may also be used.
Principle of the method? Hagen-Poiseuille law; when the flow time is sufficiently long, it simplifies to v = Ct.
When can the kinetic energy correction be neglected? It can be neglected when the flow time is greater than 200 s; for low kinematic viscosity liquids, several sizes of viscometers require a minimum flow time greater than 200 s; when C ≤ 0.01 mm²/s² and the minimum flow time does not reach 200 s, the kinetic energy correction may be relatively large.
Maximum flow time? Recommended maximum flow time 1000 s, but longer flow times may also be used.
When is surface tension correction needed? When the average diameters of the upper and lower menisci of the liquid are not the same during the flow time, and there is a substantial difference between the surface tension of the sample and the surface tension of the calibration liquid.
Which viscometers are affected by temperature? Viscometers with a fixed volume of sample liquid filled at ambient temperature: Cannon-Fenske routine viscometer, Pinkevitch viscometer, Cannon-Manning semi-micro viscometer, Cannon-Fenske opaque viscometer.
Which viscometer constants are not affected by temperature? Viscometers in which the sample volume is adjusted at the bath temperature and all suspended-level viscometers.
What are the two calibration methods? The standard viscometer method (6.2) and the viscosity standard oil method (6.3).
Formula for the constant in the standard viscometer method? C₁ = (t₂×C₂)/t₁ (Equation 1).
Formula for the constant in the viscosity standard oil method? C = v/t (Equation 3).
Conditions for gravitational acceleration correction? When the difference in g between the testing laboratory and the calibration laboratory is greater than 0.1%, correct according to Equation (2); usually when g₁/g₂ is between 0.999 and 1.001, no correction is needed.
Criteria for consistency of calibration constants? For viscometers in Appendix A and B, the difference between two C values is less than 0.2%; for viscometers in Appendix C, the difference between two C values is less than 0.3%; take the average; otherwise, retest.
Significant figures for reporting viscometer constants? Accurate to 0.1% of the measured value; 1×10⁻⁶ to 6.999×10⁻⁶ take four significant figures, 7×10⁻⁶ to 9.99×

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.