GB/T 3144-1982 Determination of hydrocarbon impurities in toluene by gas chromatography

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Add a known amount of internal standard to the sample, use a syringe to take a certain amount of this mixture and inject it into the vaporization chamber of the chromatograph. The vaporized mixture is carried by the carrier gas into the chromatographic column. Each component flowing out is detected by a flame ionization detector, and the chromatogram is recorded on a recorder. The relative retention time of the impurities is used for qualitative identification, and the peak areas of the impurities relative to the internal standard are used for quantification. The relative correction factors of the detector for each component must be considered during calculation.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

Table 3 Test conditions:

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formulas

10. Precision

Use the following values to judge the reliability of the results (95% confidence level).

11. Report

Take the arithmetic mean of two parallel determination results as the determination result of hydrocarbon impurities in the sample, and report the result to 0.01% (mass ratio).

12. Reference documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
What does this standard measure? Hydrocarbon impurities in toluene: benzene, C8 aromatics, and non-aromatics up to n-nonane.
Measurement range? Each group of impurities 0.01~1.00% (mass ratio).
Basis for qualitative identification? Relative retention time (with n-undecane as the reference).
Basis for quantification? The chromatographic peak area of the impurity relative to the internal standard (n-decane), taking into account the relative correction factor.
What is the internal standard? n-Decane, purity not less than 99% (mass ratio).
Detector? Flame ionization detector (FID).
Column specifications? Length 4 meters, internal diameter 2 millimeters; stainless steel tube, copper tube, aluminum tube, or glass tube.
Stationary phase? Poly C- glycol 1500 + 6201 support (60~80 mesh).
Column temperature? 100℃ (allowed to adjust up or down by 20℃).
Vaporization temperature? 180℃ (not lower than the boiling point of the heaviest component).
How is the carrier gas flow rate determined? Adjust so that the retention time of toluene is 6~10 minutes (not exceeding 15 minutes).
Injection volume? 1 microliters (may be increased provided separation is satisfactory).
To how many places is the report taken? To 0.01% (mass ratio).
Repeatability limit? For 0.01–0.10%, it is 0.01%; for 0.10–1.00%, it is the arithmetic mean 10%.
Reproducibility limit? For 0.01–0.10%, it is 0.02%; for 0.10–1.00%, it is the arithmetic mean 20%.
Correction factor deviation requirement? If the difference from the typical value in Table 1 exceeds 10%, a second standard sample must be used for checking until it is less than 10%.