GB/T 32384-2015 Separation and determination of aromatic hydrocarbon components in middle distillates — Solid phase extraction-gas chromatography

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

4.1 Reagents

4.2 Materials

4.3 Sample for determining the fatty acid methyl ester correction factor (6.3)

4.4 Quality control check sample (Method One, 6.4)

4.5 Quality control check sample (Method Two, 6.5)

4.6 Middle distillate reference sample (6.6)

4.7 Reference sample of diesel containing fatty acid methyl esters (6.7)

5. Instruments and equipment

6. Recommended operating conditions

Table 1 Typical chromatographic operating parameters (a):

a: The instrument parameters given in the table are optional parameter conditions. A chromatographic column of 15 m to 30 m column length and inner diameter 0.15 mm to 0.32 mm may be used, and an appropriate column temperature program may be selected, as long as it can ensure complete separation of the solvent, sample, and internal standard in the analyzed sample.

7. Sampling and sample preparation

8. Analytical procedure

8.1 Preparation of the gas chromatograph and establishment of conditions

8.2 Method 1: Determination of middle distillate samples not containing fatty acid methyl esters (9.1)

8.2.1 Solid-phase extraction separation procedure

8.2.2 Gas chromatographic analysis

8.3 Method 2: Determination of diesel samples containing fatty acid methyl esters (9.2)

8.3.1 Solid-phase extraction separation procedure

8.3.2 Gas chromatographic analysis

8.4 Quality control check (Chapter 11)

9. Qualitative and quantitative determination and calculation formulas

9.1 Method 1 (10.1)

9.2 Method Two (10.2)

10. Precision

10.1 Precision (12.1)

Table 2 Precision for the determination of aromatic and non-aromatic components (% (mass fraction))

Component Content Repeatability Reproducibility
Aromatics 4~50 0.4X^0.36 0.9X^0.15
Non-aromatics 50~95 1.2 1.5

Note: X is the average of the two results being compared, % (mass fraction).

Table 3 Precision for the determination of fatty acid methyl esters (% (mass fraction or volume fraction))

Component Content Repeatability Reproducibility
Fatty acid methyl esters 2~10 0.04X+0.1 0.5

Note 1: X is the average of the two results being compared, % (mass fraction or volume fraction). Note 2: This precision was determined by collaborative tests in 4 laboratories and is for reference.

10.2 Bias (12.2)

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
How many methods does this standard have? Method One and Method Two (Chapter 1).
What samples is Method One applicable to? Middle distillates that do not contain fatty acid methyl esters (Chapter 1).
What samples is Method Two applicable to? Diesel containing fatty acid methyl esters with a volume fraction less than 10% (Chapter 1).
Applicable distillation range? Middle distillates from 170 ℃ to 400 ℃ (Chapter 1).
Applicable range of aromatic content? Mass fraction 4%~50% (Chapter 1).
Applicable range of fatty acid methyl ester content? Diesel samples with volume fraction 2%~10% (Chapter 1).
Can samples outside the content range be determined? Yes, but precision data are not provided (Chapter 1).
What does the aromatic fraction include in Method One? Thiophene-type sulfur-containing compounds, nitrogen-containing compounds, and oxygen-containing compounds (Chapter 1).
In Method Two, which component are nitrogen-containing and oxygen-containing compounds assigned to? The fatty acid methyl ester component, which may affect the results (Chapter 1).
What is the internal standard? C30 or C32 n-alkanes dissolved in n-hexane, mass concentration 0.001 g/mL~0.005 g/mL (6.1.9).
Specifications of the solid-phase extraction column? A 3 mL solid-phase extraction column packed with about 1.5 g of stationary phase (5.1.1).
Can the solid-phase extraction column be reused? It cannot be reused (5.1.1 Note 3).
How should the solid-phase extraction column be stored? Store in a desiccator (5.1.1 Note 2).
What solvent is used to elute non-aromatics in Method One? 2 mL n-pentane and 0.5 mL dichloromethane and ethanol mixed solution with a volume ratio of 5:1 (9.1.1.3).
What solvent is used to elute aromatics in Method One? 2 mL dichloromethane and ethanol mixed solution with a volume ratio of 5:1 (9.1.1.3).
What solvent is used to elute non-aromatics in Method Two? 2 mL n-pentane and 0.5 mL dichloromethane and n-pentane mixed solution with a volume ratio of 1:1 (9.2.1.3).
What solvent is used to elute aromatics in Method Two? 2 mL dichloromethane and n-pentane mixed solution with a volume ratio of 1:1 (9.2.1.3).
What solvent is used to elute fatty acid methyl esters in Method Two? 6 mL benzene and ethanol mixed solution with a volume ratio of 1:1 (9.2.1.3).
What is the approximate elution rate? About 2 mL/min (9.1.1.3, 9.2.1.3).
Volume of internal standard added? 1.00 mL (9.1.1.4, 9.2.1.4).
Typical vaporization chamber temperature? 300 ℃ (Table 1).
Typical detector temperature? 350 ℃ (Table 1).
Typical column temperature program? 60 ℃ hold 2 min, then raise at 40 ℃/min to 300 ℃, hold 5 min (Table 1).
Typical carrier gas and column flow rate? Helium or nitrogen, about 1.0 mL/min (Table 1).
Typical split ratio? About 20∶1 (Table 1).
Typical sample amount? About 0.5 μL (Table 1).
How is the fatty acid methyl ester correction factor B calculated? According to formula (3): B = (M1/M2)×(Sw/SN16) (10.2.1).
How is the fatty acid methyl ester volume fraction calculated? According to formula (7): X6 = X5×(d2/d1), d1=880.0 kg/m³ (10.2.5).
What method is used to measure the sample density? SH/T 0604 (10.2.5).
Basis for precision statistics? GB/T 6683 (12.1.1).
Confidence level for repeatability/reproducibility judgment? 95% (12.1.1).
Can bias be determined? Cannot be determined, because there is no reference material (12.2).
To how many places are reported results taken? 0.1% (Chapter 13).
According to what standard is sampling performed? GB/T 4756 (Chapter 7).
Quality control check interval? Generally 3 months (11.1, 11.2).
What method can be used to measure the separation efficiency of a solid-phase extraction column? SH/T 0606—2005 (11.2).
Requirements for aromatics in non-aromatics and non-aromatics in aromatics? The mass fraction of both should be less than 5% (Note to 11.2).