GB/T 34101-2017 Determination of hydrogen sulfide content in fuel oils — Rapid liquid-phase extraction method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Inject a test portion of known mass into a heated test tube containing diluted base oil. Bubble air through the test portion oil liquid to extract the hydrogen sulfide gas therein. The purged hydrogen sulfide, together with the air, passes through a filter box cooled to −20℃ (Method A only) and then enters the detector, where the hydrogen sulfide content in the air is measured, thereby calculating the hydrogen sulfide content in the liquid phase of the test portion. Method B omits the relevant steps concerning the low-temperature filter box. (Chapter 3)

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and test portion preparation

8. Analysis Procedure

9. Qualitative and Quantitative Determination and Calculation Formulas

Calculate the hydrogen sulfide content of the sample in the liquid phase X according to formula (1), expressed in milligrams per kilogram (mg/kg):

X = A·M / m …… (1)

Where:

Expression of results (Chapter 13):

10. Precision

11. Report

Chapter 15 Test report; the report shall contain at least the following information:

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
How many methods does this standard have? Method A and method B, two test procedures (Chapter 1, Chapter 3, 11.1)
What is the core difference between method A and method B? Method A includes a gas-phase processor and a filter box cooled to −20℃; method B does not include a gas-phase processor and omits the low-temperature filter box step (Chapter 3, 4.1, 11.2, 11.3)
Upper limit of applicable viscosity? 50℃ kinematic viscosity not greater than 3000 mm²/s; values above this can also be measured but precision is affected (Chapter 1, Note 1)
How are mercaptan/alkyl sulfide interferences handled? When the content is higher than 5 mg/kg, the result of method B is high; method A may be used to eliminate the effect (Note 2)
Do fatty acid methyl esters interfere? They have no effect on the determination result of hydrogen sulfide content (Note 3)
Diluent oil specification? Colorless API II type base oil, 40℃ kinematic viscosity 90~110 mm²/s, no hydrogen sulfide under the test conditions (5.1)
Can the diluent oil be replaced with toluene/xylene? No, it will damage the detector (5.1 Note)
Recommended concentration of calibration liquid? 2 mg/kg (5.2)
Basis for sampling? GB/T 4756 (6.1)
Sample container capacity and sample amount? Minimum 500 mL, filled to about 95% space (6.2)
Sample storage time limit? Determine within three days after sampling; if it cannot be determined immediately, it should be placed in a refrigerator (6.4)
Order of hydrogen sulfide determination in sample testing? It should be the first test item of the sample (6.5)
Upper limit of sample heating temperature? Must not exceed 40℃ (7.2)
Air flow rate? 375 mL/min±55 mL/min (9.1, A.2.2)
Heater temperature? 60.0℃±1.0℃ (9.2, 11.2.2, 11.3.2)
Gas-phase processor temperature? −20.0℃±2.0℃ (only method A, 9.4)
Air flow rate calibration interval? At least once every month (9.1)
Heater temperature calibration interval? At least once every six months (9.2)
Detector validity check interval? At least once every month or when replacing with a new detector (9.3)
Gas-phase processor calibration interval? At least once every six months (only method A, 9.4)
Overall instrument condition check interval? Recommended at least once every three months (9.5.1)
How much deviation of the detector calibration from the nominal value requires action? If the deviation is 10%, replace the moisture filter and check cleanliness, purge 30 min, then recalibrate (9.3)
Amount of diluent oil added? 20 mL±0.5 mL (11.4.1)
How is the sampling amount determined? According to Table 1: predicted 0–10 mg/kg take 5 mL (disposable syringe); >10–20 take 2 mL (disposable syringe); >20 take 1 mL (pipette) (11.4.4, Table 1)
Sample extraction position? Below the sample surface 3 cm, avoiding contact with the bottom of the container (11.4.4)
Weighing precision? Accurate to 0.001 g (11.4.4, 11.4.8)
Total test duration? 15 min (11.4.11)
Data recording frequency? Record the mV reading at least once every 4 s (11.4.11)
Calculation formula? X = A·M / m (Chapter 12, Formula 1)
Precision of result expression? <10 mg/kg accurate to 0.01 mg/kg; ≥10 mg/kg accurate to 0.1 mg/kg (Chapter 13)
Source of precision data? Method A: 2012 year British Energy Institute; method B residue: 2009 year; method B distillate: 2011 year (14.1)
What must the report include? Standard reference and method A/B, sample type and identification, test results, deviations, test time and date (Chapter 15)
Filter box specification? See Figure A.4, individually packaged, only for method A (5.6, A.2.12)
Detector type and range? Electrochemical type, 0~50 μmol/mol, repeatability 1%, T90<30 s (A.2.11)
Test tube specification? 50 mL borosilicate glass tube, with internal inlet tube (A.2.10, Figure A.3)