GB/T 3536-2008 Petroleum products — Determination of flash point and fire point — Cleveland open cup method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Fill the test cup with the sample to the specified mark. First raise the temperature of the sample rapidly, and when approaching the flash point, raise the temperature slowly at a constant rate. At specified temperature intervals, sweep a small test flame over the test cup; the lowest temperature at which the test flame causes the vapor above the liquid surface of the sample to flash is the flash point. If the fire point is to be determined, continue the test until the test flame causes the vapor at the liquid surface of the sample to ignite and sustain burning for at least 5 s; the lowest temperature at which this occurs is the fire point. The flash point and fire point measured at ambient atmospheric pressure are corrected by formula to the flash point and fire point at standard atmospheric pressure.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and Sample Preparation

8. Analytical Procedure

Flash point test procedure (Chapter 10):

Fire point test procedure (Chapter 11):

9. Qualitative and Quantitative Determination and Calculation Formulas

10. Precision

Judge the reliability of the test results according to the following provisions (95% confidence level).

11. Report

The test report shall at least include the following contents:

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer
What does this standard measure? Flash point and fire point of petroleum products, Cleveland open cup method
Which samples are applicable? Petroleum products other than fuel oils with an open cup flash point above 79 °C
What method is used for fuel oils? Usually determined according to GB/T 261
Adoption of international standard? Modified adoption of ISO 2592:2000 (MOD)
Which version does it replace? Replaces GB/T 3536—1983
Definition of flash point? Under 101.3 kPa atmospheric pressure, the lowest temperature at which the test flame causes the vapor of the test specimen to ignite and the flame to spread to the liquid surface under specified test conditions, corrected to 101.3 kPa atmospheric pressure
Definition of fire point? Under 101.3 kPa atmospheric pressure, the lowest temperature at which the test flame causes the vapor of the test specimen to ignite and to continue burning for at least 5 s under specified test conditions, corrected to 101.3 kPa atmospheric pressure
Manual or automatic apparatus for arbitration testing? In case of dispute, the arbitration test for flash point shall use a manual apparatus
Thermometer requirements? Meets the requirements of thermometer number GB-5 in GB/T 514
Thermometer installation position? The bottom of the temperature sensing bulb is 6 mm from the bottom of the test cup, located at the midpoint between the center of the test cup and the cup edge, on the diameter perpendicular to the flame sweep arc (or line), and on the opposite side of the igniter arm
To what temperature should the test cup be cooled before use? At least 56 °C below the expected flash point
Test flame diameter? 3.2 mm~4.8 mm
Initial heating rate? 14 °C/min~17 °C/min
Heating rate when approaching the flash point? During the last 23 °C±5 °C, 5 °C/min~6 °C/min
When to start sweeping? At least 23 °C±5 °C before the expected flash point, sweep once for every 2 °C rise
Sweep radius? At least 150 mm
Conditions under which an observed flash point is invalid? The result is invalid when it differs from the initial ignition temperature by less than 18 °C
What to do if there is foam on the sample surface? If foam is still present in the final stage of the test, the result is invalid
Atmospheric pressure correction formula? T = T₀ + 0.25(101.3 − p), where p is the ambient atmospheric pressure (kPa)
Applicable pressure range of the correction formula? 98.0 kPa~104.7 kPa
How are results expressed? Report the corrected flash point or fire point, in units of °C, rounded to an integer
Repeatability r? Both flash point and fire point do not exceed 8 °C
Reproducibility R? Flash point does not exceed 17 °C, fire point does not exceed 14 °C
What is used for instrument verification? Certified reference materials (CRM), at least once a year; frequent verification with working reference materials (SWS) is recommended
Acceptance criteria for verification? Single: |x − M| ≤ R/√2; Multiple: |x̄ − M| ≤ R′/√2
Reference values of calibration fluids? n-Tetradecane 116 °C; n-Hexadecane 139 °C (Table B.1)
Storage temperature of SWS? Avoid exceeding 10 °C, store in a light-resistant container
Sample container fill level? Fill to 85%~95% of the container volume
Sample storage temperature? Avoid exceeding 30 °C
Sample division temperature? At least 56 °C below the expected flash point
Report at least includes? Standard number, product type and information, test results, differences from the specified procedure, test date

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.