GB/T 514-2005 Technical conditions for glass liquid thermometers for petroleum product testing
AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.
Card type: Product standard
1. Standard number / full title / replacement relationship / publication and implementation
- Standard number: GB/T 514—2005
- Chinese full title: Technical conditions for glass liquid thermometers for petroleum product testing
- English name: Specification for liquid-in-glass thermometers for testing of petroleum products
- Replacement relationship: replaces GB/T 514—1983 "Technical conditions for liquid thermometers for petroleum product testing"
- Previous versions: GB 514—1965, GB 514—1975, GB/T 514—1983
- Publication: issued by 2005-05-23
- Implementation: implemented by 2005-11-01
- Issuing bodies: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of China
- Proposed by: China Petrochemical Corporation
- Administered by: Sinopec Research Institute of Petroleum Processing
- Drafting organization: Sinopec Research Institute of Petroleum Processing
- Participating drafting organization: Shanghai Huachen Medical Instrument Co., Ltd.
- Main drafters: Yang Tingting, Guo Tao, Li Qian
- Main changes from the 1983 version: modified the technical specifications of 10 thermometers; added technical specifications for 39 thermometers and corresponding illustrations and provisions for thermometer calibration points; technical specifications are mainly in list form, combining tables and figures; added clauses and appendices A, B, and C on referenced documents, terms and definitions, glass manufacturing requirements, capillary gap dimensions, and correction
- Nature of appendices: Appendix A, Appendix B, and Appendix C are all informative appendices
2. Scope of application and grade classification
- Scope (Chapter 1): This standard specifies the specification requirements for glass liquid thermometers used for temperature measurement in petroleum product test methods.
- Type (Chapter 4): Divided by structure into straight-stem thermometers and enclosed-scale thermometers (see Figure 1).
- Temperature-sensing liquid: mercury (quicksilver), ethanol, or mercury-based alloy.
- Thermometer numbering: the thermometers arranged by thermometer number are listed in Table 1; the thermometers classified by temperature range and scale division are listed in Table 2.
- Temperature scale (Chapter 5): based on the 1990 International Temperature Scale (ITS-90), graduated in degrees Celsius.
- Thermometer number range: GB-1 to GB-77 (Table 1 lists GB-1 to GB-77).
3. Technical requirements (index table, item by item: item | index | test method standard number)
3.1 General requirements (Chapter 7)
| Item | Index | Test method standard number |
|---|---|---|
| Glass | Shall comply with the requirements of relevant professional standards; stress shall minimize the possibility of damage caused by heating or mechanical vibration; the temperature-sensing bulb glass shall undergo appropriate thermal stability treatment; the clarity of the engraved lines shall not be affected by glass dullness or turbidity; the meniscus shall not be distorted by glass defects or impurities | not provided in the available source |
| Temperature-sensing liquid (mercury/mercury-based alloy) | Must be pure and dry, free of bubbles and other impurities; mercury shall comply with the No. 1 mercury requirements in GB 913 | GB 913 |
| Temperature-sensing liquid (ethanol/organic liquid) | Must be pure, free of organic impurities, fibers, and other inclusions visible to the eye, and no bubbles shall precipitate out; a dye that does not adhere to glass, does not easily fade, and does not precipitate shall be used for coloring; the coloring shall be stable and obvious, and shall not leave traces on the inner wall of the capillary | not provided in the available source |
| Evaporation and condensation of temperature-sensing liquid | Shall not evaporate or condense under normal pressure within the temperature measurement range | not provided in the available source |
| Filling gas | The space above the liquid column in the capillary must be filled with dry, pure inert gas at sufficient pressure | not provided in the available source |
| Capillary | Shall be straight, with no unevenness or bending detectable by the eye; the capillary bore cross-section shall be circular or elliptical (the major axis of the ellipse at right angles to the viewing surface), with a uniform inner diameter, and no visible contraction or expansion within the reading range | not provided in the available source |
| Inner and outer surfaces of the capillary | Shall not be rough or dirty, and there shall be no bubbles affecting readings within the scale range | not provided in the available source |
| Buffer space | There must be a buffer space above the upper-limit engraved line (safety bulb or extended capillary bore), and its volume shall be greater than the volume expanded at the upper-limit temperature 20°C; low-temperature thermometers shall be able to withstand the liquid expansion amount at an ambient temperature of 60°C | not provided in the available source |
| Liquid column movement | Must be uniform, without jumping, shall not leave traces on the tube wall, and shall not split into segments | not provided in the available source |
| Straight-stem enamel band | A milky white (or colored) enamel band with a width not less than one quarter of the outer circumference of the stem shall be fused along the capillary; no enamel band is allowed above the upper-limit temperature 300°C | not provided in the available source |
| Enclosed-scale inner-core capillary | Shall pass through the center of the scale and be perpendicular to the scale plate markings; lateral displacement shall not exceed the range of the shortest engraved line length; the distance from the scale plate shall not be greater than 1 mm; longitudinal relative displacement shall not be greater than one half of the smallest scale division | not provided in the available source |
| Enclosed-scale tying | The inner-core capillary shall be tied to the scale plate with copper wire of diameter 0.2 mm to 0.3 mm, with no more than two tying points, and the copper wire shall be corrosion-resistant | not provided in the available source |
| Outer tube | The enclosed-scale outer tube shall be cylindrical, gradually contracting at the bottom end of the scale plate, and the outer tube thickness shall not be less than 0.6 mm; the inner and outer surfaces shall have no scratches or other defects affecting strength, and the front shall not be rough, nor have bubbles, nodules, or other blemishes | not provided in the available source |
| Capillary gap dimensions | See 7.6.1.1 to 7.6.1.4 and 7.6.2 (when the lowest scale line is ≤100°C, the minimum length between the temperature-sensing bulb and the lowest scale line is 13 mm; when >100°C, 30 mm; the minimum between the intermediate bulb and the scale line below it is 5 mm; between the intermediate bulb and the scale line above it, when ≤100°C, 10 mm, and when >100°C, 30 mm; above the highest scale line, when there is a safety bulb, the minimum is 10 mm, and when there is no safety bulb, 30 mm; the volume of the safety bulb shall not be less than the volume equivalent to 20 mm of normal capillary; the intermediate bulb shall be long and narrow, and its position shall be at least 10 mm below the immersion depth line) | not provided in the available source |
| Scale and engraved lines | See 7.7.1 to 7.7.9; the maximum engraved line width shall comply with the provisions of Table 4; the ratio of engraved line length to scale width shall comply with Table 5 | not provided in the available source |
| Immersion depth line | For partial-immersion thermometers, a permanent immersion depth line shall be marked on the front at a distance from the temperature-sensing bulb corresponding to the immersion depth ±0.5 mm specified in Table 1 | not provided in the available source |
| Top finish | The top of a straight-stem thermometer may be smooth-finished, ring-shaped, or button-shaped, and the outer diameter of the ring or button shall not be greater than the outer diameter of the stem; the top of an enclosed-scale thermometer shall be smooth and spherical | not provided in the available source |
| Metal sleeve | Engler viscosity No. 1 and No. 2, closed-cup flash point No. 1 and No. 2, and dropping point No. 1 and No. 2 thermometers shall all have a metal sleeve that can be fixed on the apparatus; the accuracy check of thermometers for Engler viscosity and closed-cup flash point determination shall be carried out with the metal sleeve attached | not provided in the available source |
3.2 Technical specifications for each thermometer
- The technical specifications of enclosed-scale thermometers are detailed in Table 3 (Figure 2 to Figure 10).
- The technical specifications of straight-stem thermometers are detailed in Table 4 (Figure 11 to Figure 18).
- Table 3 and Table 4 list, thermometer by thermometer: thermometer number, name, corresponding figure number, temperature range, immersion depth, scale division, long engraved line interval, numbered marking interval, indication tolerance, shortest distance between graduations/maximum engraved line width, total length, outer diameter, temperature-sensing bulb dimensions, engraved line position, safety bulb, intermediate bulb, stem expansion portion, top finish shape, notes, etc. (for specific values see the original Table 3 and Table 4; this card does not transcribe them item by item).
- Table 5: ratio of engraved line length to scale width (listed separately by smallest division 0.1°C, 0.2°C, 0.5°C, 1.0°C, 2.0°C for odd-numbered degrees, even-numbered degrees, 5C, 10C, 20C).
3.3 Thermometer numbers and names (Table 1)
| Number | Name | Temperature range/°C | Immersion depth/mm | Scale division/°C | Applicable method standard number |
|---|---|---|---|---|---|
| GB-1 | Closed-cup flash point No. 1 (enclosed-scale) | -30 to 170 | 55 | 1 | GB/T 261 |
| GB-2 | Closed-cup flash point No. 2 (enclosed-scale) | 100 to 300 | 55 | 1 | GB/T 261 |
| GB-3 | Open-cup flash point No. 1 (enclosed-scale) | 0 to 360 | 45 | 1 | GB/T 267, GB/T 509 |
| GB-4 | Open-cup flash point No. 1 | 0 to 360 | 45 | 1 | GB/T 267, SH/T 0059, SH/T 0561 |
| GB-5 | Open-cup flash point No. 2 | -6 to 400 | 25 | 2 | GB/T 3536 |
| GB-6 | Engler viscosity No. 1 (enclosed-scale) | 0 to 60 | 90 | 0.5 | GB/T 266 |
| GB-7 | Engler viscosity No. 2 (enclosed-scale) | 50 to 110 | 90 | 0.5 | GB/T 266 |
| GB-8 | Asphalt Engler viscosity | 19 to 27 | 90 | 0.1 | SH/T 0099.1 |
| GB-9 | Kinematic viscosity No. 1 (enclosed-scale) | 98 to 102 | Full immersion | 0.1 | GB/T 265 |
| GB-10 | Kinematic viscosity 2 number (internal standard type) | 78~82 | Full immersion | 0.1 | GB/T 265 |
| GB-11 | Kinematic viscosity 3 number (internal standard type) | 48~52 | Full immersion | 0.1 | GB/T 265 |
| GB-12 | Kinematic viscosity 4 number (internal standard type) | 38~42 | Full immersion | 0.1 | GB/T 265 |
| GB-13 | Kinematic viscosity 5 number (internal standard type) | 18~22 | Full immersion | 0.1 | GB/T 265、SH/T 0685 |
| GB-14 | Kinematic viscosity 6 number (internal standard type) | -2~2 | Full immersion | 0.1 | GB/T 265 |
| GB-15 | Kinematic viscosity 7 number (internal standard type) | -22~-18 | Full immersion | 0.1 | GB/T 265 |
| GB-16 | Kinematic viscosity 8 number (internal standard type) | -32~-28 | Full immersion | 0.1 | GB/T 265 |
| GB-17 | Kinematic viscosity 9 number (internal standard type) | -42~-38 | Full immersion | 0.1 | GB/T 265 |
| GB-18 | Kinematic viscosity 10 number (internal standard type) | -52~-48 | Full immersion | 0.1 | GB/T 265 |
| GB-19 | Kinematic viscosity 11 number (internal standard type) | -62~-58 | Full immersion | 0.1 | GB/T 265 |
| GB-20 | Asphalt viscosity 1 number | 58.6~61.4 | Full immersion | 0.05 | SH/T 0557 |
| GB-21 | Asphalt viscosity 2 number | 133.6~136.4 | Full immersion | 0.05 | SH/T 0557 |
| GB-22 | Brookfield viscosity 1 number | 45~-35 | Full immersion | 0.1 | GB/T 11145 |
| GB-23 | Brookfield viscosity 2 number | -35~-25 | Full immersion | 0.1 | GB/T 11145 |
| GB-24 | Brookfield viscosity 3 number | -25~-15 | Full immersion | 0.1 | GB/T 11145 |
| GB-25 | Brookfield viscosity 4 number | -15~-5 | Full immersion | 0.1 | GB/T 11145 |
| GB-26 | Dropping point 1 number (internal standard type) | 0~150 | Full immersion | 1 | SH/T 0115 |
| GB-27 | Dropping point 2 number (internal standard type) | 100~250 | Full immersion | 1 | SH/T 0115 |
| GB-28 | Dropping point 3 number | -5~300 | 76 | 1 | GB/T 4929、SH/T 0089、SH/T 0430 |
| GB-29 | Dropping point 4 number | -5~400 | 76 | 1 | GB/T 3498、GB/T 8019、SH/T 0661 |
| GB-30 | Freezing point (internal standard type) | -30~60 | 150 | 1 | GB/T 510 |
| GB-31 | Crystallization point, freezing point 1 number (internal standard type) | -80~60 | 75 | 1 | SH/T 0179、SH/T 0644 |
| GB-32 | Crystallization point, freezing point 2 number (internal standard type) | -60~60 | 75 | 1 | SH/T 0179 |
| GB-33 | Paraffin wax congealing point | 20~100 | Full immersion | 0.5 | SH/T 0132 |
| GB-34 | Melting point | 38~82 | 79 | 0.1 | GB/T 2539 |
| GB-35 | Paraffin wax dropping point, melting point | 32~127 | 79 | 0.2 | GB/T 8026、SH/T 0100 |
| GB-36 | Cloud point, pour point 1 number | -80~20 | 76 | 0.5 | GB/T 3535、GB/T 6986、SH/T 0030、SH/T 0248 |
| GB-37 | Cloud point, pour point 2 number | -38~50 | 108 | 0.2 | GB/T 3535、GB/T 6986、SH/T 0030、SH/T 0603、SH/T 0248 |
| GB-38 | Ice point | -80~20 | Full immersion | 0.2 | GB/T 2430 |
| GB-39 | Coolant freezing point 1 number | -37~2 | 100 | 0.2 | SH/T 0090 |
| GB-40 | Coolant freezing point 2 number | -54~-15 | 100 | 0.2 | SH/T 0090 |
| GB-41 | Benzene crystallization point | 4~6 | Full immersion | 0.02 | GB/T 3145 |
| GB-42 | Softening point | 30~180 | Full immersion | 0.5 | GB/T 4507、SH/T 0639 |
| GB-43 | Brittle fracture point | -38~30 | Full immersion | 0.2 | GB/T 4510 |
| GB-44 | Distillation 1 number (internal standard type) | 0~360 | Full immersion | 1 | GB/T 255 |
| GB-45 | Distillation 2 number | 0~360 | Full immersion | 1 | GB/T 255 |
| GB-46 | Distillation 3 number | -2~300 | Full immersion | 1 | GB/T 6536、SH/T 0326 |
| GB-47 | Distillation 4 number | -2~400 | Full immersion | 1 | GB/T 6536 |
| GB-48 | Lubricating oil foaming | -20~102 | Full immersion | 0.2 | GB/T 12579、GB/T 1884、SH/T 0068 |
| GB-49 | Coolant foaming | 20~150 | 76 | 0.1 | SH/T 0066 |
| GB-50 | Liquefied gas volatility | -50~5 | 35 | 0.1 | GB/T 13287 |
| GB-51 | Evaporation loss 1 number | 95~155 | Full immersion | 0.5 | GB/T 7325 |
| GB-52 | Evaporation loss 2 number | 155~170 | Full immersion | 0.1 | GB/T 5304、GB/T 11964 |
| GB-53 | Rust-preventive oil evaporation | 100~115 | Full immersion | 0.2 | SH/T 0035 |
| GB-54 | Vapor pressure 1 number | 34~42 | Full immersion | 0.2 | GB/T 6602、GB/T 8017 |
| GB-55 | Vapor pressure 2 number | 40~70 | Full immersion | 0.1 | GB/T 6602 |
| GB-56 | Demulsification | -1~105 | Full immersion | 0.1 | SH/T 0191 |
| GB-57 | Oxidation characteristics 1 number | 80~100 | 76 | 0.1 | GB/T 12581、SH/T 0565 |
| GB-58 | Oxidation characteristics 2 number | 72~126 | 100 | 0.2 | GB/T 12581, SH/T 0124, SH/T 0565 |
| GB-59 | Oxidation stability 1 number | 98~152 | 100 | 0.2 | GB/T 12580, SH/T 0722 |
| GB-60 | Oxidation stability 2 number | 95~103 | Full immersion | 0.1 | GB/T 7325, GB/T 8018, SH/T 0060, SH/T 0325, SH/T 0585, SH/T 0702 |
| GB-61 | Thermal stability 1 number | 165~180 | Full immersion | 0.5 | SH/T 0560 |
| GB-62 | Thermal stability 2 number | 145~160 | Full immersion | 0.5 | SH/T 0560 |
| GB-63 | Thermal stability 3 number | 130~145 | Full immersion | 0.5 | SH/T 0560 |
| GB-64 | Ageing characteristics | 195~205 | 100 | 0.1 | GB/T 12709, SH/T 0192 |
| GB-65 | Density 1 number | -5~25 | Full immersion | 0.1 | GB/T 13377 |
| GB-66 | Density 2 number | 20~45 | Full immersion | 0.1 | GB/T 13377 |
| GB-67 | Density 3 number | 40~65 | Full immersion | 0.1 | GB/T 13377 |
| GB-68 | Density 4 number | -1~38 | Full immersion | 0.1 | GB/T 1884 |
| GB-69 | Liquefied gas density | -15~45 | Full immersion | 0.2 | SH/T 0221 |
| GB-70 | Paraffin wax oil content | -37~21 | 76 | 0.5 | GB/T 3554, SH/T 0638 |
| GB-71 | Paraffin wax needle penetration | 25~55 | Full immersion | 0.1 | GB/T 4985 |
| GB-72 | Oil tank thermometer 1 number | -34~52 | Full immersion | 0.5 | GB/T 8927 |
| GB-73 | Oil tank thermometer 2 number | -16~82 | Full immersion | 0.5 | GB/T 8927 |
| GB-74 | Oil tank thermometer 3 number | 50~240 | Full immersion | 1 | GB/T 8927 |
| GB-75 | Aniline point 1 number | 38~42 | 50 | 0.2 | GB/T 262 |
| GB-76 | Aniline point 2 number | 25~105 | 50 | 0.2 | GB/T 262 |
| GB-77 | Aniline point 3 number | 90~170 | 50 | 0.2 | GB/T 262 |
Note: Table 1 contains garbled temperature ranges, immersion depths, scale divisions and method references in the source transcription. Verify each affected specification against the original table before selecting or using a thermometer; the transcribed table is not a verified instrument specification.
3.4 Temperature range table (Table 2)
Table 2 lists temperature ranges, immersion depths, and thermometer numbers grouped by scale division values 0.02°C, 0.05°C, 0.1°C, 0.2°C, 0.5°C, 1°C, 2°C (for details see Table 2 in the original text).
4. List of referenced test methods
- GB 913 mercury
- JJG 50 Verification regulation for liquid-in-glass thermometers for petroleum products
- Applied method standard numbers listed in Table 1, Table 2, and Appendix A (GB/T 255, GB/T 261, GB/T 262, GB/T 265, GB/T 266, GB/T 267, GB/T 509, GB/T 510, GB/T 1884, GB/T 2430, GB/T 2539, GB/T 3145, GB/T 3498, GB/T 3535, GB/T 3536, GB/T 3554, GB/T 4507, GB/T 4510, GB/T 4929, GB/T 4985, GB/T 5304, GB/T 6536, GB/T 6602, GB/T 6986, GB/T 7325, GB/T 8017, GB/T 8018, GB/T 8019, GB/T 8026, GB/T 8927, GB/T 11145, GB/T 11964, GB/T 12579, GB/T 12580, GB/T 12581, GB/T 12709, GB/T 13287, GB/T 13377, SH/T 0030, SH/T 0035, SH/T 0059, SH/T 0060, SH/T 0066, SH/T 0068, SH/T 0089, SH/T 0090, SH/T 0099.1, SH/T 0100, SH/T 0115, SH/T 0124, SH/T 0132, SH/T 0179, SH/T 0191, SH/T 0192, SH/T 0221, SH/T 0248, SH/T 0325, SH/T 0326, SH/T 0430, SH/T 0557, SH/T 0560, SH/T 0561, SH/T 0565, SH/T 0585, SH/T 0603, SH/T 0638, SH/T 0639, SH/T 0644, SH/T 0661, SH/T 0685, SH/T 0702, SH/T 0722, etc.)
5. Inspection rules, sampling
- Correction (8.1): Full-immersion thermometers or partial-immersion thermometers shall be used under the specified conditions of use; if they are not used under the specified immersion depth conditions, or if the temperature of the emergent stem when a partial-immersion thermometer is used does not match the specified average temperature of the emergent stem, refer to the correction method in Appendix C.
- Verification (8.2):
- 8.2.1 Thermometers shall be verified in accordance with the national metrological verification regulation JJG 50 and other relevant regulations and standard methods of the national metrological verification authority, and at each point listed in Table 6.
- 8.2.2 The verification period shall be carried out in accordance with the period specified in the national metrological verification regulation JJG 50 (the maximum shall not exceed 1 years).
- 8.2.3 Thermometers shall be used within the validity period of verification.
- 8.2.4 All thermometers shall strictly comply with the requirements of this technical specification and be accompanied by a verification certificate.
- Verification points for thermometers in Table 6: list the verification points for GB-1~GB-77 thermometer by thermometer (for details see Table 6 in the original text).
- Sampling: not provided in the available source.
6. Marking, packaging, transportation, storage
- Marking (9.1):
- On the scale surface of straight-form thermometers and on the front of the scale plate of enclosed-scale thermometers, the mark C (or °C) is engraved; on the back of the enamel strip of straight-form thermometers and on the back of the scale plate of enclosed-scale thermometers, the following marks are engraved:
- 9.1.1 Thermometer number (for enclosed-scale thermometers, the number is engraved on the front of the scale plate);
- 9.1.2 Manufacturer's trademark;
- 9.1.3 CMC metrological mark;
- 9.1.4 Year and month of manufacture;
- 9.1.5 Name of the thermometer.
- Packaging and storage (9.2): Each thermometer shall be placed in a suitable box, and the box shall be marked with the thermometer number and temperature range. After packaging, the thermometers shall be stored indoors at -30°C to 40°C, with a relative humidity not exceeding 80%, free from corrosive gases, and well ventilated.
- Transportation: not provided in the available source.
7. Quick Reference Q&A
| Question | Key Answer Points |
|---|---|
| Which version does this standard replace? | Replaces GB/T 514—1983; previous versions are GB 514—1965, GB 514—1975, GB/T 514—1983 |
| What does this standard apply to? | Specification requirements for liquid-in-glass thermometers used for temperature measurement in petroleum product test methods (Chapter 1) |
| How many types of thermometers are there by structure? | Solid-stem thermometers and enclosed-scale thermometers (Chapter 4) |
| What are the thermometric liquids? | Mercury (quicksilver), ethanol, or mercury-based alloy (Chapter 4) |
| What is the temperature scale based on? | The 1990 International Temperature Scale (ITS-90), graduated in degrees Celsius (Chapter 5) |
| Thermometer number range? | GB-1 to GB-77 (Table 1) |
| Where are the technical specifications for enclosed-scale thermometers? | Table 3 (Figure 2 to Figure 10) |
| Where are the technical specifications for solid-stem thermometers? | Table 4 (Figure 11 to Figure 18) |
| What shall mercury thermometric liquid comply with? | It shall comply with the requirements for No. 1 mercury in GB 913 (7.2.1) |
| Thickness requirement for the outer tube of enclosed-scale thermometers? | Not less than 0.6 mm (7.5.1) |
| Distance between the inner capillary tube and the scale plate of enclosed-scale thermometers? | Shall not be greater than 1 mm (7.4.6) |
| Diameter of the copper wire tied for enclosed-scale thermometers? | 0.2 mm to 0.3 mm, with no more than two tying points (7.4.7) |
| Safety bulb volume requirement? | Not less than the volume equivalent to 20 mm of normal capillary (7.6.1.4) |
| Position requirement for the intermediate bulb? | Shall be at least 10 mm below the immersion depth line (7.6.2) |
| Tolerance for the immersion depth line marking of partial-immersion thermometers? | Immersion depth ±0.5 mm as specified in Table 1 (7.8) |
| Which thermometers require a metal sheath? | Engler viscosity 1, 2; closed-cup flash point 1, 2; dropping point 1, 2 (7.10) |
| What regulation is verification based on? | JJG 50 Verification Regulation for Liquid-in-Glass Thermometers for Petroleum Products (8.2.1) |
| What is the maximum verification period? | Shall not exceed 1 years (8.2.2) |
| Prerequisite for using thermometers? | Shall be used within the valid verification period and accompanied by a verification certificate (8.2.3, 8.2.4) |
| What does the marking include? | Thermometer number, manufacturer's trademark, CMC metrological mark, date of manufacture, thermometer name (9.1) |
| Storage conditions? | Indoors at -30°C to 40°C, with a relative humidity not exceeding 80%, free from corrosive gases, and well ventilated (9.2) |
| What is Annex A? | Informative annex, a list of thermometers used in standard test methods for petroleum products (Table A.1) |
| What is Annex B? | Informative annex, correspondence between thermometers specified in this standard and corresponding foreign thermometers (Table B.1) |
| What is Annex C? | Informative annex, reference correction method for the emergent stem temperature of thermometers (Equations C.1, C.2, C.3, Tables C.1, C.2) |
| Value of k in the emergent stem correction formula? | For mercury and mercury-thallium alloy k, it is 0.000 16; for ethanol 0.001 04; for toluene 0.001 03; for pentane 0.001 45 (Annex C.1) |
Related public products
Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.
Search related productsSource and review status
This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.