GB/T 6539-1997 Determination of Electrical Conductivity of Aviation and Distillate Fuels

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

The following reagents are used as cleaning solvents. If water is suspected to be present, isopropanol may be used first, followed by toluene.

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

Conductivity Repeatability Reproducibility
1 1 1
15 1 3
20 1 4
30 2 6
50 3 10
70 4 13
100 5 17
200 10 32
300 14 45
500 21 69
700 29 92
1000 39 125
1500 55 177

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A Reference

Question Key Answer Points
What does this standard measure? Conductivity (static conductivity) of aviation fuels and distillate fuels, with or without antistatic additives (1.1)
What international standard is adopted? Equivalent adoption of ASTM D 2624—95 (Foreword)
Which version does it replace? Replaces GB/T 6539-86 (91) (Cover)
What is the lower measurement limit? 1pS/m or above (4.1); below 1pS/m, GB/T 12582 may be used (4.1.1)
What instruments may be used? MAIHAK MLA type and EMCEE 1152 type conductivity meters (4.1.1, 10.2)
What are the requirements for the measurement container? A cylindrical container capable of fully immersing the conductivity cell, with a volume of not less than 1L (6.3)
What is the sample amount requirement? As much as possible, at least not less than 1L (9.1.2)
How soon after sampling should measurement be performed? Measure as soon as possible, and at the latest it should not exceed 24h (9.1.4)
On-site measurement standing time? Automotive tank cars, railway tank cars 2min; <5000m³ storage tanks or oil tanks 10min; >5000m³ storage tanks or oil tanks 30min (10.2.1)
Reading recording requirements? Turn on the instrument and record the highest reading after initial stabilization, which should be completed within 3s; for multiple ranges, select the range with the highest sensitivity (10.2.2)
What should be done if the conductivity cell contacts water? It must be thoroughly rinsed with cleaning solvent, preferably first rinsed with isopropanol, then dried with a stream of air (9.1.1)
What cleaning solvents are available? Isopropanol, toluene, n-heptane, all analytical grade; a mixed solvent of 50% isopropanol + 50% n-heptane may be used instead of toluene (7)
How to report a zero reading? The measurement result may be reported as less than 1pS/m (11.1)
Applicable scope of precision data? The data in Table 1 do not include gasoline or solvents; they are limited to application at room temperature; at about -20℃ the values are (×2) (12.1, 12.2)
Is there a bias value? No, because there is neither a recognized standard substance nor an appropriate test method to determine bias (12.3)
Where is the calibration procedure? Appendix A (normative appendix), divided into MAIHAK MLA (analog type) and EMCEE 1152 (digital type) (Appendix A)
What are the differences from ASTM D 2624—95? The online continuous monitoring method in Chapter 13 is not adopted; the standing 30min provisions of API RP 2003 are not adopted; Appendix X is not adopted (Foreword)

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.