GB/T 8025-1987 Determination of trace sulfur in petroleum waxes and petroleum fats (Microcoulometric method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The analytical sample is vaporized in the vaporization section of the pyrolysis tube and mixed with carrier gas (nitrogen, original text “[gas name garbled in several forms]” requires confirmation against the source standard (source-text extraction)) to enter the combustion section, where it is mixed with oxygen and undergoes pyrolysis and oxidation. Sulfur is converted to sulfur dioxide, which enters the titration cell together with the carrier gas and reacts with triiodide ions in the electrolyte according to the following equation: I₃⁻ + SO₂ + H₂O → SO₄²⁻ + 3I⁻ + 2H⁺ (original equation source-text extraction is “Is+SOz+H0-S0+31-+2H”, requires confirmation against the source standard (source-text extraction)) The concentration of triiodide ions in the titration cell decreases, the measuring electrode senses the bias voltage and the potential change between the “reference-measuring” electrode pair, this signal is input to the microcoulometer amplifier, and after amplification the output voltage is applied to the electrolysis electrode, where the following reaction occurs at the electrolysis anode: 3I⁻ → I₃⁻ + 2e (original source-text extraction is “31-—-1+2e”, requires confirmation against the source standard (source-text extraction)) The consumed triiodide ions are replenished, and the quantity of electricity consumed is the integral of the electrolysis current over time. According to Faraday's law of electrolysis, the total sulfur amount equivalent to sulfur dioxide is obtained, and the sulfur content of the sample is calculated according to the recovery rate.

4. Reagents and materials (purity / preparation)

All reagents used in the test are of chemical purity.

5. Instruments and equipment

6. Recommended operating conditions

4.5 Adjust gas flow and furnace temperature, operating conditions are as follows:

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formulas

Calculate the test results according to one of the two formulas (3) and (4). Formula (3) is suitable for instruments equipped with coulometric integration; formula (4) is suitable for instruments not equipped with coulometric integration or for determining samples with lower sulfur content.

Appendix A Derivation of the formula (reference item):

10. Precision

11. Report

For each sample, conduct two parallel determination tests, and take the average value as the determination value of the sample. (The source “for each sample, carry out [garbled repeat count] 4 [garbled parallel-test wording], and take the mean as the sample result”, requires confirmation against the source standard (source-text extraction))

12. Normative references

not provided in the available source.

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference for questions and answers

Question Key points of answer
What does this standard apply to? Petroleum waxes and petroleum greases, including trace sulfur content in white oil, liquid paraffin, petrolatum, paraffin wax, and ozokerite
What is the determination range? 0.5~500 ppm
Interference limits? Halogen content shall not be greater than 10 times the sulfur content, silver content shall not be greater than 10%, heavy metals shall not exceed 500 ppm
What is the method principle? The sample is vaporized, cracked, and oxidized; sulfur is converted to SO₂, enters the titration cell and reacts with triiodide ions; the microcoulometer measures the electrolysis electric quantity, and the sulfur content is calculated according to Faraday's law
Pyrolysis furnace temperature conditions? Inlet section 550–700 ℃, central section 900 ℃, outlet section 800–900 ℃
Gas flow conditions? Reaction gas oxygen 90–100 mL/min, carrier gas nitrogen 100–110 mL/min
Bias voltage range? 120~160 millivolts
Recovery rate requirement? Must be greater than 75% before the test can be carried out
How is the sampling amount determined? Depending on the sulfur content of the sample, the sampling amount range is different; weighing is accurate to 0.01 milligrams
How is the injection speed controlled? Three stages: push to the furnace mouth at 10–12 millimeters/second, after the peak appears stay until about the peak top, then push into the furnace at 1 millimeters/second for 10–40 centimeters
What calculation formulas are there? Formula (3) for instruments with electric quantity integration; formula (4) for instruments without electric quantity integration or low-sulfur samples
How is precision? Repeatability and reproducibility according to the maximum allowable difference of the precision curve in Figure 3
How many determinations are taken for the report? Two parallel determinations, take the average value
How is the electrolyte prepared? 0.5 grams potassium iodide + 0.6 grams sodium azide dissolved in 500 milliliters distilled water, add 5 milliliters glacial acetic acid, dilute to 1000 milliliters, store in a brown bottle
Formula for sulfur content of standard sample? S ppm = (g·S × 10⁶)/(g+G)
Drafting unit? Fushun Research Institute of Petroleum and Petrochemicals

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.