Origin Standard

GB/T 8926-2012 Determination of Insolubles in Used Lubricating Oils

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

Table 1 Calibration errors for conical centrifuge tubes

Range/mL Minimum graduation line/mL Maximum permissible graduation error/mL
0~0.1 0.05 ±0.02
>0.1~0.3 0.05 ±0.03
≥0.3~0.5 0.05 ±0.05
>0.5~1.0 0.1 ±0.05
>1.0~2.0 0.1 ±0.10
>2.0~3.0 0.2 ±0.10
≥3.0~5.0 0.5 ±0.20
>5.0~10.0 1.0 ±0.50
>10~25 5.0 ±1.0
>25~100 25.0 ±1.0

Table 2 Rotation speeds applicable to centrifuges with different rotation diameters

Rotation diameter*/mm Rotation speed at relative centrifugal force 600/(r/min) Rotation speed at relative centrifugal force 700/(r/min)
483 1490 1610
508 1450 1570
533 1420 1530
559 1390 1500

Figure 1 Conical centrifuge tube (the source text contains dimension annotations: inner diameter φ17±1, outer diameter φ36~37.75, frosted area (for marking), maximum radius 4, and graduation annotations such as 100/75/50/25/20/15; specific dimensions are subject to the figure in the source text, requires confirmation against the source standard (source-text extraction))

6. Recommended operating conditions

7. Sampling and test portion preparation

8. Analysis Steps

8 Method A: n-Pentane and Toluene Insolubles Without Coagulant

8.1 n-Pentane Insolubles

8.2 Toluene Insolubles

9 Method B: Coagulated n-Pentane Insolubles and Coagulated Toluene Insolubles with Coagulant

9.1 Coagulated n-Pentane Insolubles

9.2 Coagulated Toluene Insolubles

9. Qualitative and Quantitative Determination and Calculation Formulas

10. Precision

Table 3

Insolubles (mass fraction)/% Repeatability (mass fraction)/%
0.0~1.0 0.07
>1.0 10% of the mean

Table 4

Insolubles (mass fraction)/% Repeatability (mass fraction)/%
0.1 0.068
0.2 0.097
0.3 0.12
0.4 0.14
0.5 0.15
0.6 0.17
0.7 0.18
0.8 0.19

Table 5

Insolubles (mass fraction)/% Reproducibility (mass fraction)/%
0.0~1.0 0.10
>1.0 15% of the mean

Table 6

Insolubles (mass fraction)/% Reproducibility (mass fraction)/%
0.1 0.14
0.2 0.20
0.3 0.24
0.4 0.28
0.5 0.31
0.6 0.34
0.7 0.37
0.8 0.39
0.9 0.42
1.0 0.44
1.1 0.46

Table 7

Insolubles (mass fraction)/% Repeatability (mass fraction)/%
0.1 0.30
0.2 0.43
0.3 0.53
0.4 0.60
0.5 0.68
0.6 0.74
0.7 0.80
0.8 0.86
0.9 0.91
1.0 0.96
1.1 1.00

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A

Question Key answer
How many methods does this standard have? Two: Method A (without coagulant), Method B (with coagulant) (see 1.1, 1.2, 1.3)
Can the results of Method A and Method B be compared with each other? No, they often give different values and are not suitable for mutual comparison; the same method should be used when comparing (see 1.3 Note)
What does Method A measure? n-Pentane insolubles, toluene insolubles (see 1.2, 8)
What does Method B measure? n-Pentane insolubles after coagulation, toluene insolubles after coagulation (see 1.3, 9)
What is the sample amount? 10.0 g±0.1 g (see 8.1.1)
What is the relative centrifugal force? 600~700 rcf (see 5.2, 8.1.2, 8.1.3, 8.2.2)
How is the centrifuge speed determined? Calculate according to formula (1) n = 1337√(rcf/d), or select according to the corresponding rotation diameter in Table 2 (see 5.2)
Centrifugation time for n-pentane insolubles? 20 min±1 min (see 8.1.2, 8.1.3)
Centrifugation time for toluene insolubles? 10 min~20 min, until no visible suspended matter floats in the solution (see 8.2.2)
Drying temperature and time? 105 ℃±3 ℃; pre-dry the centrifuge tube 30 min, dry n-pentane insolubles 30 min, dry toluene insolubles 1 h (see 5.3, 8.1.1, 8.1.5, 8.2.4)
Weighing accuracy requirements? Weighing 100 mL beaker and centrifuge tube sensitivity 0.5 mg; sample weighing sensitivity 0.1 g; weigh to the nearest 1 mg (see 5.4, 8.1.1, 8.1.5, 8.2.4)
What standard is sampling based on? GB/T 4756 or SY/T 5317 (see 7.1)
How should the sample be treated before sampling? Heat the original container to 60 ℃±5 ℃, shake until the sediment is uniformly suspended; transfer to a transparent glass bottle if necessary; filter with a 150 μm filter screen (see 7.2)
Can filtration be omitted for a clear, transparent sample with no sediment? Yes (see note after 7.2)
How long can it be left after mixing? Do not exceed 3 h (see 8.1.1)
What is the coagulant? n-Butyldiethanolamine, used as a secondary coagulant (see 6.3)
How is the n-pentane-coagulant solution prepared? 50 mL n-butyldiethanolamine + 50 mL isopropanol added to 1 L n-pentane and mixed (see 6.9)
How is the toluene-ethanol solution prepared? Mix 1 volume of toluene with 1 volume of 95% ethanol (see 6.8)
What is the calculation formula? w = 100(B − A) / m (see 10.1, denominator symbol requires confirmation against the source standard (source-text extraction))
How are insoluble gums obtained? n-Pentane insolubles minus toluene insolubles (see 11.3)
Precision confidence level? 95% (see 12.1)
What is the repeatability of toluene insolubles after coagulation by Method B? Interlaboratory precision is poor, not suitable for comparison of results between laboratories (see 12.1.2.3 Note 2)
Does the method have bias? No bias, because the mass fraction of insolubles is defined only in this standard (see 12.2)
Safety requirements for handling flammable substances? A refrigerated centrifuge or centrifuge tubes with screw caps and seals should be used (see note after 8.1.2)
How should a water-containing sample be treated? Before testing, place the sample in an open beaker and dry it in an oven at 105 ℃±3 ℃ before determination (see 9.1.2)
What water is used to calibrate the centrifuge tube? 20 ℃ distilled water (see 5.1)
Can ordinary centrifuge tubes be used for routine analysis? Ordinary graduated centrifuge tubes made of heat-resistant glass with a capacity of 100 mL may be used (see 5.1 Note)

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.