NB/SH/T 0248-2019 Determination of cold filter plugging point of diesel and domestic heating oils

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The sample is cooled under specified conditions and, through a 2 kPa controllable vacuum device, is drawn through a standard wire mesh filter into a pipette. Each time the sample is 1 ℃ below the previous temperature, this step is repeated until the amount of waxy crystals precipitated in the sample is sufficient to stop flow or reduce the flow rate; the temperature at which the time for the sample to fill the pipette exceeds 60 s or the sample cannot completely return to the test cup is recorded as the cold filter plugging point of the sample (Chapter 4).

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

5.1 Manual apparatus (6.1)

Expected cold filter plugging point Required cooling bath temperature
Higher than −20 ℃ −34 ℃±0.5 ℃
−20 ℃~−35 ℃ −34 ℃±0.5 ℃, then −51 ℃±1.0 ℃
Lower than −35 ℃ −34 ℃±0.5 ℃, then −51 ℃±1.0 ℃, finally −67 ℃±2 ℃

5.2 Automatic apparatus (6.2)

5.3 Appendix A Technical conditions for thermometers (normative appendix)

This appendix gives the dimensional specifications of the digital thermometer (6.2.2). Table A.1 gives the dimensions of each part of the platinum resistance thermometer (PRT), see Figure A.1. For platinum resistance thermometers manufactured with an all-metal stem, the supplier shall confirm that the thermometer dimensions meet the requirements of Table A.1.

Table A.1 Thermometer dimensions:

Identification Description Range/mm
A Probe diameter 4.5±1.0
B Distance from bottom to coil 3.0±2.0
C Length from coil to contact point 12.0±3.0
D Stem outer diameter 6.5±1.0

6. Recommended operating conditions

7. Sampling and test sample preparation

8. Analytical procedure

8.1 Preparation of apparatus (Chapter 10)

8.2 Calibration (Chapter 11)

8.3 Manual apparatus test procedure (12.1)

8.4 Automatic apparatus test procedure (12.2)

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

11. Report

Chapter 15 The test report shall contain at least the following information:

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key points of answer
Which version does this standard replace? Replaces SH/T 0248—2006; previous versions are SH/T 0248—1992, SH/T 0248—2006
Which international standard does this standard adopt? Modified adoption using the redrafting method of IP 309/16
What is the definition of cold filter plugging point? The highest temperature at which, when the test sample is cooled under specified conditions, the test sample cannot pass through the filter, or the time for 20 mL of the test sample to pass through the filter is greater than 60 s, or the test sample cannot flow completely back into the test cup (3.1)
What does the scope of application include? Fatty acid methyl esters (FAME), distillate fuels and synthetic diesel fuels, including fuels for diesel engines and domestic heating installations containing FAME, flow improvers or other additives (Chapter 1)
What is the applicable range of precision? Cold filter plugging point −47 ℃~11 ℃; not applicable outside this range (14.1)
What is the temperature requirement for the test sample before testing? 30 ℃±5 ℃ (Chapter 9, 12.1.5, 12.2.3)
How is the cooling bath temperature selected? According to Table 1: for above −20 ℃ use −34 ℃±0.5 ℃; for −20 ℃~−35 ℃ use −34 ℃±0.5 ℃ then −51 ℃±1.0 ℃; for below −35 ℃ use −34 ℃±0.5 ℃, −51 ℃±1.0 ℃, −67 ℃±2 ℃
Vacuum pressure difference and flow rate? 2 kPa±0.05 kPa (200 mmH₂O±1 mmH₂O), air flow rate 15 L/h±1 L/h (6.1.12, 6.1.13, 12.1.6)
Distance from the bottom of the thermometer to the bottom of the test cup? 1.5 mm±0.2 mm (12.1.4, 12.2.3)
When is the test stopped and “not plugged at −51 ℃” reported? When the test sample has cooled to −51 ℃ and the cold filter plugging point of the test sample has still not been reached (12.1.11)
Repeatability formula? r = 1.2 − 0.027 X₁ (14.2)
Reproducibility formula? R = 3.0 − 0.060 X₂ (14.3)
To how many places is the result reported? Integer (Chapter 13)
Which standards are the basis for sampling? GB/T 4756 or GB/T 27867 (Chapter 8)
Instrument calibration interval? Instruments shall be calibrated at least twice a year; thermometers at least once a year; stopwatches at least once a year; other equipment is recommended to be calibrated once every six months (11.1.1, 11.1.2, 11.2, 11.5)
Where are the technical specifications for digital thermometers? Annex A (normative annex); Table A.1 gives the dimensions of platinum resistance thermometers