NB/SH/T 0639-2024 Determination of Thermal Stability of Petroleum Wax

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Weigh 125g of test sample, place it in a 500mL heat-resistant beaker, oxidize at constant temperature at 170℃ for 30min, cool to a certain temperature, and then determine its Saybolt color number.

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

5.1 Manual apparatus

5.2 Automatic apparatus

5.3 General equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9.1 Manual apparatus

9.2 Automatic apparatus

9. Qualitative and quantitative determination and calculation formula

10. Precision

12.1 Manual apparatus

12.2 Automatic apparatus

12.3 Bias

Since the thermal stability result is obtained by defining a method, the bias of the test results of this method cannot be determined.

11. Report

12. Referenced documents

Normative referenced documents

References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer
Which version does this standard replace? Replaces SH/T 0639—1996 "Determination of thermal stability of paraffin wax".
What change has been made to the standard name? Changed from "Determination of thermal stability of paraffin wax" to "Determination of thermal stability of petroleum wax".
Which products are covered by the scope? Refined paraffin wax products such as semi-refined paraffin wax, fully refined paraffin wax, packaging wax, food additive paraffin wax, etc., as well as non-light white oil products with a flash point (open cup) not exceeding 140℃.
What is the method summary? Weigh 125g of test sample, place in a 500mL heat-resistant beaker, oxidize at constant temperature for 170℃ at 30min, and after cooling measure the Saybolt colour number.
What is the sample amount? 125g±1g.
Oxidation temperature and time? Manual instrument 170℃±2℃, 30min±10s; automatic instrument 170℃±1.0℃, 30min.
Stirring speed? 100r/min±2r/min.
Temperature rise requirements for manual instrument? Within 5min, raise from about 90℃ to 170℃, start timing at 169℃.
Temperature rise requirements for automatic instrument? Within 5min, raise from 90℃ to 169℃, start timing at 169℃.
Post-cooling treatment for paraffin wax sample? Cool at room temperature for 5min, then place in a 85℃±5℃ constant-temperature oven and keep for 15min.
Post-cooling treatment for white oil sample? Cool to room temperature.
Colour measurement method for manual instrument? Determine Saybolt colour according to GB/T 3555.
Colour measurement method for automatic instrument? Determine Saybolt colour according to NB/SH/T 0905; may also use GB/T 3555, but precision has not been investigated.
How are results reported? For manual instrument, take the average of 2 replicate determinations, round to an integer, and report as Saybolt colour number; for automatic instrument, report the Saybolt colour number measured by the instrument.
Which method prevails in arbitration? The thermal stability result from the manual instrument prevails.
Repeatability for manual instrument? The difference between two independent test results is not greater than 1 Saybolt colour number.
Reproducibility for manual instrument? The difference between two independent test results is not greater than 2 Saybolt colour number.
Repeatability for automatic instrument? The difference between two independent test results is not greater than 1.1 Saybolt colour number.
Reproducibility for automatic instrument? The difference between two independent test results is not greater than 2.6 Saybolt colour number.
Frequency of quality control sample determination? Preferably once every 20 samples or once per month for 1.
Requirements for establishing quality control charts? According to NB/SH/T 0843, at least 20 determinations on the same instrument, using control charts or equivalent statistical analysis techniques to obtain the mean and standard deviation.
Which standards are the basis for sampling? For solids, according to SH/T 0229; for liquids, according to GB/T 4756 or GB/T 27867.
Melting requirements for paraffin wax sample? Melt in a water bath or oven to not higher than its melting point 17℃, stir evenly, and the molten state should not exceed 1h.
Beaker specifications? Made of stainless steel or glass, 500mL, diameter 77mm±2mm.
Electronic balance accuracy? 0.1g.
Thermometer requirements? Thermometer for softening point from 30℃ to 180℃, conforming to GB/T 514—2005 number GB-42; or other thermometers containing 170℃, accuracy 0.5℃.
How is bias specified? The bias of the test results of this method cannot be determined.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.