NB/SH/T 0859-2013 Determination of thermal stability of chemical substances by thermal analysis

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

9. Qualitative and quantitative and calculation formulas

10. Precision

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
What method does this standard use? DSC Determine the approximate temperature and enthalpy value of the enthalpy change; DTA determine the approximate temperature of the enthalpy change (1.1)
What forms of test samples are applicable? Solid, liquid, or slurry (1.2)
What is the temperature range? 27 ℃~1200 ℃ (1.3, 11.3)
What is the sample amount range? 1 mg~50 mg (10.1); DSC generally 5 mg, DTA generally 50 mg (11.1)
Recommended sample amount for the first test of an unknown sample? 1 mg or less (11.1 note)
Recommended heating rate? Generally 10 ℃/min–20 ℃/min; when exotherm immediately follows endotherm, 2 ℃/min–6 ℃/min is recommended (11.2)
Common temperature range (most organic substances)? 20 ℃~500 ℃ (12.3)
Gas supply rate? 50 mL/min±5 mL/min (7.2.1.4)
Gas purity requirements? When oxygen has an effect, use nitrogen, helium, or argon with a purity of 99% or above (7.2.1.4 note 1)
What are the reference materials? High-temperature calcined alumina, glass beads, silicone oil, or an empty container (10.3)
Enthalpy change calculation formula? ΔH = A/m, ΔH unit J/g, m unit mg, A unit mJ (13.3 formula 1)
Which temperatures are reported? DSC: extrapolated onset temperature, peak temperature; DTA: peak temperature (4.5, 14.5)
Relative standard deviation of repeatability of reaction enthalpy change? 3.5% (15.2.1)
Relative standard deviation of reproducibility of reaction enthalpy change? 4.7% (15.3.1)
Repeatability standard deviation of extrapolated onset temperature? 0.2 ℃ (15.2.2, requires confirmation against the source standard (source-text extraction))
Reproducibility standard deviation of extrapolated onset temperature? 3.4 ℃ (15.3.2)
Repeatability standard deviation of peak temperature? 27 ℃~527 ℃: 0.4 ℃; 527 ℃~1200 ℃: 0.87 ℃ (A.2, requires confirmation against the source standard (source-text extraction))
Reproducibility standard deviation of peak temperature? 27 ℃~527 ℃: 1.00 ℃; 527 ℃~1200 ℃: 7.98 ℃ (A.3)
Calibration method? According to the method recommended in the instrument manufacturer's user manual or according to GB/T 13464 (9)
Balance requirement? Weighing not less than 100 mg, sensitivity at least 10 μg (7.4)
Temperature sensor accuracy? ±0.5 ℃ (7.2.1.2)
Differential sensor sensitivity? 0.1 ℃ or 0.2 mW (7.2.1.3)
Is the pressure test included? Not included, deleted (foreword)
Main differences from ASTM E537-07? Added DTA method; deleted pressure test; temperature range changed to 27 ℃~1200 ℃; added DTA curve and enthalpy change definitions; added typical DTA curve schematic diagram; deleted initial temperature content; added peak temperature test conditions, procedure and precision; calibration added GB/T 13464; verified with polymers, lubricating oil, lubricating grease, kaolin, molecular sieve, indium and zinc; added appendix A (foreword)

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.