NB/SH/T 0864-2023 Determination of elements in grease by inductively coupled plasma optical emission spectrometry

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

Table 1 Determination range and recommended wavelengths of the elements to be determined

Element Mass fraction range/(mg/kg) Recommended wavelength/nm
Aluminum 10~600 167.038, 308.215, 396.152, 309.271
Antimony 10~2300 206.833, 217.581, 231.147
Barium 50~800 223.540, 233.527, 455.404, 493.409
Boron 500~12600 208.959, 249.677, 249.773
Calcium 20~50000 315.887, 317.933, 364.441, 396.847, 422.673
Iron 10~360 238.204, 259.941
Lithium 300~3200 670.780, 610.364, 460.289
Magnesium 30~10000 279.079, 279.553, 280.278, 285.213
Molybdenum 50~22000 135.387, 202.030, 281.615
Phosphorus 50~2000 177.495, 178.287, 213.618, 214.914, 253.398
Silicon 10~15000 288.158, 251.612
Sodium 30~1500 589.595
Sulfur 1600~28000 182.034, 180.731, 182.625
Zinc 300~2200 202.548, 206.191, 213.856, 334.557, 481.053

Note: When using wavelengths below 190nm to determine elements such as phosphorus and sulfur, vacuum or inert gas purging of the optical path is required.

3. Method principle

Accurately weigh the grease sample, decompose the sample by sulfate ash digestion or closed-vessel microwave digestion, and make up to volume with acid solution. Introduce the acid-soluble standard solution and sample solution into the ICP instrument by siphon or peristaltic pump for measurement, and calculate the content of the elements to be determined in the sample solution by comparing the emission intensities of the elements in the acid-soluble standard solution and the sample solution. (See Chapter 4 Method summary)

4. Reagents and materials (purity / preparation)

All reagents shall use the specified purity. If verified in advance, other grades of reagents that can achieve sufficient purity without affecting the accuracy of the determination may also be used.

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and Sample Preparation

8. Analytical Procedure

Table 2 Concentration and recommended wavelength for internal standard use

Internal standard Approximate concentration when used / (mg/kg) Recommended wavelength / nm
Cobalt 5 238.892
Indium 10 230.606
Scandium 1~2 361.383, 255.237
Yttrium 1~5 371.029, 317.306, 224.306

Note: Other internal standards, wavelengths, and concentrations may also be used.

9. Qualitative and Quantitative Analysis and Calculation Formulas

Calculate the mass fraction of the element to be determined in the test sample according to formula (1).

ω = (ρ × V) / m ····················································· (1)

Where:

Calculate the mass concentration of the element to be determined in the test solution ρ according to formula (2).

ρ = K (Ie − IBe) ·················································· (2)

Where:

(Note: The original formula (1) source-text extraction is displayed in the form “ω = ρV/m”, the formula symbol typesetting has noise, requires confirmation against the source standard (source-text extraction).)

10. Precision

Table 3 Precision of elemental analysis of grease, in milligrams per kilogram

Element Range Repeatability r Reproducibility R
Aluminum 10~600 0.2163X₁^0.9 6.8156X₂^0.9
Antimony 10~2300 0.3051X₁^0.8191 4.6809X₂^0.8191
Barium 50~800 0.3165X₁^0.7528 2.9503X₂^0.7528
Boron 500~12600 0.9727X₁^0.6765 0.1927X₂^0.9992
Calcium 20~50000 2.2853X₁^0.7067 3.0571X₂^0.7067
Iron 10~360 0.8808X₁^0.7475 2.5737X₂^0.7457
Lithium 300~3200 0.0720X₁^1.0352 0.1476X₂^1.0352
Magnesium 30~10000 0.6620X₁^0.6813 2.6155X₂^0.6813
Molybdenum 50~22000 0.1731X₁^0.9474 0.4717X₂^0.9474
Phosphorus 50~2000 1.2465X₁^0.6740 4.0758X₂^0.6740
Silicon 10~15000 1.3859X₁^0.9935 4.8099X₂^0.9935
Sodium 30~1500 6.5760X₁^0.5 11.571X₂^0.5
Sulfur 1600~28000 1.0507X₁^0.8588 1.5743X₂^0.8588
Zinc 300~2200 0.1904X₁^0.8607 0.5912X₂^0.8607

Note: X₁—average of two replicate test results/(mg/kg); X₂—average of two single, independent test results/(mg/kg).

(Note: The original text of the iron element reproducibility index is “0.7457”, which is inconsistent with the repeatability index “0.7475”, requires confirmation against the source standard (source-text extraction). The boron element reproducibility coefficient “0.1927” is obviously smaller than the repeatability coefficient “0.9727”, requires confirmation against the source standard (source-text extraction).)

11. Report

When the element content is less than 1000mg/kg, report the result to three significant figures, in mg/kg; when the element content is greater than or equal to 1000mg/kg, report the result to two significant figures, in mass fraction (%); when the element content is greater than or equal to 10000mg/kg, report the result to three significant figures, in mass fraction (%). (See Chapter 15)

12. Normative References

Normative references:

Bibliography:

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A

Question Key points of answer
Which elements does this standard apply to? Metals: aluminum, antimony, barium, calcium, iron, lithium, magnesium, molybdenum, sodium, zinc; non-metals: boron, phosphorus, silicon, sulfur. It also applies to bismuth, cadmium, chromium, copper, lead, manganese, potassium, titanium, etc., but the precision has not been determined.
What samples does it not apply to? Grease samples that produce visible insoluble matter after sample preparation (assuming that the insoluble matter contains part of the element to be determined).
What are the two sample digestion methods? Sulfated ash digestion method (9.2) and closed vessel microwave digestion method (9.3).
Which digestion method is not suitable for the determination of sulfur? Sulfated ash digestion method (adding sulfuric acid interferes with sulfur determination, and sulfur volatilizes and is lost during combustion ashing).
Which digestion method is not suitable for the determination of barium? Sulfated ash digestion method (adding sulfuric acid interferes with barium determination).
Which digestion method is not suitable for the determination of boron? Sulfated ash digestion method.
Sample mass and ashing temperature for the sulfated ash method? Weigh 1g~2g sample; muffle furnace temperature 525℃±25℃, generally requires 2h.
Sample mass range for the microwave digestion method? 0.1g~1g; when containing volatile or easily oxidizable organic components, the initial sample mass shall not exceed 0.1g.
Requirement for the correlation coefficient of the standard curve? R² value ≥0.9999.
Tolerance for check standard solution? The analysis results of all elements are within the error range of the expected concentration 5%.
Frequency of testing the check standard solution? Test once after every 5 samples analyzed.
Rinsing requirements for samples containing boron? Run water for rinsing for 60s or longer between sample determinations.
Rules for significant figures in result reporting? <1000mg/kg report three significant figures (mg/kg); ≥1000mg/kg report two significant figures (%); ≥10000mg/kg report three significant figures (%).
What are the recommended internal standard elements? Cobalt, indium, scandium, yttrium.
What should be noted when using wavelengths below 190nm to determine phosphorus and sulfur? Vacuum or inert gas purging of the optical path is required.
What is the precision based on? Determined according to GB/T 6683.1, through statistical analysis of interlaboratory analysis results (95% confidence level).