NB/SH/T 0874-2013 Crude petroleum and liquid petroleum products — Laboratory determination of density — Weighing digital liquid density meter method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The main components of the weighing-type density meter include a force sensor, a float, a suspension wire, a temperature sensor, a data processing and display instrument, and a density measuring cylinder. The float, connected to the force sensor by the suspension wire, is statically immersed in the test portion. According to Archimedes' principle, a change in the density of the test portion is converted into a change in the buoyant force exerted by the test portion on the float. The force sensor senses this change in buoyant force through the suspension wire and converts it into an electrical signal. The display instrument receives and processes the electrical signal and displays the density of the test portion on the display instrument.

According to the actual needs of density measurement, the test portion is heated or cooled so that it reaches the test temperature and has sufficient fluidity; the test portion is poured into the density measuring cylinder and stirred thoroughly so that the temperature of the test portion is uniformly distributed; the float is immersed and kept static in the test portion, and after the reading stabilizes, the apparent density and test temperature displayed by the weighing-type density meter are quickly read and recorded. The standard density is converted using petroleum measurement tables (or a weighing-type density meter with petroleum measurement tables embedded). When a stable temperature is required, especially when the difference between the test temperature and the ambient temperature is large, the density measuring cylinder should be placed in a constant-temperature bath or a density measuring cylinder capable of maintaining constant temperature should be used to contain the test portion.

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

The reliability of the density measurement results of this method shall be judged according to the following provisions (confidence level 95%).

12.1 Repeatability: In the same laboratory, by the same operator using the same instrument, and according to the same method, under normal and correct operation on the same sample, the difference between two consecutive determination results shall not exceed the value in Table 1.

Table 1 Repeatability of Petroleum Product Density

Example of petroleum product Temperature range Unit Product characteristics Repeatability
Kerosene, diesel 23~30 °C kg/m³ At normal temperature, weak flow or weak volatility 0.2
Kerosene, diesel 23~30 °C g/cm³ At normal temperature, weak flow or weak volatility 0.0002
Gasoline, lubricating oil, light crude oil with low pour point 20~35 °C kg/m³ At normal temperature, good fluidity and strong volatility, or poor fluidity and weak volatility 0.8 (requires confirmation against the source standard (source-text extraction), the values in the original table are unclear in the source text)
Gasoline, lubricating oil, light crude oil with low pour point 20~35 °C g/cm³ At normal temperature, good fluidity and strong volatility, or poor fluidity and weak volatility 0.0008 (requires confirmation against the source standard (source-text extraction))
High pour point crude oil 40~50 °C kg/m³ At normal temperature, flows or has poor fluidity and light components are easily lost after heating 0.4 (requires confirmation against the source standard (source-text extraction))
High pour point crude oil 40~50 °C g/cm³ At normal temperature, flows or has poor fluidity and light components are easily lost after heating 0.0004 (requires confirmation against the source standard (source-text extraction))

(Note: The original layout of Table 1 source-text extraction is disordered. The above correspondence is organized according to the legible content of the original text. The specific values and row correspondence require confirmation against the source standard (source-text extraction).)

12.2 Reproducibility: In different laboratories, by different operators using different instruments, and according to the same method, under normal and correct operation on the same sample, the difference between the total test results obtained shall not exceed the value in Table 2.

Table 2 Reproducibility of Petroleum Product Density

Example of petroleum product Temperature range Unit Product characteristics Reproducibility
Kerosene, diesel 24~30 °C kg/m³ At normal temperature, good fluidity, nonvolatile or weakly volatile 0.4
Kerosene, diesel 24~30 °C g/cm³ At normal temperature, good fluidity, nonvolatile or weakly volatile 0.0004
Gasoline, lubricating oil, light crude oil with low pour point 20~35 °C kg/m³ At normal temperature, good fluidity and strong volatility, or poor fluidity and weak volatility 0.9
Gasoline, lubricating oil, light crude oil with low pour point 20~35 °C g/cm³ At normal temperature, good fluidity and strong volatility, or poor fluidity and weak volatility 0.0009
High pour point crude oil 40~50 °C kg/m³ At normal temperature, flows or has poor fluidity and light components are easily lost after heating 1.4
High pour point crude oil 40~50 °C g/cm³ At normal temperature, flows or has poor fluidity and light components are easily lost after heating 0.0014

Note: The standard does not provide density data for test temperatures exceeding the ranges specified in Table 1 and Table 2. When the test temperature exceeds the corresponding temperature range, the precision data in Table 1 and Table 2 are for reference only.

Note: The precision data in Table 1 and Table 2 were obtained using a weighing-type density meter with a maximum permissible error of 0.2 kg/m³. When the maximum permissible error of the weighing-type density meter is greater than 0.2 kg/m³, its precision results are expected to be equal to or lower than the data in Table 1 and Table 2; when the maximum permissible error of the weighing-type density meter is less than 0.2 kg/m³, its precision results are expected to be equal to or higher than the data in Table 1 and Table 2. (The original text "9.24em" is suspected to be source-text extraction noise, requires confirmation against the source standard (source-text extraction))

11. Report

The 13 test report shall include at least the following:

12. Normative References

Normative references (Chapter 2):

Bibliography:

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A Reference

Question Key Answer Points
What is the method number of this standard? NB/SH/T 0874—2013
Which samples does this standard apply to? Liquid crude oil, petroleum products, and mixtures of petroleum products and non-petroleum products; crude oil and liquid petroleum products with RVP less than 100 kPa
What is the method principle? Archimedes' principle; the change in buoyancy of the float is converted into an electrical signal by a force sensor, displaying apparent density and test temperature
What is the resolution requirement for density measurement? Not lower than 0.05 kg/m³
What are the resolution and accuracy requirements for temperature measurement? Resolution not lower than 0.01 °C, accuracy not lower than ±0.1 °C
What are the requirements for the outer diameter and length of the float? Outer diameter 20 mm~30 mm, length 80 mm~100 mm
What is the minimum sample volume of the density measuring cylinder? Not less than 300 mL
According to which standard is sampling performed? GB/T 4756 or GB/T 27867
How are volatile samples with RVP greater than 50 kPa homogenized? Homogenize the sample within the original sample container and closed system
What are the heating requirements before homogenization of waxy crude oil? When the pour point is higher than 10 °C or the cloud point is higher than 15 °C, preheat to above the pour point by 9 °C or above the cloud point by 3 °C
What are the heating requirements before homogenization of waxy distillate oil? Preheat to above the cloud point by 3 °C
How is the test temperature for crude oil selected? 20 °C, a higher temperature above the pour point by 9 °C or above the cloud point by 3 °C
What is the requirement for ambient temperature variation? During the test, it shall not exceed 2 °C; otherwise, place the density measuring cylinder in a constant-temperature bath
What are the installation position requirements for the temperature sensor? Horizontally located at the midpoint of the distance from the inner wall of the measuring cylinder to the outer wall of the float, vertically at the middle of the sample height, and at a distance of 5 mm or more from the outer wall of the float
What is the stability criterion for reading the temperature indication? Temperature fluctuation within 30 s shall not exceed 0.1 °C
What is the stability criterion for reading the density and temperature indications? Within 30 s, density fluctuation shall not exceed 0.05 kg/m³, and temperature fluctuation shall not exceed 0.05 °C
How to handle an excessive difference between two temperature measurements? When the difference is greater than 0.5 °C, re-read according to 9.3~9.5 until stable within 0.5 °C
Which table is used to look up standard density? GB/T 1885—1998 Petroleum measurement tables: crude oil—Table 59A, products—Table 59B, lubricating oils—Table 59D
What is the apparent density correction formula? ρt = ρ't [1 + αf (20 − t)], αf is usually 25×10⁻⁶ °C⁻¹
What is the density calculation formula at other temperatures? ρt = ρ20 × VCF20
What are the units and number of digits for reporting results? Report to 0.1 kg/m³ (or 0.0001 g/cm³), 20 °C
How is the final result determined? Take the arithmetic mean of two measured density results
What is the confidence level for precision? 95%
Definition of repeatability? Same laboratory, same operator, same instrument, same method; the difference between two consecutive test results shall not exceed the value in Table 1
Definition of reproducibility? Different laboratories, different operators, different instruments, same method; the difference between two test results shall not exceed the value in Table 2
What is the maximum permissible error of the density meter corresponding to the precision data? 0.2 kg/m³
What does the test report include at least? Sample name, standard number, test results, apparent density and test temperature, other circumstances, test date and test personnel
What are the normative references? GB/T 1885—1998, GB/T 2013—2010, GB/T 4756, GB/T 27867, SH/T 0316—1998
What are the bibliography items? GB/T 1884—1998, standardization research by the Research Institute of Petroleum Processing (2010 year 8 month), JJG 999—2005