NB/SH/T 0886-2014 Determination of pour point of petroleum products — Automatic tilt method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

After the preheating preparation work, place the test sample into the automatic pour point tester. After the program starts, the test sample is cooled according to the cooling steps listed in Table 1 and Figure 1, and the pour point is detected once every 1 ℃ or 3 ℃ by the automatic pour point tester; the lowest flowing temperature measured for the test sample is taken as the pour point. During the cooling process, the instrument tilts the pour point tube, while an optical instrument is used to check the movement of the test sample surface. Note: if the preheating step of the automatic pour point tester is used, after the test sample is placed and the program is started, the instrument will automatically run the preheating preparation work.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and test sample preparation

8. Analysis procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

11. Report

According to this standard, report the pour point result of the sample according to the temperature recorded in 11.10, the test temperature interval, and the instrument model.

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
What does this standard apply to? Determination of pour point of petroleum products, applicable range -66 ℃ to 51 ℃
What does it not apply to? Determination of pour point of crude oil
What international standard is adopted? Adopted by redrafting method with modifications to ASTM D5950-12
Main differences from ASTM D5950-12? Some referenced standards adopt the corresponding national standards of our country; added 8.3 clause on residual fuel treatment method; deleted 13.5 clause
What is the method principle? After preheating preparation, the sample is placed in an automatic pour point tester, cooled according to the cooling procedure in Table 1 and Figure 1, and the pour point is detected every 1 ℃ or 3 ℃; an optical instrument checks the movement of the sample surface, and the lowest flow temperature is taken as the pour point
What are the measurement temperature intervals? 1 ℃ or 3 ℃
What is the resolution? 1.0 ℃
How is pour point defined? Under specified conditions, the lowest temperature at which the observed specimen can flow
How is no-flow point defined? Under test conditions, the temperature at which the specimen, due to wax structure, increased viscosity, or both acting simultaneously, prevents movement of the specimen surface
How is tilt defined? The movement technique in which the pour point tube is moved from a vertical position toward a horizontal position, causing the specimen to move
How is the pour point calculated? The temperature of the no-flow point plus 1 ℃ or 3 ℃ (depending on the selected test interval temperature)
What standard is sampling based on? GB/T 4756 or SY/T 5317
How is residual fuel handled before testing? If it has been heated above 45 ℃ before testing or its heating history is unknown, it must be left at room temperature for 24 h
What is the upper heating limit for the test sample? It must not exceed 70 ℃
How is pretreatment done when the expected pour point is ≤ -33 ℃? Heat in a constant-temperature bath or oven at 48 ℃ to 45 ℃
How is pretreatment done when the expected pour point is > -33 ℃? Heat to 9 ℃ above the expected pour point, or above 45 ℃, but not higher than 70 ℃
What is the position of the cork gasket? 25 mm±3 mm from the bottom of the pour point tube
What are the temperature detector requirements? IEC 60751 grade A, able to operate normally from +70 ℃ to -80 ℃, with the platinum resistance tip immersed 3 mm below the oil surface
What are the pour point tube specifications? Cylindrical glass, flat bottom, outer diameter 34 mm±0.5 mm, wall thickness 1.4 mm±0.15 mm, height 120 mm±0.5 mm, bottom thickness 2.0 mm±0.5 mm, with a scribe line 54 mm±0.5 mm from the inner bottom
What are the specifications of the pour point tube jacket? Copper, cylindrical, flat bottom, depth 113 mm±0.2 mm, inner diameter 45 mm±0.1 mm
What are the cooling system requirements? An external cooling system equipped with a circulation pump, with the cold bath temperature at least 10 ℃ lower than the pour point tube jacket temperature; or an internal cooling system maintaining the required jacket temperature
What is the repeatability? 3 ℃ interval r=3.9 ℃; 1 ℃ interval r=2.7 ℃
What is the reproducibility? 3 ℃ interval R=6.1 ℃; 1 ℃ interval R=4.5 ℃
Has bias been established? It has not been established, because there is no recognized reference material to determine bias
What is the difference in results between the 1 ℃ and 3 ℃ intervals? The test results for the 1 ℃ interval are on average 1.1 ℃ lower than those for the 3 ℃ interval
What is the report content? The temperature recorded in accordance with clause 11.10, as well as the test temperature interval and instrument model
Has the application to residual fuel been confirmed? Note: The application of this standard to residual fuel samples has not yet been confirmed; for details see chapter 13