NB/SH/T 0889-2014 Determination of normal paraffin content and carbon number distribution in liquid paraffin products and raw materials — Gas chromatography

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Inject the sample into a gas chromatograph. The vaporized sample is separated in the chromatographic column according to boiling point and then enters a flame ionization detector for detection. After the detected signal is processed by baseline maintenance integration and valley-to-valley integration, the normal paraffin content and carbon number distribution are obtained.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

Table 1 Chromatographic operating conditions (original text Table 1)

Item Condition
Injector temperature/℃ 290
Split ratio 150:1
Sample amount/μL 0.2
Column temperature program: initial temperature/℃ 100
Equilibration time/min 0.5
Initial hold time/min 0
Heating rate/(℃/min) 5.5
Final temperature/℃ 300
Final temperature hold time/min 5
Detector type Flame ionization
Detector temperature/℃ 320
Carrier gas type Helium/hydrogen (original text “ammonia/hydrogen”, requires confirmation against the source standard (source-text extraction))
Carrier gas flow rate/(mL/min) 0.5

Formula (1): R = 2(tR C9 − tR C8) / [1.699(w C9 + w C8)]

Where:

7. Sampling and sample preparation

8. Analysis procedure

Formula (2): S = A / (C16 × V)

Where:

9. Qualitative and quantitative analysis and calculation formulas

Formula (3): Ci = (Ai / A) × 100

Where:

Formula (4): C = ΣCi

Where:

10. Precision

Judge the reliability of test results according to the following provisions (95% confidence level).

12.1 Repeatability (r): the difference between two consecutive determination results on the same sample by the same laboratory, the same operator, using the same instrument and equipment, under the same experimental conditions shall not exceed the following values.

Where: X—— average of two determination results, % (mass fraction).

12.2 Bias: because there is no acceptable reference standard material, the bias of these steps cannot be determined.

11. Report

Report the mass fraction of each normal alkane and the total mass fraction of normal alkanes in the sample, accurate to 0.01% (mass fraction).

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
Standard number and year NB/SH/T 0889—2014
Whether first release Yes, this standard is the first release
Release and implementation dates Released on 2014-06-29, implemented on 2014-11-01
Applicable objects Normal alkane content and carbon number distribution in liquid paraffin products and raw materials
Method principle Gas chromatographic separation, FID detection, baseline-maintained integration and valley-to-valley integration processing
Carbon number range of standard solution for qualitative analysis nC8~nC22 (qualitative step 8); standard solution preparation selects even-carbon-number normal alkanes in the range nC5~nC22 (requires confirmation against the source standard (source-text extraction))
Chromatographic column Methyl cross-linked siloxane capillary column, 50 m × 0.20 mm × 0.50 μm (requires confirmation against the source standard (source-text extraction))
Injection volume 0.2 μL
Split ratio 150:1
Injector temperature 290℃
Detector type and temperature FID, 320℃
Carrier gas Helium or hydrogen, flow rate 0.5 mL/min (requires confirmation against the source standard (source-text extraction))
Column oven temperature program 100℃ initial, equilibration 0.5 min, initial hold 0 min, 5.5℃/min rise to 300℃, final temperature hold 5 min
Chromatographic column resolution requirement R not less than 30
Detector minimum response requirement S greater than 0.08 pA·s/ng (nC16) (requires confirmation against the source standard (source-text extraction))
Quantification formula Ci=(Ai/A)×100; C=ΣCi
Reporting precision Accurate to 0.01% (mass fraction)
Repeatability Three-tier formulas for C13~C15, C14~C22, and total; total r=1.20% (mass fraction) (subscript requires confirmation against the source standard (source-text extraction))
Bias No available reference standard material, bias cannot be determined
Sampling reference GB/T 4756 Petroleum liquids—Manual sampling

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.