Origin Standard

NB/SH/T 0916-2015 Determination of biodiesel (fatty acid methyl esters) content in diesel fuel — Infrared spectroscopy method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis Steps

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

11. Report

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
What does this standard measure? Biodiesel (fatty acid methyl esters FAME) content in diesel fuels.
Applicable concentration range? Volume fraction 1%~20%; can be extended to 1%~100% with appropriate ATR accessories, but precision is not applicable when exceeding 20%.
Is it applicable to FAEE? Only applicable to FAME; biodiesel containing FAEE will produce a negative bias.
Method principle? Fourier transform mid-infrared spectroscopy + attenuated total reflection (ATR) sample cell + partial least squares (PLS).
Spectral acquisition range? 4000 cm⁻¹~650 cm⁻¹.
Resolution requirement? 4 cm⁻¹.
Sample temperature requirement? 15 ℃~27 ℃.
Spectral intervals for low-concentration model? 1800 cm⁻¹–1692 cm⁻¹ and 1327 cm⁻¹–940 cm⁻¹, without baseline correction.
Spectral intervals for medium-concentration model? 1800 cm⁻¹–1700 cm⁻¹ and 1399 cm⁻¹–931 cm⁻¹, without baseline correction.
Spectral intervals for high-concentration model? 1851 cm⁻¹–1670 cm⁻¹ and 1371 cm⁻¹–1060 cm⁻¹, without baseline correction.
Number of latent variables for modeling? 3 (principal factors); 4 may also be used but the reason should be recorded.
Reporting precision? Volume fraction (%), to the nearest 0.01%.
Repeatability formula? Repeatability = 0.01505 (X + 14.905).
Reproducibility formula? Reproducibility = 0.04770 (X + 14.905).
Sampling basis? GB/T 4756.
Density determination basis? GB/T 1884.
BD100 specification basis? GB/T 20828—2014.
Middle distillate fuel specification basis? GB 252—2011.
Quality control frequency? Measure at least one biodiesel quality control sample every day.
When to calibrate/inspect the instrument? At initial use, recalibration, and repair.
Prohibition for cleaning the ZnSe crystal? Do not clean with acid or alkali.