NB/SH/T 0812-2010 Determination of 2- furfural and related components in mineral insulating oil

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formulas

10. Precision

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A

Question Key answer points
What are the method number and title of this standard? NB/SH/T 0812—2010 Determination of 2- furfural and related components in mineral insulating oil
What is the status of adoption of the international standard? Modified adoption of IEC 61198:1993 (MOD), first issue
What is the applicable object? Detection of furfural content in used mineral insulating oil; may also be applicable to new mineral insulating oil
Method principle? Liquid–liquid extraction of furan derivatives, the extract is introduced into a high-performance liquid chromatograph (HPLC) for analysis
Detector? Ultraviolet (UV) detector, a diode array UV detector is recommended
Chromatographic column? Reversed-phase octyl column or octadecyl column; a guard column with the same packing as the analytical column may be used
Injection volume? 10 μL~20 μL; precision syringe capacity 50 μL~250 μL
Balance sensitivity? 0.1 mg
Sample amount for extraction? About 20 mL, accurate to 0. (value incomplete, requires confirmation against the source standard (source-text extraction))
Extraction shaking time? Shake on a mechanical shaker for at least 2 min, let stand until the acetonitrile and oil phases completely separate
Mobile phase? 60% water—40% methanol or acetonitrile (volume fraction, requires confirmation against the source standard (source-text extraction))
Detection wavelength? 276 nm (Figure A.2 also labels 220 nm and 276 nm, quantitative wavelength requires confirmation against the source standard (source-text extraction))
Quantification method? Single-point external standard method (formula incomplete, requires confirmation against the source standard (source-text extraction))
Result unit? mg/kg
Lower detection limit? Minimum detectable concentration of each analyte to be determined 0.05 mg/kg or smaller; oxidized samples may not reach it
Repeatability? <1.0 mg/kg is 0.1X; ≥1.0 mg/kg is 0.05X (X is the average value)
Reproducibility? <1.0 mg/kg is 0.15X; ≥1.0 mg/kg is 0.1X (X is the average value)
Sampling reference standards? GB/T 4756, GB/T 7597
Laboratory water reference standard? GB/T 6682
Which contents of IEC were not adopted? Not adopted IEC 61198: the solid-liquid extraction content of method B in 1993 and the related appendix A and figure 1 contents
Appendix situation? Appendix A and Appendix B are both informative appendices