NB/SH/T 0908-2015 Determination of Stability and Compatibility of Heavy Fuel Oils and Crude Oils Optical Detection Method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The stability and compatibility parameters are determined by optically detecting the initial asphaltene flocculation point during titration. First, prepare three stock solutions of xylene-sample in different proportions, then titrate with n-heptane to promote asphaltene precipitation. During titration, the mixed liquid continuously circulates through the optical detector, and asphaltene flocculation is judged by detecting changes in the scattering intensity of visible light. The stability parameters are calculated from the amounts of sample, xylene, and n-heptane used: xylene equivalent value SE, P value, and FRs1. If the density is known, the compatibility parameters of crude oil (SBN and IN) can also be calculated.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

9. Qualitative and quantitative and calculation formulas

10. Precision

Table 1 Repeatability requirements

Parameter Repeatability r Recommended precision applicability range
P 0.103X 1.28~4.54
Pa 0.03 0.33~0.83
Po 0.2 0.43~1.31
FRs1 0.03 0.05~0.62

Note 1: X—— the average of two determination results. Note 2: Method for determining the recommended precision range: (minimum sample average - repeatability) to (maximum sample average + repeatability).

11. Report

The report shall provide the following information:

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A Reference

Question Key points of answer
What does this standard measure? Stability and compatibility parameters of heavy fuel oils and crude oils (SE, P, FRs1; for crude oil, IN and SBN can also be calculated)
What is the principle of the method? The xylene-sample standby solution is titrated with n-heptane, and the initial asphaltene flocculation point is optically detected
Which samples are applicable? Atmospheric and vacuum distillation, thermal processing, catalytic, and hydrocracking residual oils; GB 25989—2010 in F-R1~F-R4; GB/T 29114—2012 petroleum fuels containing heavy fractions; crude oil
Which samples are not applicable? Samples with asphaltene mass fraction < 0.05%; unstable samples that have already undergone asphaltene flocculation cannot obtain a titration endpoint
What are the aromatic solvent and n-alkane solvent respectively? Xylene and n-heptane
How is the standby solution prepared? 40 g sample + 20 g xylene, weighed to 0.01 g, shaken for at least 2 min, heated if necessary (not exceeding 80 ℃) or refluxed
How is the titration method selected? For high-stability samples, choose a low flocculation ratio (dilution ratio 1:2, 1:1, 2:1); for low-stability samples, choose a high flocculation ratio (dilution ratio 2:1, 3:1, 4:1)
What is the recommended cup capacity? Maximum 55 mL (actual cup capacity 100 mL)
Pump calibration frequency? At initial installation and during use; once a month during continuous use
Evaporation correction factor calibration frequency? Once a quarter during continuous use
QC sample requirements? Stable and uniform residual fuel oil or crude oil with asphaltene mass fraction 1%~10%, asphaltenes determined according to SH/T 0266
How to check repeatability? See Table 1: P is 0.103X, Pa is 0.03, Po is 0.2, FRs1 is 0.03
Is there reproducibility? Not investigated, because it is irrelevant to the application of this method
Is there bias? No bias (the results are defined only in this standard)
What should the report include? Sample name, SE/P/FRs1, IN/SBN (if measured), this standard number, measurement temperature when above room temperature
What standard is sampling based on? GB/T 4756 or GB/T 27867
How should samples be stored? Sealed, light-blocking container (metal can or brown glass bottle), minimize exposure to air and light above 25 ℃