SH/T 0209-1992 Determination of Thermal Stability of Hydraulic Oil

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The sample is maintained at 135℃ for 168h in the presence of copper and steel rod catalysts. The thermal stability of the sample is evaluated through analysis and determination of the sample, copper rod, and steel rod before and after the thermal test.

4. Reagents and materials (purity / preparation)

4.1 Reagents

Number Reagent Purity / preparation
5.1 Petroleum ether Analytical reagent, 60~90℃
5.2 Acetone Analytical reagent
5.3 Sodium hydroxide Analytical reagent, prepared as 200g/L solution
5.4 Potassium cyanide Prepared as 100g/L solution

Note: In 7.2.3.3, 200g/L sodium hydroxide solution is additionally used.

4.2 Materials

5. Instruments and equipment

Number Instrument Specification / description
4.1.1 Electric heating automatic convection constant temperature oven Can control 135℃±1℃; volume not less than 470mm×425mm×372mm
4.1.2 Aluminum bath Structure as shown in the figure, material is Al-3 aluminum ingot
4.1.3 M-50 type sand core filtration movable device Note: Its glass sand core can be replaced with a self-made porous (pore size unclear in the source text (source-text extraction)) heat-resistant plastic filter plate
4.1.4 Vacuum pump Specification not provided in the available source
4.1.5 Beaker 100mL, 250mL
4.1.6 Acid-resistant funnel G4 type
4.1.7 Constant temperature drying oven Can maintain temperature at 105~110℃
4.1.8 Thermometer 0~100℃, graduation value 1℃; 0~250℃, graduation value 1℃
4.1.9 Graduated cylinder 200mL
4.1.10 Weighing bottle 40mm×70mm
4.1.11 Pipette 25mL
4.1.12 Wide-mouth bottle 200mL (the volume unit is unclear in the source transcription and remains unverified)
4.1.13 Wash bottle Specification not provided in the available source

Other vessels: culture dish, watch glass, desiccator, suction filtration flask, glass microporous filter (specification not provided in the available source).

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

8.1 Preparation work

8.2 Thermal test process (7.1)

8.3 Evaluation of copper rod and steel rod (7.2)

8.4 Sample evaluation (7.3)

9. Qualitative and Quantitative and Calculation Formulas

8.1 200mL Copper Rod Mass Loss of Sample X1 (mg/200mL)

X1 = (m1 − m6) / 1000 …… (1)

Where:

8.2 200mL Steel Rod Mass Loss of Sample X2 (mg/200mL)

X2 = (m2 − m8) / 1000 …… (2)

Where:

8.3 200mL Mass of Deposit on Steel Rod of Sample Y1 (mg/200mL)

Y1 = (m7 − m8) × 1000 …… (3)

Where:

8.4 100mL Total Sediment Mass of Sample Y2 (mg/100mL)

Y2 = [(m9 − m3)/2 + (m10 − m4) × 4 + (m5 − m6)/2] × 1000 …… (4)

Where:

8.5 Percentage Change in Kinematic Viscosity of Sample Before and After Heat Test Z1 (%)

Z1 = (v2 − v1) / v1 × 100 …… (5)

Where:

8.6 Percentage Change in Acid Number of Sample Before and After Heat Test Z2 (%)

Z2 = (A2 − A1) / A1 × 100 …… (6)

Where:

Qualitative Evaluation

10. Precision

11. Report

Take the arithmetic mean of two duplicate determination results for the sample and catalyst test items respectively as the determination result of thermal stability of hydraulic oil. These items are: the calculated values of clauses 8.1 to 8.6 and the appearance of copper and steel rods, eight items in total.

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key Points of Answer
What oils does this standard apply to? It applies to various hydraulic oils based on mineral oil or synthetic oil.
What are the heat test conditions? Under catalysis by copper rod and steel rod, 135℃±1℃, 168h, no air blowing required.
What is the sample amount? 200mL sample is poured into a 250mL beaker.
What are the catalysts? Copper rod (T2Y copper, containing copper 99.90%, φ6.25mm×76.20mm) and steel rod (T10A steel, containing carbon 1%, φ6.25mm×76.20mm).
When can the metal rods no longer be used? When the diameter is less than 6.20mm or when they cannot be polished bright.
Which items need to be measured before the heat test? 100℃ kinematic viscosity (GB/T 265, recorded as v1), acid number (GB/T 7304, recorded as A1), weighing of copper rod and steel rod (m1, m2).
How is the copper rod appearance evaluated? Visually evaluated against the reference card for copper rod appearance in the determination of thermal stability of hydraulic oil.
How is the steel rod appearance evaluated? Visually evaluated against the reference card for steel rod appearance in the determination of thermal stability of hydraulic oil.
How is the copper rod deposit removed? At room temperature, place in 100g/L potassium cyanide solution to completely remove the deposit.
How is potassium cyanide waste liquid treated? Every 10mL of waste liquid can be completely decomposed by 1.8g potassium chlorate in alkaline conditions; the solution can be reused three to four times.
How is the lacquer-like precipitate on the steel rod removed? 200g/L sodium hydroxide solution, 82℃±1℃, 15min, then wipe off with silk cloth.
How is the sediment determined? Use a M-50 type sand core filtration movable device for vacuum filtration, wash with petroleum ether, and constant weight at 105 to 110℃.
What is the formula for total sediment mass? Y2 = [(m9−m3)/2 + (m10−m4)×4 + (m5−m6)/2] × 1000.
What is the formula for percentage change in viscosity? Z1 = (v2−v1)/v1 × 100.
What is the formula for percentage change in acid number? Z2 = (A2−A1)/A1 × 100.
How is precision specified? Viscosity same as GB/T 265, acid number same as GB/T 7304, this standard does not specify precision.
What items are included in the report? The calculated values of clauses 8.1 to 8.6 and the appearance of copper and steel rods, eight items in total, taking the arithmetic mean of two duplicate determination results.
How many decimal places are retained in the results? 8.1 to 8.4 retain one decimal place; 8.5 to 8.6 retain two decimal places.
What is the time limit for analysis after the heat test? Sample analysis shall be completed within 24h; catalyst analysis shall be completed within 48h.
Which standard does it replace? It replaces ZB E39 007-89.