SH/T 0228-1992 Determination of barium, calcium, and zinc content in lubricating oils (atomic absorption spectrometry)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The sample is diluted with an organic solvent in a certain proportion and directly sprayed into the flame of an atomic absorption spectrometer to measure absorbance. The absorbance of the sample is compared with the absorbance of the reference solution to calculate the barium, calcium, and zinc content in the sample.

4. Reagents and materials (purity / preparation)

4.1 Materials

4.2 Reagents

4.3 Solution preparation

5. Instruments and equipment

6. Recommended operating conditions

Operate the instrument according to the conditions in Table 2 (source note: instrument performance varies among different models; the above conditions are for reference only):

Element Analytical line wavelength nm Lamp current mA Slit nm Burner height mm Air flow rate L/min Acetylene flow rate L/min Expansion factor Extraction amount mL/min
Ba 553.5 9 0.95 10 10 1.8 5 2.6
Ca 422.7 8 0.38 6 10 2-3 1 2-6
Zn 213.9 6 0.38 not provided in the available source 10 2.1 1 2.6

Note: In Table 2, the acetylene flow rate "2-3", extraction amount "2-6" for Ca, and the burner height for Zn are missing or unclear in the source, requires confirmation against the source standard (source-text extraction).

7. Sampling and sample preparation

8. Analytical procedure

8.1 Prepare standard working solutions (7.1)

Respectively take appropriate amounts of barium, calcium, and zinc standard stock solutions into 25 mL volumetric flasks, dilute with blank solution to reach the concentrations given in Table 1.

Table 1 Concentrations of standard working solutions (μg/mL):

Element Concentration points
Ba 20, 40, 60, 80, 100
Ca 2, 4, 6, 8, 0.1, 0.8, 1.2, 1.6, 2.0 (the source arrangement is suspected to be misaligned, requires confirmation against the source standard (source-text extraction))
Zn no explicit value provided in the available source (requires confirmation against the source standard (source-text extraction))

Note: ① The standard working solutions must be prepared on the day of the test. ② The calcium and zinc standard working solutions can be brought to the required concentrations by secondary dilution.

8.2 Measure absorbance (7.3)

  1. Operate the instrument according to the conditions in Table 2.
  2. Zero the instrument with the blank solution.
  3. Measure the absorbance of the standard working solution with the lowest barium, calcium, and zinc content.
  4. Measure the absorbance of the standard working solution with medium barium, calcium, and zinc content.
  5. Measure the absorbance of the standard working solution with the highest barium, calcium, and zinc content.
  6. Measure the absorbance of the sample solution, continuously measuring twice.
  7. Measure the absorbance of the standard working solution with the highest barium, calcium, and zinc content.
  8. Measure the absorbance of the standard working solution with medium barium, calcium, and zinc content.
  9. Measure the absorbance of the standard working solution with the lowest barium, calcium, and zinc content.
  10. Zero with blank.
  11. Repeat steps 7.3.3~7.3.10, and use the average of the two absorbance values as the measurement result.

9. Qualitative and quantitative determination and calculation formula

The barium, calcium, and zinc content X [% (m/m)] in the sample is calculated according to the following formula:

X = (A / A₀) × C × 25 / (f × m × 10⁶) × 100

Where:

(The source formula source-text extraction contains noise; the symbols shall be interpreted according to the context, requires confirmation against the source standard (source-text extraction).)

10. Precision

Use the following provisions to judge the reliability of the measurement results (95% confidence level).

10.1 Repeatability (Table 3)

The difference between two results obtained by the same operator in repeated measurements shall not exceed the value specified in Table 3:

Element Content % (m/m) Percentage of arithmetic mean %
Ba 0.10~0.80 5
Ca 0.03~0.40 5
Zn 0.02~0.16 5

10.2 Reproducibility (Table 4)

The difference between two results submitted by different laboratories shall not exceed the value specified in Table 4:

Element Content % (m/m) Percentage of arithmetic mean %
Ba 0.10~0.80 12
Ca 0.03~0.40 12
Zn 0.02~0.16 not provided in the available source (requires confirmation against the source standard (source-text extraction))

11. Report

Take the arithmetic mean of two repeated measurement results as the measurement result; if the result is greater than 0.10% (m/m), take two decimal places; if the result is less than 0.10% (m/m), take three decimal places.

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
What does this standard measure? Barium, calcium, and zinc content in unused lubricating oils.
Method principle? After dilution with organic solvent, directly sprayed into the flame; absorbance measured by atomic absorption spectrometry and quantified by comparison with a reference solution.
Applicable concentration range? Ba 0.1–0.8%, Ca 0.03–0.4%, Zn 0.02~0.16% (m/m).
What oil products are not applicable? Oil products containing aluminum additives and acetate additives.
Standard stock solution concentration? Barium, calcium, and zinc standard stock solutions of 5000 ppm.
When is the standard working solution prepared? Must be prepared on the day of the test.
Sample weighing amount? 0.1~0.3 g, accurate to 0.1 mg; for trace analysis, can be increased to 2 g.
How is the blank solution prepared? About 0.18 g benzyltrimethylammonium chloride + 1 g sodium dodecyl sulfonate dissolved in 1 L solvent naphtha, heated to saturation at 95±5 ℃, use the clear liquid.
Repeatability requirement? The difference between two results is 5% of the arithmetic mean of ≤ (95% confidence level).
Reproducibility requirement? Ba, Ca is 12%; Zn not provided in the available source (requires confirmation against the source standard (source-text extraction)).
Result value rule? >0.10% take two decimal places; <0.10% take three decimal places.
What are the referenced standards? SH 0005, SH 0007.
Which standard does it replace? ZB E60 007—89.
Which foreign standard is referenced? DIN 51391 (81) T1.