Origin Standard

SH/T 0242-1992 Determination of lead content in light petroleum products (atomic absorption spectrometry)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Lead in the sample is extracted with iodine monochloride. The extract is boiled and cooled to room temperature; then excess potassium iodide is added, forming a lead iodide complex, and the iodine produced is reduced with ascorbic acid. Finally, the iodine complex is back-extracted with methyl isobutyl ketone and determined with an air-acetylene flame atomic absorption spectrophotometer.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formula

The lead content of the sample X [ng/g (ppb)] is calculated according to the following formula:

X = (C - C₀) × V₁ × 10³ / (V × ρ₂₀)

Where:

(Note: The original formula source-text extraction contains noise; the symbols have been inferred and organized from context, requires confirmation against the source standard (source-text extraction).)

10. Precision

Judge the reliability of test results according to the following provisions (95% confidence level).

Sample lead content range Repeatability Reproducibility
10 to 50 7 11
>50 to 120 10% of the arithmetic mean 15% of the arithmetic mean

(Note: The original table source-text extraction contains noise; the units and meanings of "7" and "11" require confirmation against the source standard (source-text extraction).)

11. Report

Take the arithmetic mean of two repeated determination results as the determination result of this standard.

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference Q&A

Question Key answer points
What does this standard apply to? Determination of lead content in light petroleum products such as naphtha and unleaded gasoline.
What is the method principle? Lead is extracted with iodine monochloride, potassium iodide forms a lead iodide complex, ascorbic acid reduces it, MIBK back-extraction, air-acetylene flame atomic absorption determination.
Lowest detectable concentration? The lowest measurable lead concentration for a treated 500g sample is 1ng/g (10ppb) (requires confirmation against the source standard (source-text extraction)).
Analytical line wavelength? 283.3nm.
Working curve range? 0.5 to 4.0μg/mL.
Lamp current? 5 to 10mA.
Air to acetylene flow ratio? 8:4 (fuel-lean flame), after injection of MIBK it becomes a stoichiometric flame.
How soon after sampling is the analysis performed? For tank sampling, follow GB/T 4756; the sample taken should be analyzed and determined within 24h.
Within what time must the sample solution be measured? This sample solution must be measured within 8h.
Limit for blank determination value? Should not be greater than 0.2μg/mL.
How is the result value taken? Take the arithmetic mean of two results from duplicate determinations.
Precision conditions? 95% confidence level, divided into repeatability and reproducibility.
Which standards are cited? GB/T 1884, GB/T 4756.
Precautions for iodine monochloride? Rubber stoppers must not be used; contact with ammonia or ammonium salts is absolutely prohibited.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products