SH/T 0245-1992 Determination of aromatic hydrocarbon content in solvent naphtha (chromatographic method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Use a polar chromatographic column, detect by backflush with a flame ionization detector, then calculate the aromatic hydrocarbon content by the external standard area method.

4. Reagents and materials (purity / preparation)

4.1 Reagents

4.2 Materials

4.3 Preparation of external standard sample (5.2)

Weigh a certain amount of 1,3,5- trimethylbenzene (accurate to 0.0002g) in a stoppered conical flask, dilute with standard n-heptane or aromatic-free C₅~C₇ alkanes to a mass percentage concentration close to the aromatic hydrocarbon content of the test sample, then weigh (accurate to 0.0002g). Calculate its exact concentration, then inject into an ampoule for storage.

5. Instruments and equipment

6. Recommended operating conditions

Typical operating conditions (6.1):

Determination of chromatographic column resolution (5.4): under the specified conditions, inject with a microsyringe a n-heptane solution containing 1%(V/V) each of toluene and decalin; the measured resolution between toluene and decalin shall be not less than 1.0.

7. Sampling and test sample preparation

8. Analysis procedure

6.1 Start the instrument; after the baseline is stable, analysis can be carried out. 6.2 First, use a clean 10μL microsyringe to draw 2μL of the external standard sample, and accurately inject all of it into the injection port; after observing on the recording paper that the non-aromatic hydrocarbon chromatographic peaks have completely emerged and the baseline has recovered, immediately turn the “backflush valve” until the aromatic hydrocarbon chromatographic peaks have completely emerged. Adjust the “attenuation” according to the size of the aromatic hydrocarbon chromatographic peak so that the aromatic hydrocarbon peak height is not less than 2mV. 6.3 Use a clean 10μL microsyringe to draw 2μL of the test sample, carry out the determination of aromatic hydrocarbon content in the test sample according to clause 6.2, and after the aromatic hydrocarbon chromatographic peaks have completely emerged, turn off the recorder switch. A typical chromatogram is shown in Figure 2 (1— non-aromatic hydrocarbons; 2— aromatic hydrocarbons).

9. Qualitative and quantitative determination and calculation formula

The aromatic hydrocarbon content x [%(m/m)] in the test sample is calculated according to the following formula:

x = (A₁ / A) × P

Where:

10. Precision

Repeatability: the difference between two results of repeated determinations by the same operator shall not exceed the following values.

Aromatic hydrocarbon content, %(m/m) Repeatability, %(m/m)
<1 0.1
1~<3 0.2
3~10 0.5
>10 0.6

11. Report

Take the arithmetic mean of the two results of repeated determinations as the aromatic hydrocarbon content of the test sample.

12. Referenced documents

not provided in the available source.

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
What is this standard applicable to? Solvent naphtha of the 145~200℃ fraction.
What detector is used? Flame ionization detector.
What quantitative method is used? External standard area method.
Chromatographic column specifications? Stainless steel, length 2m, internal diameter 4mm.
What is the stationary liquid? N,N- bis(β- cyanoethyl)formamide.
What is the support? 6201 support, 60~80 mesh.
What is used to prepare the external standard sample? 1,3,5- trimethylbenzene, diluted with standard n-heptane or aromatic-free C₅~C₇ alkanes.
What is the injection volume? 2μL.
Resolution requirement? The resolution between toluene and decalin shall be not less than 1.0.
Calculation formula? x = (A₁/A) × P.
What is the precision? Segmented by aromatic hydrocarbon content: <1 is 0.1; 1~<3 is 0.2; 3~10 is 0.5; >10 is 0.6.
What value is taken for the report? The arithmetic mean of the two results of repeated determinations.
Which standard does it replace? SY 2307-82.
In what year was it first issued? 1978.