Origin Standard

SH/T 0293-1992 Determination of Saturated Vapor Pressure of Vacuum Greases

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

A method (Chapter 3)

B method (Chapter 11)

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

A method (Chapter 6)

B method (Chapter 14)

9. Qualitative and Quantitative and Calculation Formulas

A Method (Chapter 7)

B Method

10. Precision

A Method (Chapter 8)

B Method (Chapter 15)

11. Report

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
How many methods does this standard contain? Two: A method suspension pendulum method; B method evaporation method.
Which samples is it applicable to? Vacuum pump oil (grease), diffusion pump oil (grease), and silicone oil for diffusion pumps.
In what are the results expressed? Expressed in millimeters of mercury at 20°C.
Constant-temperature oil bath temperature for the A method? Can be kept constant at 160±0.5°C.
Sample amount for the A method? 5~10ml, added to the evaporation chamber from the nozzle.
How many times is the origin determination repeated in the A method? Repeated five times, and the average is taken as the origin S₀.
Temperature rise interval and constant-temperature time for each point in the A method? Approximately for every increase of 5°C, keep constant temperature for 3min and determine one point.
Minimum number of points to measure in the A method? Each test determines at least six points or more.
Calculation formula for the A method? P = 0.00136 × (m·S·d)/A (original formula source-text extraction is incomplete, requires confirmation against the source standard (source-text extraction)).
Precision requirement? The difference between two results of repeated determination (extrapolated to 20°C) should not be greater than one order of magnitude.
How to handle out-of-tolerance? Perform a third determination, and use the arithmetic mean of the two results not greater than one order of magnitude, giving one significant figure.
What value is taken for the report? Take the arithmetic mean of two results of repeated determination.
Pressure measurement condition for the B method? Perform the measurement when the system pressure is lower than 4×10⁻⁴mmHg (requires confirmation against the source standard (source-text extraction)).
Photoelectric balance current range for the B method? 30~80μA (original text "30~80A", requires confirmation against the source standard (source-text extraction)).
Which old standard does it replace? Replaces SY 2672-77.
In what year was it first published? 1966 year.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.