1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Under specified conditions, the sample is oxidized, and the mass change of metal sheets in the sample before and after oxidation, the change in kinematic viscosity at 50℃, the n-pentane insolubles after oxidation, and the acidity/alkalinity of the sample vapor are scored, and the total score is used to represent the oxidation stability of the sample. The lower the total score, the better the oxidation stability.

The test conditions are divided into two types:

(Note: the original text "oxygen flow rate 20010mL/min" is source-text extraction noise; based on the context and clause 7.2 "200±10mL/min", it is determined to be 200±10 mL/min.)

4. Reagents and materials (purity / preparation)

4.1 Materials (listed in the source as 4.2 materials)

4.2 Reagents (listed in the source as Chapter 5)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative Determination and Calculation Formulas

The oxidation stability of the sample is calculated as the sum of the scores of six individual indicators. The scores of the six individual indicators are calculated according to their respective specified calculation standards.

10. Precision

Repeatability: For samples with a total score below 50 points, the difference between two results of repeated determinations by the same operator and their arithmetic mean shall not exceed 15% of their arithmetic mean. (The source wording for the results and comparison is unclear; verify this repeatability requirement against the original standard)

11. Report

Take the arithmetic mean of two results of repeated determinations as the determination result of the oxidation stability of internal combustion engine oils.

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
Standard number and replacement relationship? SH/T 0299-92, replaces SY 2681-82 (original text "SY 268182", requires confirmation against the source standard (source-text extraction))
Scope of application? Internal combustion engine oils
Method summary? Oxidize the sample under specified conditions, score by metal strip mass change, 50℃ kinematic viscosity change, n-pentane insolubles, and vapor acidity/alkalinity; the lower the total score, the better the oxidation stability
How many test conditions are there? Two: non-intensified (160℃, 200±10 mL/min, 6 h); intensified (165℃, 200±10 mL/min, 12 h, used for internal combustion engine oils with relatively good oxidation stability)
Sample amount? Inject into the oxidation tube up to the 40 mL mark
Metal strip specifications? Iron strip DT1: 76×12×0.5 mm; copper strip T1 (containing copper 99.95%): 76×16×0.5 mm; lead strip Pb1 or Pb2 (containing lead 99.99%): 76×12×0.5 mm; a small hole of diameter 3 mm is drilled in the center at 5 mm above
Which metal strip must use a new strip? Lead strip
Metal strip weighing precision? Accurate to 0.0002 g
Oxygen inlet tube outlet position? 3~4 mm from the bottom of the oxidation tube
Litmus paper check time? When oxidation proceeds to 5.5 h, place moist blue litmus paper at the tube mouth for about 30 s
n-Pentane insolubles test? Take 5 mL of the oxidized sample, add 45 mL n-pentane and stir evenly, centrifuge at 1500 r/min for 20 min, record the precipitate volume
n-Pentane insolubles scoring rule? Scoring is performed only when reaching and exceeding 0.1%; 0.1% counts as 5 points, and for each increase of 0.1%, 1 points are added
Vapor acidity/alkalinity scoring? If the litmus paper turns red, it is acidic and 3 points are counted; if it does not change color, it is alkaline and zero points are counted
Light transmittance measurement conditions? Dilute with petroleum ether at four times the sample mass, use petroleum ether as reference, 650 nm, 1 cm colorimetric cell
What method is used for 50℃ kinematic viscosity before/after oxidation? GB/T 265
Precision requirement? For samples with a total score below 50 points, the difference between two results of repeated determinations and the arithmetic mean shall not exceed 15% of the arithmetic mean
Reported value? Arithmetic mean of two results of repeated determinations
Drafting unit? Metals and Chemistry Research Institute, Railway Science Research Institute, Ministry of Railways
Year of first publication? 1978

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.