SH/T 0330-1992 Determination of mechanical impurities in lubricating grease (extraction method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and specimen preparation

8. Analysis procedure

9. Qualitative and quantitative determination and calculation formula

10. Precision

11. Report

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference Q&A

Question Key answer points
What does this standard apply to? It is applicable to the determination of the content of substances in lubricating grease that are insoluble in ethanol-benzene mixed liquid and hot distilled water.
What is the method principle? Extract the lubricating grease with ethanol-benzene mixed liquid, treat the precipitate on the filter with hot distilled water, and determine the mass of the insoluble residue.
What is the volume ratio of the ethanol-benzene mixed liquid? 1:40.
What is the specimen weighing amount? 1.5~2.0 g, to the nearest 0.0002 g.
What is the extraction rate requirement? The rate at which the condensate flows into the filter is 3~5 drops per second.
What is the extraction time requirement? At least 1.5 h, until the solution in the extractor changes from yellow to colorless.
What is the drying temperature? 105~110 ℃.
What is the difference for constant weight determination? The difference between two consecutive weighings is not greater than 0.0004 g.
What is the precision requirement? The difference between two results of repeated determinations shall not exceed 0.1%(m/m).
How are the results reported? Take the arithmetic mean of two results of repeated determinations.
Which old standard does it replace? Replaces SY 2709-62(82).