SH/T 0557-1993 Standard Test Method for Viscosity of Petroleum Asphalt (Vacuum Capillary Method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / supersession relationship

2. Applicable scope and non-applicability

3. Method principle

Under strictly controlled vacuum and temperature, measure the time required for a certain volume of liquid to flow through a capillary. The viscosity of the liquid is expressed in units of Pa·s and is the flow time in seconds multiplied by the viscometer calibration coefficient.

As the liquid in the capillary rises, the shear rate decreases; using different vacuum levels or capillaries of different diameters also changes the shear rate. Therefore, this method is suitable for determining the viscosity of both Newtonian and non-Newtonian fluids.

4. Reagents and materials (purity / preparation)

Table D1 Viscosity standard samples (excerpt from the source text)

Viscosity number Approximate viscosity, Pa·s (20℃) Approximate viscosity, Pa·s (38℃)
N30000 150 24
N190000 5 000 (requires confirmation against the source standard, source-text extraction) 160
(third row, number requires confirmation against the source standard) 24 not provided in the available source

Note: In Table D1, the 20℃ value of N190000 and the content of the third row source-text extraction have obvious noise and require confirmation against the source standard (source-text extraction).

5. Apparatus and equipment

Figure 1 Basic components of the vacuum system: 1— open mercury differential manometer; 2— regulating valve or stopcock valve; 3—1 liter buffer tank; 4— pressure stabilizer.

6. Recommended operating conditions

7. Sampling and sample preparation

4.1 Carefully heat the sample and stir continuously to ensure the sample is heated evenly and to prevent local overheating.

4.2 Transfer at least 20 mL of sample into a suitable container, heat to 135℃±5℃ (requires confirmation against the source standard, source-text extraction the source text reads "135`2 1 592"), stir occasionally to prevent local overheating, and stir carefully and slowly to prevent air from entering.

8. Analytical procedure

5.1 Maintain the water bath temperature within the range of the test temperature ±0.03℃, and make necessary corrections to all thermometer readings.

5.2 Select a clean, well-dried viscometer with a flow time greater than 60 s, and preheat to 135℃±5℃.

5.3 Fill the viscometer with the prepared sample to the height of the filling line E±2 mm (see Figure 2).

5.4 Place the viscometer filled with sample into an oven or oil bath at 135℃±5℃, and keep it at 10 min±2 min to expel large bubbles.

5.5 Remove the viscometer from the oven or oil bath, and within 5 min clamp the viscometer on a rack and place it vertically into the water bath so that the uppermost timing mark scale is at least 20 mm below the water bath liquid surface.

5.6 Establish a vacuum system of 39996 Pa±67 Pa (300 mmHg±0.5 mmHg), and connect the vacuum system to the viscometer using the regulating valve or stopcock valve installed on the viscometer line.

5.7 After the viscometer has been placed in the water bath for 30 min±5 min, open the regulating valve or stopcock valve on the line leading to the vacuum system so that the asphalt flows into the capillary.

5.8 Measure the time required for the leading edge of the meniscus to pass the two timing mark scales, to the nearest 0.1 s, report the first flow time between the two timing mark scales that exceeds 60 s, and note the numbers of this pair of timing mark scales.

5.9 After the test is completed, remove the viscometer from the water bath and place it upside down in an oven at 135℃±5℃ until the asphalt flows out of the viscometer. Thoroughly clean the viscometer with toluene, then rinse with petroleum ether. Pass a small flow of clean dry air through the capillary (for about 2 min or until the solvent has completely evaporated) to dry the tube. Periodically clean the viscometer with chromic acid cleaning solution to thoroughly remove organic deposits. Rinse with distilled water and residue-free acetone and blow dry with clean air.

Figure 2 Viscometer selected by this standard: 1— filling line E; 2— first timing mark F; 3—B section; 4— second timing mark G; 5—C section; 6— third timing mark H; 7—D section; 8— fourth timing mark I; 9— filling tube A; 10—24/40 ground joint; 11— vacuum tube M.

9. Qualitative and quantitative determination and calculation formula

Select the calibration coefficient corresponding to this pair of timing mark scales (e.g., 5.8), use the following formula to calculate and report the viscosity (Pa·s), retaining three significant figures:

P = k · t

Where:

(The source formula source-text extraction is "P two ‘ "c"; based on the context and the formula form in Annex D, it is arranged as P = k·t, requires confirmation against the source standard (source-text extraction).)

Annex D Calibration formula:

D3.7 The capillary calibration coefficient is calculated according to formula (D1):

k = P₁ / t

Where:

D4.3 The calibration coefficient k₁ is calculated according to formula (D2):

k₁ = (t₂ · k₂) / t₁

Where:

10. Precision

11. Report

When reporting the viscosity test results, the test temperature and vacuum shall also be stated. For example: viscosity [60℃, 39996 Pa±67 Pa, Pa·s] (the source example source-text extraction is "viscosity [printed ℃, 39996Pa±67Pa, Pa- s five 60t (300manHg, poise)]", requires confirmation against the source standard (source-text extraction)).

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
What does this standard measure? Dynamic viscosity of petroleum asphalt at 60℃, vacuum capillary method
Can other substances be measured? Other substances with viscosity in the range 0.0036~20.000 Pa·s can be measured
What is the vacuum degree? 39996 Pa±67 Pa (300 mmHg±0.5 mmHg)
Water bath temperature control requirement? Test temperature ±0.03℃
Sample heating temperature? 135℃±5℃
How long is the viscometer placed in the water bath? 30 min±5 min
Flow time requirement? Select greater than 60 s; report the first flow time exceeding 60 s
Calculation formula? P = k·t (k is the calibration coefficient, t is the flow time)
How many digits are retained in the result? Three significant figures
Repeatability limit? Not exceeding 7% of the average value
Reproducibility limit? Not exceeding 10% of the average value
What additional information is required in the report? Test temperature and vacuum degree
Which standard is identically adopted? ASTM D2171-85
How is the viscometer calibration coefficient obtained? Determined with a standard viscosity sample according to Appendix D, or by comparison with a standard vacuum viscometer
Is the calibration coefficient related to temperature? The calibration coefficient of the capillary is independent of temperature (D3.8)
What solvents are used for cleaning? Toluene, petroleum ether, chromic acid cleaning solution, distilled water, residue-free acetone

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.