SH/T 0560-1993 Test method for thermal stability of lubricating oils

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope and exclusions

3. Principle

At the temperature specified in the product standard, place two test cups, each containing 20g of sample, on the turntable in a lubricating-oil thermal-stability oven that has reached the specified temperature. Rotate at 5~6r/min. After the time specified in the product standard, visually examine the cooled samples and the bottoms of the test cups for sediment to assess thermal stability.

4. Reagents and materials (purity / preparation)

5. Apparatus

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation

10. Precision

Not provided in the available source.

11. Report

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick questions and answers

Question Key answer
Which oils does this standard cover? Mineral-oil-based and synthetic lubricating oils.
What is the sample amount? 20g±1g, weighed to the nearest 0.1g.
What is the turntable speed? 5~6r/min.
What is the temperature-control accuracy? Within ±1℃ of the specified test temperature.
When does timing start? When the sample temperature reaches 1℃ below the specified temperature.
What are the test-cup specifications? Internal diameter 51~55mm, external height 55~59mm, thickness 1.0~2.0mm; heat-resistant to no lower than 200℃; transparent flat-bottomed glass cup with a fire-finished level rim.
Where is the thermometer mercury bulb positioned? Its lower end is approximately 2~3mm above the cup bottom.
How is the result assessed? Report 'no sediment' if neither of the two parallel samples nor their cup bottoms show sediment; otherwise report 'sediment present'.
What if the results disagree? Repeat the procedure in 6.1~6.3 until the results agree.
What must be reported with the result? Test temperature and test duration.
Can samples requiring different test durations be tested simultaneously? No.
Which sample is used for temperature measurement when testing different samples simultaneously? The sample with the highest viscosity or density.
Can a test cup still containing deposits after cleaning be used? It must not be reused.
Which foreign standard was used as a reference? JIS K2540-1989, Test method for thermal stability of lubricating oils.
Which standard is cited? SH 0005 Solvent oil for the paint industry.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.