SH/T 0604-2000 Determination of density of crude petroleum and petroleum products (U-shaped oscillating tube method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

A small amount of sample (generally less than 1 mL) is injected into a temperature-controlled sample tube, the vibration frequency or period is recorded, and the sample density is calculated using the sample tube constant obtained in advance. The sample tube constant is determined from the vibration frequency when the sample tube is filled with a calibration liquid of known density (Chapter 4).

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

Judge the reliability of the results according to the following provisions (95% confidence level) (Chapter 14).

11. Report

Report the final density result to 0.1 kg/m³ or 0.0001 g/cm³ (Chapter 13).

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
What is the method number of this standard? SH/T 0604-2000
Which standard does it replace? SH/T 0604-1994
Which international standard is it equivalent to? ISO 12185:1996
Scope of application? Crude oil and petroleum products that can be handled as a single-phase liquid at the test temperature and pressure, with a density range of 600~1100 kg/m³
Can it be used for online density meter calibration? No
Method principle? A small amount of sample is injected into a temperature-controlled sample tube, the oscillation frequency or period is recorded, and the density is calculated using the sample tube constant
Sample amount? Generally less than 1 mL
Density meter resolution requirement? ±0.1 kg/m³ or higher
Constant-temperature bath temperature control accuracy? ±0.05℃
Temperature sensor accuracy? At least ±0.1℃
At least how many calibration liquids? Two
Calibration interval? If more than 7 days have passed since the last calibration, recalibrate before use
What can the standard temperature be? 20℃ or 15℃
Number of digits reported? 0.1 kg/m³ or 0.0001 g/cm³
Repeatability (transparent middle distillates)? 0.2 kg/m³
Repeatability (crude oil and other petroleum products)? 0.4 kg/m³
Reproducibility (transparent middle distillates)? 0.5 kg/m³
Reproducibility (crude oil and other petroleum products)? 1.5 kg/m³
Sampling according to which standard? GB/T 4756
Which standard is used to convert density to standard temperature? GB/T 1885 Petroleum Measurement Tables
In which appendix is the water density table? Appendix A (normative appendix)
In which appendix is the air density table? Appendix B (normative appendix)
Main differences from ISO 12185:1996? The definition of "wax precipitation temperature" and related content are not adopted; ISO Appendix A is changed to the main text; IP compilation Appendices G, H are used as Appendices A, B of this standard

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.