SH/T 0614-1995 Determination of composition of industrial propane and butane (gas chromatography)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

The sample is carried into the chromatographic column by the carrier gas, separated into the corresponding components within the column, detected by a thermal conductivity or flame ionization detector, and its chromatogram is recorded. The components are identified qualitatively using relative retention values, and the content of each component is calculated by the area normalization method.

4. Reagents and materials (purity / preparation)

4.1 Reagents

4.2 Materials

4.3 Preparation of stationary phase (7.1)

According to the liquid phase loading (see Table 1), first pour in the stationary liquid and add an appropriate solvent to dissolve it. Make the resulting solution volume slightly larger than the volume of the support, add the support while stirring, then gently stir and evaporate to remove the solvent.

5. Instruments and equipment

6. Recommended operating conditions

Table 1 Chromatographic column and operating conditions

Chromatographic column Dibutyl maleate + β,β′-oxydipropionitrile tandem column Dibutyl maleate column Sebaconitrile column
Stationary liquid Dibutyl maleate, β,β′-oxydipropionitrile Dibutyl maleate Sebaconitrile
Solvent n-Pentane or chloroform n-Pentane or chloroform Dichloromethane or toluene or diethyl ether
Support 6201 type 6201 type 6201 type
Support particle size, μm 180~250 180~250 180~250
Liquid phase loading, %(m/m) 25 25 20
Column tube material Stainless steel, copper, aluminum or glass tube Stainless steel, copper, aluminum or glass tube Stainless steel, copper, aluminum or glass tube
Column length, m 8 3 9 (the original source-text extraction is "9a7", requires confirmation against the source standard (source-text extraction))
Column inner diameter, mm 2~5 2~5 2~5
Column temperature, ℃ 20~40 20~40 20~40

Note: The column length, column inner diameter and column tube material can be selected by the operator when they can satisfy the resolution requirement of 7.4.

9.1 Test conditions

7.3 Aging of the chromatographic column

Before use, the chromatographic column should be purged with carrier gas for 5 h at a temperature slightly higher than the column temperature (below the maximum operating temperature of the stationary liquid).

7.4 Resolution of components

Between propane and propylene in industrial propane, and between propylene and isobutane in industrial butane, a resolution of formula (1) should be achieved. See Figure 1.

$$R = \frac{2(d'{A} - d'{B})}{W_A + W_B} \geq 1.5 \quad \cdots\cdots\cdots\cdots(1)$$

Where: A and B—— represent propane and propylene or propylene and isobutane components respectively; R—— the resolution of peak A and peak B; d'_A and d'_B—— are the adjusted retention times of components A and B respectively, expressed as distances on the chart, mm; W_A and W_B—— represent the peak widths of components A and B respectively.

Adoption notes:

  1. ISO 7941 the solvents are n-pentane, toluene or dichloromethane; this standard recommends the use of chloroform and diethyl ether.
  2. The length of the sebaconitrile column is increased by 3 m compared with ISO 7941.
  3. The column temperature of the sebaconitrile column, ISO 7941 specifies 20±1℃, this standard recommends the use of 20~40℃.

7. Sampling and sample preparation

8. Analysis procedure

9.2 Calibration

9.2.1 Qualitative analysis

The components can be identified by passing a standard reference mixture or several pure hydrocarbons through the column, or by comparing with the typical chromatograms of Figure 2, Figure 3, Figure 4 and the relative retention values shown in Table 2 (see 4.4.3), thereby identifying the components qualitatively.

Table 2 Relative retention values

Component Dibutyl maleate column Dibutyl maleate column + β,β′-oxydipropionitrile column Sebaconitrile column
Air + methane 0 0 0
Ethane 0.11 0.10 0.11
Ethylene 0.11 0.10 0.11
Propane 0.33 0.31 0.32
Propylene 0.42 0.41 0.52
Isobutane 0.68 0.66 0.64
n-Butane 1.00 1.00 1.00
1-butene 1.20 1.26 1.50
Isobutylene 1.20 1.26 1.61
trans-2--butene 1.55 1.59 1.95
cis-2--butene 1.77 1.83 2.31
1,3--butadiene 1.96 — 3.17
Isopentane 2.21 — 2.19
n-Pentane 2.86 — 2.83

Note on adoption: This retention value was obtained from a large number of tests and is consistent with the typical chromatogram of ISO 7941. The retention values in ISO 7941 are not consistent with their own typical chromatograms.

9.2.2 Quantitative analysis

The calculation method is the normalization method. Table 3 gives the peak area correction factors used. These values are only a guide or approximation; laboratories with experience in preparing calibration gas mixtures should determine the correction factors themselves.

9.3 Determination

9.3.1 Liquid injection

Place the sampling bottle or gas cylinder upright so that the outlet valve is at the bottom. Using a non-plasticized transparent armored or pressure-resistant tube (the tube should be "grounded"), connect the valve to the injector through a metal filter (6.6) (see Figure 5). Install a pressure reducing valve downstream of the injection device vent system to avoid any vaporization upstream when the flow rate reaches equilibrium. Open the outlet valve and control the flow through the transparent tube so that the liquid completely fills the transparent tube. Inject the sample into the column and close the outlet valve.

9.3.2 Gas injection

9.4 Chromatogram

9.4.1 Typical chromatogram

Figure 2, Figure 3, and Figure 4 represent typical chromatograms of industrial propane, butane, and a reference mixture containing liquefied petroleum gas components.

9.4.2 Qualitative analysis

For industrial propane, use a dibutyl maleate + β,β′-oxydipropionitrile column. For industrial butane, use a dibutyl maleate column. When determining both propane and butane, use a sebaconitrile column.

9.4.3 Quantitative analysis

9. Qualitative and quantitative analysis and calculation formulas

10 Expression of results

Calculate the concentration of each component in the sample using the following formula.

Thermal conductivity detector

The mass percentage X_i %(m/m) of component i in the sample is calculated according to formula (2):

$$X_i = \frac{f_i \cdot A_i}{\sum (f_i \cdot A_i)} \times 100 \quad \cdots\cdots\cdots\cdots(2)$$

Where: A_i—— peak area of component i; f_i—— mass correction factor of component i, see Table 3; n—— number of components in the mixture.

The mole percentage X_i %(mol) of component i in the sample is calculated according to formula (3):

$$X_i = \frac{f'_i \cdot A_i}{\sum (f'_i \cdot A_i)} \times 100 \quad \cdots\cdots\cdots\cdots(3)$$

Where: A_i—— peak area of component i; f'_i—— mole correction factor of component i, see Table 3; n—— number of components in the mixture.

The calculation results are rounded to two decimal places.

Flame ionization detector

The mass percentage X_i %(m/m) of component i in the sample is calculated according to formula (4):

$$X_i = \frac{f_i \cdot A_i}{\sum (f_i \cdot A_i)} \times 100 \quad \cdots\cdots\cdots\cdots(4)$$

Where: A_i—— peak area of component i; f_i—— mass correction factor of component i, see Table 3; n—— number of components in the mixture.

The mole percentage X_i %(mol) of component i in the sample is calculated according to formula (5):

$$X_i = \frac{A_i / n_{Ci}}{\sum (A_i / n_{Ci})} \times 100 \quad \cdots\cdots\cdots\cdots(5)$$

Where: A_i—— peak area of component i; n_{Ci}——effective carbon number in component i; n—— number of components in the mixture.

The calculation results are rounded to two decimal places.

Note on adoption: The formula in ISO 7941 is incorrect; upon consulting the literature, this formula should add ×100.

Table 3 Relative correction factors of components

Component Thermal conductivity detector - mass correction factor (carrier gas hydrogen) Thermal conductivity detector - mass correction factor (carrier gas helium) Thermal conductivity detector - mole correction factor (carrier gas hydrogen) Thermal conductivity detector - mole correction factor (carrier gas helium) Flame ionization detector - mass correction factor (carrier gas hydrogen) Flame ionization detector - mass correction factor (carrier gas helium)
Methane 0.56 0.65 2.03 2.37 1.11 0.99
Ethane 0.74 0.86 1.44 1.66 1.03 0.99
Ethylene 0.74 0.84 1.52 1.74 0.97 0.99
Propane 0.89 0.97 1.17 1.28 1.01 0.93
Propylene 0.90 0.94 1.24 1.29 0.97 0.99
Isobutane 1.03 1.02 1.03 1.02 1.00 0.99
n-Butane 1.00 1.00 1.00 1.00 1.00 0.99
1--Butene 1.00 1.00 1.03 1.03 0.97 0.99
Isobutylene 1.01 1.00 1.04 1.04 0.97 0.99
trans-2--Butene 0.99 0.96 1.02 1.00 0.97 0.99
cis-2--Butene 0.99 0.94 1.02 0.98 0.97 0.99
1,3--Butadiene 1.01 0.99 1.08 1.07 0.93 0.99
Isopentane 1.14 1.05 0.92 0.85 0.99 0.99
n-Pentane 1.10 1.01 0.89 0.82 0.99 0.99

(Note: Some values in Table 3 are misaligned due to source-text extraction typesetting; the above has been organized according to row-column correspondence and requires confirmation against the source standard (source-text extraction))

10. Precision

Judge the reliability of test results according to the following provisions (95% confidence level).

11.1 Repeatability

For the same operator, using the same instrument, on the same sample under the same operating conditions, with normal and correct operating methods, the difference between two results obtained from repeated determinations shall not exceed the value shown in Table 4.

11.2 Reproducibility

For different operators, using different instruments, on the same sample under the same operating conditions, with normal and correct operating methods, the difference between the two results obtained by each shall not exceed the values shown in Table 4.

Table 4 Repeatability and Reproducibility

Operating conditions Product and components Repeatability Reproducibility
Gas injection, with flame ionization or thermal conductivity detector Industrial propane, all components 0.25 1
Gas injection, with flame ionization or thermal conductivity detector Industrial butane, all components 0.25 2
Liquid injection, with thermal conductivity detector Industrial propane component concentration: >0.1% and <1% 0.05 0.20
Liquid injection, with thermal conductivity detector Industrial propane component concentration: ≥1% and <5% 0.20 0.50
Liquid injection, with thermal conductivity detector Industrial propane component concentration: ≥5% 0.5 1
Liquid injection, with thermal conductivity detector Industrial butane component concentration: <25% not provided in the available source (source-text extraction missing) not provided in the available source (source-text extraction missing)
Liquid injection, with thermal conductivity detector Industrial butane component concentration: ≥25% 1 1.5

Note: Component concentration ranges are expressed in mass percent; precision data are expressed in mass percent (absolute value).

11. Report

12 Test report: The report shall include the following contents:

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A Reference

Question Key points of answer
What international standard does this standard adopt? Equivalent adoption of ISO 7911:1988 "Analysis of commercial propane and butane (gas chromatography method)".
What is the scope of application? Determination of industrial propane, butane and their hydrocarbon components, excluding hydrocarbon components with concentrations below 0.1%(m/m); also applicable to various hydrocarbon components and mixtures in liquefied petroleum gas.
What is it not applicable to? Not applicable to "online" chromatography.
What is the principle of the method? The sample is carried by carrier gas into the chromatographic column for separation, detected by thermal conductivity or flame ionization detector, qualified by relative retention value, and quantified by area normalization method.
How many chromatographic columns are there? Three: dibutyl maleate + β,β′-oxydipropionitrile tandem column, dibutyl maleate column, sebaconitrile column.
Which column is used for measuring industrial propane? Dibutyl maleate + β,β′-oxydipropionitrile column.
Which column is used for measuring industrial butane? Dibutyl maleate column.
Which column is used for measuring both propane and butane? Sebaconitrile column.
What is the resolution requirement? The resolution between propane and propylene, and between propylene and isobutane is R ≥ 1.5.
Which components cannot be separated? Air and methane, ethane and ethylene, 1- butene and isobutylene (under recommended conditions).
What is the quantification method? Area normalization method.
Formula for mass percent with thermal conductivity detector? X_i = f_i·A_i / Σ(f_i·A_i) × 100, see formula (2).
Formula for mole percent with flame ionization? X_i = (A_i/n_Ci) / Σ(A_i/n_Ci) × 100, see formula (5).
How many decimal places are retained in the results? To two decimal places.
What standard is sampling based on? GB/T 13290.
Chromatographic column aging conditions? Slightly above column temperature (below the maximum use temperature of the stationary liquid) with carrier gas flow 5 h.
Recommended carrier gas flow rate? 30 mL/min or the flow rate required to obtain good resolution.
Vaporization chamber temperature? 40±5℃ or a suitable temperature selected by the operator.
Detector temperature range? Thermistor thermal conductivity 40~50℃; hot wire thermal conductivity 100~150℃; flame ionization 100~150℃.
Repeatability (gas injection, all components)? Industrial propane 0.25, industrial butane 0.25.
Reproducibility (gas injection, all components)? Industrial propane 1, industrial butane 2.
What should the report include? All information about the test sample (batch number, date, sampling location, etc.), test results (to two decimal places), abnormal phenomena.