SH/T 0629-1996 Determination of arsenic content in naphtha (potassium borohydride-silver nitrate spectrophotometric method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Arsenic in the sample is extracted with sulfuric acid and hydrogen peroxide. The extract is heated to destroy organic matter. In a tartaric acid medium, potassium borohydride is used to generate arsine. Impurities in the arsine are removed by purification. The arsine is absorbed with a silver nitrate-polyvinyl alcohol-ethanol solution to form a yellow silver sol. The absorbance is measured at 410nm.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

Table 1 Sample amount table

Arsenic content µg/kg Sampling amount mL Separatory funnel volume mL Absorption solution volume mL
1~5 50 1000 3
>5~20 300 500 5
>20~60 100 250 5
>60~200 30.0 125 5
>200 10.0 No extraction 5

Note: If the arsenic content is greater than 200 µg/kg, when taking a sample less than or equal to 10mL, extraction may be omitted; directly take the sample into a 100mL Erlenmeyer flask, add 30% hydrogen peroxide 10mL and sulfuric acid solution 30mL, mix thoroughly with the sample, place on an electric furnace and heat, then proceed as in 8.3.

(In Table 1, “[unreadable value]- 21[unreadable characters]”, “2([unreadable characters])”, “,10[unreadable character]” etc. are source-text extraction noise, inferred from context, requires confirmation against the source standard (source-text extraction))

8. Analysis steps

9. Qualitative and Quantitative and Calculation Formula

The arsenic content in the sample As (µg/kg) is calculated according to formula (1):

As = X / (V × ρ) ………………(1)

Where:

(The original formula layout is “[garbled symbols] X [unreadable symbol] X - Vxp”, suspected to be source-text extraction noise, arranged according to context, requires confirmation against the source standard (source-text extraction))

10. Precision

Judge the reliability of the results according to the following provisions (95% confidence level).

Table 2 Precision (µg/kg)

Arsenic content Repeatability Reproducibility
1~5 1.0 2.5
>5~20 2.0 5.0
>20 0.1X 0.25X

Note: X is the arithmetic mean of the two results.

(In Table 2, “O.IX” is arranged according to context as 0.1X, requires confirmation against the source standard (source-text extraction))

11. Report

Take the arithmetic mean of two repeated measurement results as the determination result, retaining three significant figures.

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
What petroleum products does this standard apply to? Naphtha, and it also applies to reforming feedstock oil.
What is the applicable arsenic content range? The original text is 1~5501 µg/kg; the values are suspected to be source-text extraction noise, requires confirmation against the source standard (source-text extraction).
What is the principle of the method? Sulfuric acid-hydrogen peroxide extraction, heating to destroy organic matter, generation of arsine from potassium borohydride in a tartaric acid medium, after purification absorption with silver nitrate-polyvinyl alcohol-ethanol solution to form a yellow silver sol, 410nm measure absorbance.
How is the color-developing absorption liquid prepared? Silver nitrate solution, polyvinyl alcohol solution, and anhydrous ethanol are mixed in a volume ratio of 1:1:2, and can be stable for 3h (6.7).
Measurement wavelength and cuvette? 410nm, 1cm cuvette.
Within how long after sampling should analysis be performed? Within 24h.
According to what standard is sample density measured? GB/T 1884.
According to what standard is sampling with other containers performed? GB/T 4756.
Can high-arsenic samples be not extracted? When the arsenic content is greater than 200 µg/kg and the sample amount is ≤10mL, extraction may be omitted and digestion performed directly (Note to Table 1).
What should be noted during digestion? Do not evaporate to dryness, so as to avoid volatilization loss of arsenic; the test shall be carried out in a fume hood.
What is the calculation formula? As = X / (V × ρ), X is the arsenic content corresponding to the working curve (µg), V is the sample volume (mL), and ρ is the density at the test temperature (g/mL).
How many significant figures are retained in the result? Three.
Precision conditions? 95% confidence level; repeatability and reproducibility according to Table 2.
Who provides the arsine generation apparatus? It can be produced and provided by Fushun Petroleum Institute, and supervised by the petroleum products standardization technical centralized unit.
Highly toxic precautions? Arsenic trioxide is a highly toxic substance; arsine gas is a highly toxic substance, and the reaction shall be carried out in a fume hood or a well-ventilated place; chromic acid cleaning solution is a strong oxidant, hygroscopic agent, and toxic substance, and protective measures shall be taken during use.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.