SH/T 0631-1996 Determination of barium, calcium, phosphorus, sulfur and zinc in lubricating oils and additives (X-ray fluorescence spectrometry)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

4.1 Reagents (6.1)

4.2 Materials (6.2)

5. Instruments and equipment

6. Recommended operating conditions

Element Wavelength, nm Analyzing crystal Peak angle, 2θ Background angle, 2θ Detector
Barium 0.278 LiF200 87.13 85.70 FS
Calcium 0.355 LiF200 113.10 114.50 F
Phosphorus 0.615 Germanium 140.92 142.90 F
Sulfur 0.537 Graphite 106.22 108.00 F
Zinc 0.143 LiF200 41.79 43.60 FS

7. Sampling and sample preparation

Standard sample number Barium Calcium Phosphorus Sulfur Zinc
1 1.0 0.1 0.8 0.1 0.5
2 1.0 0.8 0.8 0.4 1.0
3 1.0 0.5 0.8 0.25 2.0
4 0.1 0.1 1.8 1.0 1.0
5 0.8 0.8 0.1 0.5 requires confirmation against the source standard (source-text extraction)
Standard sample number Barium Calcium Phosphorus Sulfur Zinc
7 0.1 0.1 0.6 1.8 0.6
8 0.6 0.1 0.1 1.8 0.4
9 1.0 0.1 0.3 1.7 0.1
10 0.1 0.9 0.1 1.3 0.7
11 0.6 0.3 0.5 0.8 0.1
12 0.4 0.8 0.1 1.8 0.1
13 0.3 0.7 0.7 1.6 1.0
14 0.2 1.0 0.1 0.1 0.7
15 0.8 0.4 0.2 1.3 0.1
16 1.0 1.0 0.6 0.1 0.2
17 1.0 1.0 0.1 0.1 0
18 0.6 0.5 0.1 1.4 0
19 0.1 0.1 0.1 0.6 0
20 0.06 0.2 0.4 1.0 0.5
21 0.8 0 0.1 0.1 0.5
22 ◎ (requires confirmation against the source standard (source-text extraction)) 0 0.8 0 0
23 0 0 0 0.5 0.5
24 0 0.6 0 0 0.5
25 1.0 0.8 0.5 0 0.25
26 1.0 0.8 0.5 0.1 0.5
27 1.0 0 0 0.1 0
28 0.5 0.4 0.5 1.5 0
29 1.0 0.8 0.25 0.75 requires confirmation against the source standard (source-text extraction)
30 0 0 0.5 1.5 requires confirmation against the source standard (source-text extraction)
31 0 requires confirmation against the source standard (source-text extraction) 0.5 0.75 requires confirmation against the source standard (source-text extraction)
32 0 requires confirmation against the source standard (source-text extraction) 0 requires confirmation against the source standard (source-text extraction) requires confirmation against the source standard (source-text extraction)

8. Analysis Procedure

9. Qualitative and Quantitative Calculation Formulas

10. Precision

Element Repeatability r Reproducibility R
Barium 0.020X13 (requires confirmation against the source standard (source-text extraction)) 0.138×0.93 (requires confirmation against the source standard (source-text extraction))
Calcium 0.023 8X9.86 (requires confirmation against the source standard (source-text extraction)) 0.113 6X0.85 (requires confirmation against the source standard (source-text extraction))
Phosphorus 0.034 8X9.92 (requires confirmation against the source standard (source-text extraction)) 0.064 2X0.56 (requires confirmation against the source standard (source-text extraction))
Sulfur X=0.1 (requires confirmation against the source standard (source-text extraction)) 0.155 9 (X+0.021 4)
Zinc X=0.1 (requires confirmation against the source standard (source-text extraction)) 0.116 5X

11. Report

not provided in the available source.

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key Answer Points
Which elements does this standard determine? Barium, calcium, phosphorus, sulfur, zinc
Applicable concentration range? 0.03%~1.0% (m/m); sulfur is 0.01%~2.0% (m/m)
Which method of ASTM D4927-93 is adopted? B method (mathematical correction method)
Main differences between this standard and the ASTM B method? Standard samples: 32 (ASTM 20); domestic substitution of reagents; downward irradiation fills half a cup, upward irradiation fills a full cup
Core of the method principle? Measure net intensity (analyte line minus background), use regression analysis to obtain inter-element correction coefficients, and calculate concentration according to formula (1)
Counting time requirement? The counting time for the analyte line at each wavelength position shall be at least 60s
Sample loading requirement? For a downward-irradiation spectrometer, fill half a cup; for an upward-irradiation spectrometer, fill a full cup
Dilution solvent requirement? Does not contain metals or phosphorus, sulfur content lower than 0.01% (m/m), such as kerosene, white oil, trimethylbenzene, etc.
What to do if the upper concentration limit is exceeded? When barium, calcium, phosphorus, zinc >1% (m/m) or sulfur >2% (m/m), dilute and repeat 9.1~9.3
Precision confidence level? 95%
Is bias determined? Not determined, because there is no suitable reference material with known composition
Issue/implementation date? Issued on 1996-05-24, implemented on 1996-12-01

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.