1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

9. Qualitative and quantitative determination and calculation formulas

10. Precision

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A

Question Key answer points
Standard number and year SH/T 0703—2001
Card type Test method
Test temperature 150℃±0.1℃
Tube wall apparent shear rate 1.4×10⁶ s⁻¹ (the original text source-text extraction is 1.4x108s-1, requires confirmation against the source standard (source-text extraction))
Method principle Under nitrogen (or carbon dioxide) pressurization, the test sample flows out through a capillary; the apparent viscosity is obtained from the efflux time and pressure via a calibration curve
Applicable objects High-temperature high-shear (HTHS) apparent viscosity of engine oils
Equivalent adoption ASTM D5481—96
Differences from ASTM Added relevant normative references; added instrument cleaning agent requirements
Viscometer cell specifications Capillary diameter 0.15 mm; length 15 mm–18 mm; pressure 350 kPa–3500 kPa; pressure control ±1%
Calibration oils HT39, HT75, HT150, FT240, HT390, 150℃ approximate viscosity 2.0~7.0 mPa·s
Viscosity range of Newtonian oils for calibration At 150℃, 2 mPa·s–5 mPa·s (9.1.2); for Annex B, use 2.5–4.0 mPa·s
Sampling standard GB/T 4756
Impurity filtration 10 μm filter paper (the original text source-text extraction is 10ym, requires confirmation against the source standard (source-text extraction))
Constant temperature time Viscometer at least 30 min; test sample 150℃ maintained for 15 min
Repeatability Arithmetic mean × 1.6%
Reproducibility Arithmetic mean × 5.4%
Reporting unit and number of decimal places mPa·s, retain two decimal places, and report the test temperature
Bias This test method has no bias
Cleaning agents Toluene or other high-aromatic solvents; lubricating base oils with kinematic viscosity at 40℃ less than 13 mm²/s, or transformer oil conforming to GB 2536
Cleaning temperature limit When cleaning with toluene or high-aromatic solvents, the viscometer cell temperature must not exceed 100℃
Temperature verification frequency At least once per year
Calibration curve stability check frequency Before each new round of tests or once every twenty tests; for non-Newtonian calibration oils, at least once per month
Annexes Annex A (determination of viscometer cell volume), Annex B (calculation of formula coefficients), both are normative annexes

Related public products