SH/T 0705-2001 Standard test method for determination of vanadium in heavy oils by spectrophotometry

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Source limitations: The source has unresolved oxidation-state notation, standard-solution concentration, volume symbols and precision coefficients. These uncertainties are retained below and summarized in section 13. Check the original standard before using the values or calculation.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope and exclusions

3. Principle

Prepare a sample solution containing V³⁺ and VO²⁺ by the treatment procedure. Pass a measured portion through a cation-exchange resin so that V³⁺, VO²⁺ and certain other cations are retained. Elute the column with acidified hydrogen peroxide to oxidize the species to V⁵⁺, which leaves the cation-exchange column as VO₃⁻. The source card lists the species being oxidized as V(Ⅳ), V(Ⅳ), with the extracted wording 'VII, V4+'; this unresolved oxidation-state wording requires verification against the original standard. Collect and treat the eluate, then react its V⁵⁺ with a mixed tannic acid–thioglycolic acid solution to form an indigo-blue complex. Measure at a wavelength of 600 nm.

4. Reagents and materials (purity / preparation)

Unless otherwise stated, all reagents are analytical grade and the water meets the grade-two requirements of GB/T 6682.

6.1 Preparation of solutions

6.2 Treatment of the cation-exchange resin

Sieve the cation-exchange resin to obtain the 60~80 mesh fraction. Place it in 20% (V/V) hydrochloric acid solution and soak for 1 h. Wash with distilled water until neutral and until a test with silver nitrate solution shows no chloride ions. Measure 40 mL of the activated hydrogen-form resin and pack it into the cation-exchange column for use.

6.3 Regeneration of the cation-exchange column

Soak the used cation-exchange column with 40 mL of 1:2 (V/V) hydrochloric acid solution for 1 h, then rinse with distilled water until neutral. Test with silver nitrate solution for absence of chloride ions. Shake the resin column to loosen the packing for the next use.

5. Apparatus

6. Recommended operating conditions

7. Sampling and sample preparation

8. Procedure

7.1 Preparation of the calibration curve

Into 6 separate 25 mL volumetric flasks, add respectively 0, 0.5, 1.0, 1.5, 2.0 and 2.5 mL of the 10 μg/mL vanadium standard solution. Add a small piece of Congo red test paper and adjust with aqueous ammonia until the paper turns red. Then add, in sequence, 10 mL of acetic acid–ammonium acetate buffer solution and 2.5 mL of mixed tannic acid–thioglycolic acid solution. Dilute to the mark with distilled water and mix. After standing for 25 min, measure the absorbance of these standard solutions at 600 nm using a spectrophotometer with a 3 cm optical-path cell and the blank solution as reference. Plot vanadium concentration of the standard solutions on the horizontal axis and absorbance on the vertical axis.

7.3 Measurement of the sample solution

Pipette a measured volume of sample solution through the cation-exchange column at 2 mL/min~3 mL/min. When vanadium content does not exceed 1 mg/kg, pass the entire 25 mL sample solution through the column. Then pass 35 mL of water through the column to flush out all the sample solution and discard all this effluent. Next elute the column successively with 50 mL of hydrogen peroxide eluent and 25 mL of distilled water. Collect the eluate in a 100 mL beaker and add 1 mL of 1:1 (V/V) sulfuric acid solution. Concentrate on the electric heater until sulfur trioxide appears as white fumes. After cooling, transfer the solution to a 25 mL volumetric flask. Rinse the beaker three times with distilled water, using a total volume not exceeding 10 mL, and add the rinsings to the flask. Add a small piece of Congo red test paper and adjust with aqueous ammonia until the paper turns red. Add, in sequence, 10 mL of acetic acid–ammonium acetate buffer solution and 2.5 mL of mixed tannic acid–thioglycolic acid solution. Dilute to the mark with distilled water, mix and leave for 25 min. Measure absorbance at 600 nm on a spectrophotometer using a 3 cm optical-path cell and the blank solution as reference. Read the vanadium content of the sample solution from the calibration curve.

Notes

  1. If the sample vanadium content exceeds 5.00 mg/kg, increase the sample dilution factor for the determination.
  2. Iron (Fe³⁺), aluminium (Al³⁺) and cobalt (Co²⁺) ions do not interfere when their contents are no greater than 75, 50 and 50 times the vanadium content, respectively. Otherwise, a longer cation-exchange column may be used for treatment.

9. Determination and calculation

Calculate the sample vanadium content X (mg/kg) using equation (1):

X = (c × 25 × 25) / (m × V₁)

Where:

Note: the extracted formula appears as 'cx [unreadable character]x V / M X V₁' and 'cx25x25 / M X V₁'. The source card uses c×25×25 as the numerator and m×V₁ as the denominator, but the formula remains subject to verification against the original standard.

10. Precision

Assess the reliability of test results using the following criteria (95% confidence level).

9.1 Repeatability (r)

The difference between two results obtained by the same operator, on the same instrument and for the same sample shall not exceed the value specified in Table 1.

Table 1 Repeatability (r)

Vanadium content, mg/kg Repeatability (r), mg/kg
X < 1 0.104X (the source extraction reads '104 X'; verification pending)
1 < X < 5 0.076X + 0.213 (the source extraction reads '0.076X 0,213'; verification pending)

Reproducibility (R)

The difference between two results obtained by different operators in different laboratories, using different instruments for the same sample, shall not exceed the value specified in Table 2.

Table 2 Reproducibility (R)

Vanadium content, mg/kg Reproducibility (R), mg/kg
X < 1 0.202X (the source extraction reads '0.202X 0.}'; verification pending)
1 < X < 5 0.123X (the source extraction reads '0 .123 x'; verification pending)

11. Report

Report the arithmetic mean of two repeat determinations as the sample vanadium content, to two decimal places.

12. References

13. Source limitations and verification

14. Quick reference questions

Question Key points
Which oils are covered? Heavy oils and crude oils; heavy oils include atmospheric residues, vacuum residues and wax-oil fractions (1.1 and note).
Minimum detection limit? 0.06 mg/kg (1.2).
Measurement wavelength? 600 nm (3, 7.1, 7.3).
Absorption-cell optical path length? 3 cm (4.10, 7.1, 7.3).
Colour reagent? Mixed tannic acid–thioglycolic acid solution (3, 6.1.4).
Colour of the complex? Indigo blue (3).
Cation-exchange resin grade? 001×7 (GB/T 1631) (5.13).
Resin particle size? 60~80 mesh (4.12, 6.2).
Ashing temperature? 550℃±25℃ (4.3, 7.2).
Ashing time? Generally 8 h~12 h or longer, depending on the oil sample (7.2).
Flow rate through the exchange column? 2 mL/min~3 mL/min (7.3).
Standing time after adding the colour reagent? 25 min (7.1, 7.3).
Vanadium standard-solution concentrations? 0.1 mg/mL; after dilution, 10 μg/mL (6.1.7; source verification pending).
Calculation formula? X = (c × 25 × 25) / (m × V₁) (section 9; source verification pending).
Reporting decimal places? Two decimal places (section 11).
Interfering ions and permitted ratios? Fe³⁺ ≤ 75 times vanadium content; Al³⁺ ≤ 50 times vanadium content; Co²⁺ ≤ 50 times vanadium content (section 7, note 2).
Handling high-content samples? If vanadium content exceeds 5.00 mg/kg, increase the sample dilution factor (section 7, note 1).
Referenced standards? GB/T 1631 and GB/T 6682 (section 12).