Origin Standard

SH/T 0751-2005 Determination of viscosity at high temperature and high shear rate (tapered plug viscometer method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative and calculation formulas

r = [M·Σxiyi − (Σxi)(Σyi)] / { √[M·Σxi² − (Σxi)²] · √[M·Σyi² − (Σyi)²] } …………(A-1)

Where:

This formula has been stored in the computer connected to the instrument. (The original formula source-text extraction has noise; the symbols are subject to the above arrangement, requires confirmation against the source standard (source-text extraction))

10. Precision

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
What are the method number and name of this standard? SH/T 0751-2005 Determination of viscosity at high temperature and high shear rate (tapered plug viscometer method)
Which foreign standard is adopted? Modified adoption of ASTM D4741-00
What oils are applicable? Internal combustion engine oils (1.4)
Test temperature and shear rate? 150℃ and 1×10⁶ s⁻¹; 100℃ and 1×10⁶ s⁻¹ (1.1)
What model of instrument is used? BE/C (single-speed) or BS/C (multi-speed) high shear rate tapered plug viscometer (1.1, 6.2)
Can BE/C perform 100℃? No, due to torque limitation range (6.1 note)
What calibration oils are there? RL102, RL103, RL104, RL105, RL106, RL107 (7.1)
What is the check oil? Non-Newtonian check oil RL174 (7.2)
Calibration oil viscosity range? 150℃: about 1.8~5.9 mPa·s; 100℃: about 4.2~18.9 mPa·s (1.2, 10.1.2)
Calibration curve requirements? 6 Newtonian calibration oils, linear regression, straight line through the origin; correlation coefficient ≥0.99970 (10.1.4, 10.1.5)
Check oil bias requirements? 150℃: ±0.04 mPa·s; 100℃: ±0.06 mPa·s; subsequent ±0.06 mPa·s and same direction (10.1.7)
After how many samples must a check be performed? After every 3~6 test samples, verify with check oil; after every 6 samples tested, a check must be performed (10.1.6, 11.3)
Repeat test requirements? Repeat at least twice; if the difference between the two exceeds 1%, a third or even fourth test is required (10.1.11)
Sample injection amount? When viscosities are completely different, four full funnels; otherwise two full funnels; for repeatability tests, one full funnel; one full funnel is about 10 mL (10.2.3, 10.2.4)
Reporting unit and number of digits? mPa·s, accurate to two decimal places, rounded (12.1, 12.2)
Repeatability (r)? 150℃: ≤ average value 1.7%; 100℃: ≤ average value 1.0% (14.1.1)
Reproducibility (R)? 150℃: ≤ average value 2.6%; 100℃: ≤ average value 2.4% (14.1.2)
Precision effective viscosity range? 150℃: about 2.6~4.7 mPa·s; 100℃: about 4.9~11.8 mPa·s (14.1.1, 14.1.2)
Bias? Less than 0.06 mPa·s (14.4)
Is there precision for used oils? No precision specification (8.2)
Filtration precautions? Maximum particles >3 μm need filtration; filters smaller than 1 μm cannot be used; for used oils, pressure filtration rather than vacuum filtration is recommended (8.1, 8.2.1)
What are the normative references? GB 1922, SH/T 0024, ASTM D4683 (Chapter 2)