SH/T 0771-2005 Petroleum products — Determination of pour point (automatic pressure pulse method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formula

10. Precision

Instrument model 1 ℃ interval 3 ℃ interval
70Y model (original text source-text extraction as "7OY model") 1.6 2.2
30 model, 50 model, 70 model 2.9 4.1
Instrument model 1 ℃ interval 3 ℃ interval
70Y model (original text source-text extraction as "7OV model") 3.2 3.8
30 model, 50 model, 70 model 6.2 6.3

11. Report

12. Reference documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference for questions and answers

Question Key points of answer
What is the method number and name of this standard? SH/T 0771—2005 Standard test method for pour point of petroleum products (automatic pressure pulsing method)
Which foreign standard is adopted? Modified adoption of ASTM D5949-01, redrafted based on it
Applicable temperature range? -57 ℃ to 51 ℃ (original source-text extraction unclear in the source text, requires confirmation against the source standard)
Is it applicable to crude oil? Not applicable (1.4)
Can 1 ℃ or 3 ℃ intervals be used? Yes (1.3, 5.5, 11.7)
What is the cooling rate? 1.5 ℃/min ± 0.1 ℃/min (4, 6.4, 11.8)
What is the sample amount? 0.150 mL ± 0.005 mL (7.2.2, 11.2, 11.4)
Sample heating conditions? When the expected pour point is higher than 36 ℃ or the sample is solid at room temperature, heat to 45 ℃ to 70 ℃; the sample may be heated or transferred into the sample cup only when the sample temperature is below 70 ℃ (8.2, 11.2)
What gas is used for the pulsed nitrogen? Dry nitrogen, with a dew point lower than the lowest temperature the sample can reach (7.2.1)
Pressure pulse unit dimensions? Length 250 mm ± 2 mm, inner diameter 1.1 mm ± 0.1 mm stainless steel tube, outlet distance from sample center 8 mm ± 2 mm (6.3)
Temperature sensor accuracy? Reading 0.1 ℃, minimum accuracy 0.2 ℃, below the bottom surface of the sample cup by 0.1 mm (A.1.2)
Light source wavelength? 660 nm ± 10 nm (A.1.4)
How is the pour point determined? When pulsed nitrogen is applied, if no movement of the sample surface is observed, that is the no-flow point; the last temperature at which movement can be observed is the pour point (3.2, 11.8)
Repeatability values? See Table 1: Type 70Y 1 ℃ interval 1.6 ℃, 3 ℃ interval 2.2 ℃; Type 30/50/70 1 ℃ interval 2.9 ℃, 3 ℃ interval 4.1 ℃
Reproducibility values? See Table 2: Type 70Y 1 ℃ interval 3.2 ℃, 3 ℃ interval 3.8 ℃; Type 30/50/70 1 ℃ interval 6.2 ℃, 3 ℃ interval 6.3 ℃
Is bias determined? Bias not determined (13.2)
Difference between results at 1 ℃ and 3 ℃ intervals? Average bias 1.1 ℃, results at 1 ℃ intervals are on average lower than those at 3 ℃ intervals by 1.1 ℃ (13.3.2)
Sampling basis? GB/T 4756 (8.1)
What is reported? Report the temperature in 11.8 as the pour point (12)
Nature of Appendix A? Normative appendix, detailed description of the apparatus