SH/T 0797-2007 Determination of benzene content in spark-ignition engine fuels — Mid-infrared spectroscopy

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

9. Qualitative and Quantitative and Calculation Formulas

10. Precision

11. Report

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference for questions and answers

Question Key points of answer
What is the method number and name of this standard? SH/T 0797—2007 Determination of benzene content in spark-ignition engine fuels — Mid-infrared spectroscopy
Which foreign standard is adopted/modified adopted? Modified adoption of ASTM D6277-01, redrafted based on it
Applicable benzene content range? 0.1% (volume fraction) ~ 5% (volume fraction)
What is the principle of the method? The sample is placed in a liquid sample cell, an infrared beam passes through the sample, and the detector measures the response; the spectral response is measured in the benzene measurement band range, and a multivariate calibration model is used to convert the response to benzene content
What are the main interferences? Toluene and other mono-substituted aromatic hydrocarbons
What instrument is used? Fourier transform mid-infrared spectrometer (FTIR), equipped with an infrared light source, transmission cell or ATR cell, interferometer, detector, and sample introduction device
Scan range and resolution? 4000cm⁻¹ ~ 600cm⁻¹; resolution 4cm⁻¹
Signal-to-noise ratio requirement? At 674cm⁻¹, S/N > 300:1 (RMS)
Noise level requirement? Within the range 700cm⁻¹ ~ 664cm⁻¹, not greater than 0.0015AU
Sample test temperature? 15℃ ~ 38℃
Sample storage temperature? 0℃ ~ 5℃
According to which standard is sampling performed? GB/T 4756
According to which standards is density determined? GB/T 1884 or SH/T 0604; converted according to GB/T 1885
What calibration method is used? Classical least squares fitting method (CLS, K matrix method), spectral region 710cm⁻¹ ~ 660cm⁻¹
Fitting region and function for the net benzene absorption peak? 691cm⁻¹ ~ 660cm⁻¹, Lorentzian band shape function with linear background, see formula (A.1)
Formula for converting mass fraction to volume fraction? φ = c₀ × (ρ₂₀ / 0.8789), see formula (1)
Density value of benzene at 20℃? 0.8789 g/cm³
Reporting precision? Accurate to 0.1% (volume fraction)
Repeatability (0.1%~1.8%) formula? r = 0.047 + 0.043X, see formula (2)
Reproducibility (0.1%~1.8%) formula? R = 0.099 + 0.031X, see formula (3)
Repeatability/reproducibility at 4% (volume fraction)? Repeatability not exceeding 0.18%; reproducibility not exceeding 0.23% (volume fraction)
Bias statement? Since no suitable reference material is available, no bias statement has been made
Quality control standard sample requirements? Prepare at least standard samples with benzene mass fraction 1.2% ± 0.2% (volume fraction approximately 1.0%); standard samples prepared from the same batch of raw materials shall meet at least 30 times QC checks
QC out-of-tolerance criterion? If the measured volume fraction differs from the known value by more than 0.12% (volume fraction), it indicates a problem with the measurement system, and measurement is not allowed before correction
How many verification standard samples are needed at minimum for instrument verification? Using the CLS method calibration method, at least 20 verification standard samples
Basis for instrument verification? According to Annex A.3, determine instrument validity based on the relevant provisions of ASTM E2056
Nature of Annex A? Normative annex
PSEQ and DOF of Table A.3? PSEQ = 0.0812; DOF (PSEQ) = 270

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.