SH/T 0108-1992 Determination of Average Molecular Weight and Molecular Weight Distribution of Certain Polymeric Additives (Size Exclusion Chromatography)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Inject the sample solution dissolved in tetrahydrofuran into a chromatographic column packed with special porous packing; in the column, the sample is separated according to molecular size, and its concentration is detected. From the molecular weight calibration curve of polystyrene standards and the viscosity parameters K and a of the polystyrene standards and sample in tetrahydrofuran at the same temperature, the molecular weight calibration curve of the sample is obtained. From the detected concentrations of the sample and their corresponding molecular weights, the average molecular weight and molecular weight distribution of the sample are calculated. The test temperature of this standard is 30℃; other temperatures may also be used if the K and a values at that temperature are available.

Universal calibration principle: the elution volume VR of a linear random coil polymer in size exclusion chromatography depends on the product of the polymer molecular weight M and the intrinsic viscosity [η]. For two polymers with different chemical properties, M₁·[η]₁ = M₂·[η]₂ (Equation 3). The intrinsic viscosity is calculated according to the Mark–Houwink equation [η] = K·Mᵃ (Equation 4). Substituting Equation (4) into Equation (3) and taking logarithms gives Equation (5), from which the molecular weight of polystyrene standards can be converted to the molecular weights of other polymers.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and Quantitative Calculation Formulas

10. Precision

For polymethacrylates, polyisobutylene and ethylene-propylene copolymers, the precision was determined by tests; the reliability of test results is judged by the following provisions (95% confidence level).

11. Report

The report shall include the following contents:

12. Reference Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key Answer Points
What is the method number of this standard? SH/T 0108-92
What samples are applicable? Linear polymeric additives soluble in tetrahydrofuran
Measurement range? Sample molecular weight corresponding to the elution volume of polystyrene standard samples with sample elution volume and molecular weight 5×10³~3×10⁶ g/mol (values require confirmation against the source standard source-text extraction)
Test temperature? 30℃ (other temperatures may also be used if K and a values are available)
What solvent is used? Tetrahydrofuran (analytical grade, purified before use)
Chromatographic column requirements? Four stainless steel columns of length 300 mm connected in series, plate number ≥4800, resolution ≥1.30
Flow rate? 1 mL/min ± 0.1 mL/min
Standard sample concentration? 1 g/L
Reference range of sample solution concentration? 5~10 g/L
Injection amount? Polymer amount generally 1~4 mg; solution volume generally not exceeding 300 μL
Baseline drift requirement? Within 3 h, not exceeding 5% of the recorder paper full scale
How to calculate number-average molecular weight? Mn(u) = ΣHi / Σ(Hi/Mi) (formula 6)
How to calculate weight-average molecular weight? Mw(u) = Σ(Hi·Mi) / ΣHi (formula 7)
How to calculate molecular weight distribution? D(u) = Mw(u) / Mn(u) (formula 8)
When can spreading effect correction be ignored? When X₂<0.05 and X₁<1.05, the correction value is less than 5%, and the apparent values can be directly used
Criterion for serious distortion of spreading effect? When X₂>0.15, the experimental conditions should be reconsidered
Repeatability requirement? The difference between two results of repeated determinations of number-average and weight-average molecular weights shall not exceed 12% of the arithmetic mean
Reproducibility requirement? Number-average molecular weight not exceeding 24%, weight-average molecular weight not exceeding 22%
Expression of results? Take the arithmetic mean of two repeated determination results, three significant figures
Which foreign standard is referenced for adoption? ASTM D3593-80(86)
Drafting unit? Refinery of Shanghai Gaoqiao Petrochemical Company
Main drafter? Deng Shuming
Technical committee? Petrochemical Research Institute