SH/T 1550-2012 Determination of purity and impurities of methyl tert-butyl ether (MTBE) for industrial use — Gas chromatography

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

3.1 Single-column (polymethylsiloxane column and 624 column) analysis method

Under the conditions specified in this standard, an appropriate amount of sample is injected into a gas chromatograph equipped with a flame ionization detector (FID) for analysis. The peak areas of each impurity and the main component are measured, and the mass percentage of each component is calculated by the corrected area normalization method. The correction factors are determined by analyzing standard samples with concentrations similar to those in the sample to be tested.

3.2 Heart-cutting analysis method

Under the conditions specified in this standard, an appropriate amount of sample is injected into a gas chromatograph equipped with a heart-cutting device and dual FID detectors. The sample first passes through a polymethylsiloxane column, and the components are separated according to boiling point. Based on the elution times of the difficult-to-separate substances, the time period for heart-cutting is determined, and they are cut to a polyethylene glycol capillary column so that the difficult-to-separate substances are effectively separated. The peak areas of each impurity and the main component are measured, and the mass percentage of each component is calculated by the corrected area normalization method. The correction factors are determined by analyzing standard samples with concentrations similar to those in the sample to be tested.

Water and other impurities that cannot be determined by this standard must be determined by corresponding standard methods, and the results obtained shall be normalized.

4. Reagents and materials (purity / preparation)

Warning: The reagents and materials involved in this chapter are all high-pressure compressed or flammable gases and flammable, toxic liquids, and should be used with attention to safety.

5. Instruments and equipment

6. Recommended operating conditions

Table 1 Recommended chromatographic columns and typical operating conditions (original table source-text extraction is misaligned; the following is organized according to the discernible content, with doubtful points marked as requiring confirmation against the source standard (source-text extraction)):

Item Polymethylsiloxane column 624 column (main analytical column) Heart-cutting auxiliary analytical column (polyethylene glycol column) Damping column (empty capillary column)
Column length/m 50 60 50 10
Liquid film thickness/μm 0.5 1.4 0.5 0.5
Inner diameter/mm 0.20 0.25 0.20 0.32
Carrier gas Helium (He) Nitrogen (N) (original text source-text extraction writes “nitrogen (N)”, requires confirmation against the source standard (source-text extraction)) Helium (He) Helium (He)
Average carrier gas flow rate/(mL/min) 0.6 0.7 0.8 2.0
Initial temperature/℃ 40 50 15 40
Initial temperature hold time/min 13 10 13 13
Heating rate/(℃/min) 10 10 10 10
Final temperature/℃ 180 180 180 180
Final temperature hold time/min 3 3 3 3
Switching time/min — — 4.7 (on); 5.1 (off); 6.6 (on); 7.4 (off); 9.4 (on); 10.4 (off) —
Injector temperature/℃ 200 200 200 —
Detector temperature/℃ 250 250 250 —
Split ratio 200:1 100:1 100:1 —
Injection volume/μL 0.1~0.5 1.0 0.4~0.8 —

Note: The three sets of switching times listed in the table respectively correspond to the heart-cutting times for the three pairs of difficult-to-separate substances: methanol and butene, tert-butanol and 2-pentene, and methyl sec-butyl ether and sec-butanol.

7. Sampling and Test Sample Preparation

8. Analytical Procedure

7 Instrument Preparation (original numbering is 7)

8 Measurement Procedure (original numbering is 8)

Table 2 MTBE Typical retention times, relative mass correction factors, and densities of samples in different analytical methods

(The original table source-text extraction is severely misaligned; only identifiable component names are listed below; retention times, correction factors, and density values cannot be reliably matched due to the misalignment and are all marked as requires confirmation against the source standard (source-text extraction))

Identifiable component names: isobutane, methanol, isobutene + 1-butene, 1-butene, n-butane, trans-2-butene, cis-2-butene, 3-methyl-1-butene, acetone, isopentane, 2-propanol, 1-pentene, 2-methyl-1-butene, n-pentane, trans-2-pentene, tert-butanol, cis-2-pentene, 2-methyl-2-butene, cyclopentene, methyl tert-butyl ether, 2,3-dimethyl-1-butene, 4-methyl-cis-2-pentene, 2-methyl-pentane, methyl ethyl ketone, 3-methylpentane, methyl sec-butyl ether, sec-butanol, ethyl tert-butyl ether, methyl-2-butenyl ether, isooctane, methyl tert-amyl ether, 3,5-dimethyl-1-hexene, 2,4,4-trimethyl-1-pentene, 2,4,4-trimethyl-2-pentene, 3,4,4-trimethyl-trans-2-pentene, 2,3,4-trimethyl-2-pentene, 4,4-dimethyl-2-neopentyl-1-pentene, 2,2,4,6,6-pentamethyl-3-heptene.

Notes to Table 2 (identifiable in the original):

9. Qualitative and Quantitative Determination and Calculation Formulas

9.1 Calculation of Mass Content

The mass content wᵢ of each component is expressed as a mass fraction, with the value expressed in %, and is calculated according to formula (2):

wᵢ = (Aᵢ × fᵢ) / Σ(Aᵢ × fᵢ) × (100 - w) (the original source-text extraction formula is garbled, requires confirmation against the source standard (source-text extraction))

Where:

For the few impurity components for which correction factors cannot be obtained, their relative mass correction factor fᵢ may be set to 1.00.

9.2 Calculation of Volume Content

If it is necessary to report the volume content Vᵢ of each component, expressed as a volume fraction, with the value expressed in %, conversion may be carried out according to formula (3):

Vᵢ = wᵢ × ρ / ρᵢ (the original source-text extraction formula is garbled, requires confirmation against the source standard (source-text extraction))

Where:

10. Precision

11 Repeatability (original number is 11)

In the same laboratory, by the same operator using the same equipment, according to the same test method, and within a short period of time, the absolute difference between two independent test results obtained from mutually independent tests on the same test object shall not exceed the repeatability limit listed in Table 3 (r), on the premise that the cases exceeding the repeatability limit do not exceed 5%.

Table 3 Repeatability requirements

Component/% (mass fraction) Repeatability limit (r)
Impurity 0.002≤w<0.020 20% of its average value
Impurity ≥0.020 10% of its average value
MTBE purity ≥99.0 0.02%
MTBE purity 99.0~ (original source-text extraction reads "99.0~43 95.0", interval unclear, requires confirmation against the source standard (source-text extraction)) 0.1%

Note: The original foreword states that "a self-determined repeatability limit (r) was adopted, and the typical data for reproducibility were deleted", therefore the reproducibility is not provided in the available source.

11. Report

The report includes the following contents:

12. Normative references

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key points of answer
What is the number and name of this standard? SH/T 1550-2012 Determination of purity and impurities of industrial methyl tert-butyl ether (MTBE) by gas chromatography
Which version does it replace? Replaces SH/T 1550-2000; previous versions SH/T 1550—93, SH/T 1550—2000
Does it adopt an international/foreign standard? Not equivalently adopts ASTM D5441-98 (2008)
Dates of publication and implementation? Published on 2012-05-24, implemented on 2012-11-01
What is it applicable to? Determination of the purity of industrial MTBE and its impurities, such as C4 to C.in hydrocarbons, methanol, isopropanol, tert-butanol and sec-butanol, methyl sec-butyl ether and methyl tert-amyl ether, acetone, and methyl ethyl ketone
Minimum detectable concentration? Polymethylsiloxane column 0.02% (mass fraction); 624 column and center cutting 0.002% (mass fraction)
Method principle? Corrected area normalization method; single-column method uses FID; center cutting method uses a center cutting device and dual FID
What is the carrier gas? Helium or nitrogen, purity ≥99.99% (volume fraction)
What is the combustion gas? Hydrogen, purity ≥99.99% (volume fraction)
Purity requirements for reference materials? The purity of MTBE shall be greater than 99.7% (mass fraction), and the purity of other substances shall be greater than 99.0% (mass fraction)
Weighing accuracy for calibration mixtures? Weigh accurately to 0.0001 g, and calculate the content accurately to 0.0001% (mass fraction)
Basis for sampling? Take samples in accordance with GB/T 4756
Sample storage conditions? Glass pressure-resistant sampling bottle, stored at a low temperature of about 4℃
What chromatographic columns are there? Polymethylsiloxane column, 624 column, center cutting auxiliary polyethylene glycol column, damping empty capillary column
Three pairs of difficult-to-separate substances in center cutting? Methanol and butene, tert-butanol and 2- pentene, methyl sec-butyl ether and sec-butanol
Correction factor formula? See formula (1), the original source-text extraction is garbled, requires confirmation against the source standard (source-text extraction)
Mass content formula? See formula (2), corrected area normalization method, the original source-text extraction is garbled, requires confirmation against the source standard (source-text extraction)
Volume content formula? See formula (3), Vᵢ = wᵢ × ρ / ρᵢ, the original source-text extraction is garbled, requires confirmation against the source standard (source-text extraction)
What if the correction factor cannot be obtained? Its relative mass correction factor fᵢ may be set to 1.00
Precision of result expression? MTBE purity accurate to 0.01%; impurity concentration by the polymethylsiloxane column method accurate to 0.01%; impurity concentration by the 624 column and center cutting method accurate to 0.001%
Basis for numerical rounding? Perform numerical rounding in accordance with GB/T 8170
Repeatability limit? See Table 3; impurity 0.002≤w<0.020 is 20% of the average value, impurity ≥0.020 is 10% of the average value; MTBE purity ≥99.0 is 0.02%, and the other interval is 0.1% (interval source-text extraction unclear, requires confirmation against the source standard (source-text extraction))
Reproducibility? Not provided in the available source; the foreword states that the typical data for reproducibility were deleted
How is water measured? Determined in accordance with GB/T 6283 and other methods, and the results are normalized
What does the report include? Sample information, standard number, analysis results, abnormal phenomena, name of the analyst, date of analysis