Origin Standard

SH/T 0197-1992 Determination of Iron Content in Lubricating Oils

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

A method

B method

C method

5. Instruments and equipment

A method

B method

C method

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis procedure

A method

B method

C method

9. Qualitative and quantitative and calculation formulas

A method

B method

C method

10. Precision

11. Report

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
How many determination methods does this standard have? Three: A method—colorimetric method, B method—gravimetric method, C method—volumetric method.
Which method is used for arbitration testing? Colorimetric method (A method).
Which oils are applicable? Lubricating oils and used lubricating oils.
What is the determination wavelength of the A method? 510nm.
What is the light path of the cuvette in the A method? 1cm.
What are the color development heating conditions for the A method? Heat in a boiling water bath for 2min.
What is the reference solution for the A method? Distilled water.
What amounts of iron standard solution are added for the working curve in the A method? 0, 0.5, 1.0, 1.5, 2.0, 2.5mL.
What amount of sample is weighed for the A method? 10-20g, accurate to 0.1g.
Which standard is followed for ashing in the A method? GB/T 508.
What is the precipitation heating temperature for the B method? 70~80℃.
What are the calcination temperature and time for the B method? 775℃±25℃, 1.5~2.0h.
What is the weighing accuracy for the B method? Accurate to 0.0002g; the constant weight difference is not greater than 0.0004g.
What volume of sample solution is taken for the B method? 25mL.
What is the standing time after adding potassium iodide in the C method? 20min.
What standard solution is used for titration in the C method? Sodium thiosulfate standard titration solution, c(Na2S2O3)=0.1mol/L.
What volume of sample solution is taken for the C method? 10mL.
When is the starch indicator solution added in the C method? Add 3mL when the liquid becomes pale yellow during titration.
How is the reported result taken? For all three methods, the arithmetic mean of two results of repeated determinations is taken.
How are the precision grades for the A method divided? 1~5ppm: arithmetic mean 30%; >5~10ppm: 10%; >10ppm: 6.5%.
What is the precision of the B method? Repeatability is not greater than 10% of the smaller result.
What is the precision of the C method? Repeatability is not greater than 15% of the smaller result (requires confirmation (source-text extraction)).
What are the referenced standards? GB/T 508 Petroleum products ash determination method.
What does this standard replace? Replaces LB E34 006-88.
In what year was it first published? 1965.