AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

After the silver probe is heated to 810°C and held for 2~3min, it is quickly quenched into the sample, and the function curve of the silver probe temperature decrease with time is continuously recorded; the cooling performance of the sample is determined from the cooling curve.

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

See section 7 (the source text lists preparation work and test procedure separately in chapters 5 and 6; this section summarizes according to the source text 6.1~6.4):

9. Qualitative and quantitative determination and calculation formulas

10. Precision

Judge the reliability of test results according to the following provisions (95% confidence level).

Table 1

Item Repeatability Reproducibility
Characteristic temperature, °C 5 10
Time from 800°C cooled to 400°C, s 0.3 0.5

11. Report

Take the arithmetic mean of the two results that meet the requirements of Chapter 8 as the characteristic temperature and the time from 800°C cooled to 400°C.

12. Normative References

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference Q&A

Question Key points of answer
What oil does this standard apply to? Heat treatment oil.
What is the principle of the method? The silver probe is heated to 810°C and held for 2~3min, then quenched into the test sample; the temperature-time cooling curve is recorded, and the cooling performance is determined from it.
Silver probe heating temperature and holding time? 810°C ± 2°C, held for 2~3min.
Sample amount and heating temperature? 250mL, heated to 80°C ± 2°C.
Quenching time requirement? Within 2s.
Recorder paper speed? 1s/mV setting.
Distance between the lower end of the silver probe and the bottom of the beaker? 15mm.
What is the calibration liquid? Dioctyl phthalate, chemically pure.
Calibration liquid cooling performance requirements? Characteristic temperature 510°C ± 10°C; time from 800°C cooled to 400°C 5.3s ± 0.3s.
How is the characteristic temperature determined from the cooling curve? Draw two tangents along the upper abrupt-change region of the curve; from the intersection point draw a line parallel to the X axis to meet the curve at one point; the temperature corresponding to that point is the characteristic temperature.
How is the time from 800°C cooled to 400°C determined? At Y axis 400°C, draw a parallel line to meet the curve at one point; the time corresponding to that point is the time from 800°C cooled to 400°C.
Precision requirements? 95% confidence level; characteristic temperature repeatability 5°C, reproducibility 10°C; time from 800°C cooled to 400°C repeatability 0.3s, reproducibility 0.5s.
What value is reported? Take the arithmetic mean of the two results that meet the requirements of Chapter 8.
Silver probe material requirements? Grade one (99.99%) or grade two (99.95%) pure silver (GB 4135).
Thermocouple wire specifications? Diameter 0.5mm nickel-chromium/nickel-silicon thermocouple wire, thermoelectric potential conforming to EU-2 (GB/T 2614).
Silver probe hot junction deviation requirement? The deviation between the centre of the thermocouple hot junction and the centre line of the silver rod diameter 10mm shall not exceed 0.2mm.
Standard oil quality indicators? Kinematic viscosity (50°C) 19–20 mm²/s; flash point (open cup) not lower than 180°C; molecular mass 310–325; water content none.
What method is Appendix B? Determination of number-average molecular mass of petroleum products and additives (vapour pressure osmometry), also called the VPO method.
Applicable molecular mass range of Appendix B? 100–2.5×10⁴ (original text is 100–2.5×10*, requires confirmation against the source standard (source-text extraction)).
Test temperature of Appendix B? Benzene as solvent 45°C; chloroform or acetone as solvent 37°C.
Which foreign standard is referenced for adoption? Japanese Industrial Standard JIS K 2242—1980 'Heat treatment oils', clause 5.2 'Cooling performance test method'.

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.