SH/T 0024-1990 Determination of Precipitation Value of Lubricating Oils

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

At room temperature, take 10mL test sample, dilute with precipitation petroleum ether to 100mL. After mixing uniformly, under the action of a relative centrifugal force of 600~700, rotate the centrifuge tube for 10min. The milliliters of precipitate formed at the bottom of the centrifuge tube is the precipitation value of the test sample.

4. Reagents and materials (purity / preparation)

Test item Requirement Determination method
Density (15℃), g/cm³ 0.692~0.702 GB/T 1884 and GB/T 1885
Aniline point, ℃ 58~60 GB/T 262
Initial boiling point, ℃ ≥50 (source text ",50", requires confirmation against the source standard (source-text extraction)) GB/T 6536
50% point, ℃ 70~80 GB/T 6536
End point, ℃ ≤130 (source text "‘130", requires confirmation against the source standard (source-text extraction)) GB/T 6536

5. Instruments and equipment

Table 1 Allowable graduation error of centrifuge tubes (mL):

Range Minimum division Volume allowable error
0.0~0.1 0.05 ±0.02
>0.1~0.3 0.05 ±0.03 (source text "[unreadable sign]0 [unreadable digits]", requires confirmation against the source standard (source-text extraction))
>0.3~0.5 0.05 ±0.05
>0.5~1.0 0.10 ±0.05
>1.0~2.0 0.10 ±0.10
>2.0~3.0 0.20 ±0.10
>3.0~5.0 0.50 ±0.20
>5.0~10 1.0 ±0.50
>10~25 5.0 ±1.00
>25~100 2 (source text "2", requires confirmation against the source standard (source-text extraction)) ±1.00

Table 2 Rotation speeds of centrifuges with different rotation diameters (the original data source-text extraction has significant noise; the following is the identifiable content, requires confirmation against the source standard (source-text extraction)):

Rotation diameter mm Rotation speed when relative centrifugal force is 600 n r/min Rotation speed when relative centrifugal force is 700 n r/min
32 1820 1960
35 1750 1890
41 1703 (requires confirmation against the source standard (source-text extraction)) 1840
45 1640 1770
48 1540 1670
50 1490 1610
53 1450 1570
56 1420 1530
59 1390 1500

(In the original Table 2, "32 5 350 370", "41[unreadable character] [unreadable character]50 483 508 533 [unreadable character]59", etc. are source-text extraction noise and cannot be reliably restored, requires confirmation against the source standard (source-text extraction))

6. Recommended operating conditions

7. Sampling and test sample preparation

8. Analysis procedure

9. Qualitative and quantitative determination and calculation formula

10. Precision

When the precipitation value is in the range of 0.00~1.20mL, use the following provisions to judge the reliability of the test results (95% confidence level).

11. Report

Take the arithmetic mean of two results from repeated determinations as the determination result.

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick reference Q&A

Question Key answer points
What does this standard apply to? Determination of precipitation value in lubricating oils.
What is the test sample amount? 10mL.
What reagent is used for dilution? Precipitation petroleum ether, dilute to the 100mL mark.
What is the centrifugation relative centrifugal force? 600~700.
How long is the centrifugation time? 10min.
Water bath temperature and time? 32~35℃, leave for 5min.
Number of inversions? Each time 20 times or more, two rounds in total.
Reading precision? Read to 0.1mL, estimate to 0.05mL.
Permissible difference between readings of the two centrifuge tubes? If not greater than 0.1mL, take the average; if greater than 0.1mL, then perform two more runs and take the average of four.
Repeatability requirement? The difference between two results by the same operator shall not exceed 10% of the average; when the sediment value is below 0.5mL, it is 0.05mL.
Reproducibility requirement? The difference between two results from different laboratories shall not exceed 30% of the average; when the sediment value is below 0.20mL, it is 0.05mL.
Density requirement for precipitation petroleum ether? At 15℃, 0.692~0.702 g/cm³.
Aniline point requirement for precipitation petroleum ether? 58~60℃.
Distillation range requirement for precipitation petroleum ether? Initial boiling point ≥50℃, 50% point 70~80℃, end point ≤130℃ (some values require confirmation against the source standard (source-text extraction)).
What standard is sampling based on? GB/T 4756.
Which foreign standard is referenced and adopted? ASTM D 91-87.