SH/T 0061-1991 Determination of magnesium content in lubricating oils (atomic absorption spectrometry)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

After the sample is diluted with solvent naphtha, it is directly sprayed into the flame of an atomic absorption spectrometer, and the absorbance is measured. The absorbance of the sample is compared with the absorbance of the reference solution to calculate the magnesium content in the sample.

4. Reagents and materials (purity / preparation)

4.2 Materials

5 Reagents

6 Preparatory work

5. Instruments and equipment

4.1 Instruments

6. Recommended operating conditions

Operate the instrument according to the conditions in Table 2:

Element Analytical line wavelength nm Lamp current mA Burner height mm Air flow rate L/min Acetylene flow rate L/min Expansion factor Extraction amount mL/min
Magnesium 285.2 5 0.38 4 10 1.5 2.8

Note: The performance of different instrument models varies; the above conditions are for reference only.

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formula

The magnesium content in the sample X[%(m/m)] is calculated according to the following formula:

X = (A / A₀) × C × f × 25 × 100 × 10⁻³ / (m × 0.1) (requires confirmation against the source standard (source-text extraction), formula layout has noise)

Where:

10. Precision

Use the following provisions to judge the reliability of test results (95% confidence level).

Repeatability: The difference between two results obtained by repeated determination by the same operator shall not exceed the value specified in Table 3.

Table 3

Magnesium content %(m/m) Percentage of arithmetic mean, %
0.02~0.3 The value at source-text extraction in the source text is missing, requires confirmation against the source standard (source-text extraction)

Reproducibility: The difference between two results submitted by different laboratories shall not exceed the value specified in Table 4.

Table 4

Magnesium content %(m/m) Percentage of arithmetic mean, %
0.02~0.3 The value at source-text extraction in the source text is missing, requires confirmation against the source standard (source-text extraction)

11. Report

Take the arithmetic mean of two repeated determination results as the determination result. If the result is equal to or greater than 0.1%(m/m), take two decimal places; if the result is less than 0.1%(m/m), take three decimal places.

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick-reference Q&A

Question Key points of answer
Standard number and year? SH/T 0061-91, approved by 1991-06-28, implemented by 1992-07-01, confirmed in (2001)
Method type? Atomic absorption spectrometry (flame method)
Applicable concentration range? 0.02%~0.3%(m/m)
Is it applicable to additives? Applicable; it can determine the magnesium content in additives
How is the sample diluted? Diluted with blank solution and made up to volume in a 25mL volumetric flask
How is the blank solution prepared? Petroleum sodium sulfonate dissolved in 1L solvent naphtha so that the sodium content is about 200 mg/mL (requires confirmation against the source standard (source-text extraction))
Concentration of magnesium standard stock solution? About 5000mg/kg
Analytical line wavelength? 285.2 nm
Lamp current? 5 mA
Air/acetylene flow rates? Air 4 L/min, acetylene 10 L/min
Burner height? 0.38 mm
Expansion factor? 1.5
Extraction amount? 2.8 mL/min
When are the standard working solutions prepared? They must be prepared on the day of the test
How are high-content samples handled? Change the optical path length, reduce the weighed sample amount, or dilute
How is trace analysis handled? Increase the sample amount to 2g, and add an appropriate amount of white oil to the standard sample to make the viscosity consistent
Calculation formula? X = (A / A₀) × C × f × 25 × 100 × 10⁻³ / (m × 0.1) (requires confirmation against the source standard (source-text extraction)))
Precision values? The values in Table 3 and Table 4 are missing at source-text extraction in the source text, requires confirmation against the source standard (source-text extraction)
Reporting rounding rules? ≥0.1%(m/m) take two decimal places; <0.1%(m/m) take three decimal places
Referenced standards? SH 0005, SH 0007
Reference adopted standard? DIN 51431 T1(86)
Drafting organization? Dalian Petrochemical Company
Main drafter? Liang Chunfang

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.